Heat generation by ultrasonic bone curette comparing with high-speed drill

被引:27
|
作者
Suzuki, Kengo [1 ,2 ]
Wanibuchi, Masahiko [1 ]
Minamida, Yoshihiro [1 ]
Akiyama, Yukinori [1 ]
Mikami, Takeshi [1 ]
Fujishige, Masahito [2 ]
Yamamura, Akinori [2 ]
Nakagawa, Toshio [2 ]
Mikuni, Nobuhiro [1 ]
机构
[1] Sapporo Med Univ, Dept Neurosurg, Sch Med, Chuo Ku, South 1,West 16, Sapporo, Hokkaido 0608543, Japan
[2] Shinsapporo Neurosurg Hosp, Dept Neurosurg, Sapporo, Hokkaido, Japan
关键词
High-speed drill; Ultrasonic bone curette; Heat generation; Thermography; CUSA; ANTERIOR CLINOIDECTOMY; SURGICAL ASPIRATOR; SPINAL SURGERY; NERVE;
D O I
10.1007/s00701-017-3445-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Ultrasonic bone curettes have been used as with high-speed drills. However, the amount of heat generated by the ultrasonic bone curette is not well known. This study quantitatively assessed the heat generated by an ultrasonic bone curette and compared it to that by a high-speed drill. The thermal change in a swine skull during bone curetting using an ultrasonic device and a high-speed drill were assessed. The investigation focused on the type of surgical manipulation (brush-like strokes vs. pushing motion) and irrigation (room temperature vs. cold water; low-volume irrigation vs. high-volume irrigation). The thermal elevation during drill use was suppressed when using brush-like strokes compared to pushing motion (brush-like strokes, 44.7 A degrees C; pushing motion, 69.2 A degrees C; p < 0.01). Cold-water irrigation while drilling had a small effect compared to room temperature (RT) water (RT, 44.7 A degrees C; cold, 35.2 A degrees C; p = 0.12). The temperature generated by the curette was higher than that generated by the drill (curette, 72.5 A degrees C; drill, 44.7 A degrees C; p < 0.01). High-volume irrigation was required to reduce the heat generated by the curette (no irrigation, 88.6 A degrees C; low-volume, 72.5 A degrees C; high-volume, 60.5 A degrees C; p < 0.01). The ultrasonic bone curate generated more heat than the high-speed drill. During surgical manipulation, the use of brush-like strokes by both the high-speed drill and the ultrasonic bone curette is necessary to avoid excess thermal elevation. Irrigation with RT water is sufficient to avoid heat generation. High-volume irrigation is required to reduce the heat generated by the curette.
引用
收藏
页码:721 / 725
页数:5
相关论文
共 28 条
  • [21] High-speed and thermal imaging of the mechanism of action of the cavitron ultrasonic surgical aspirator (CUSA)
    Verdaasdonk, R
    van Swol, C
    Grimbergen, M
    Priem, G
    SURGICAL APPLICATIONS OF ENERGY, PROCEEDINGS OF, 1998, 3249 : 72 - 84
  • [22] Autologous bone harvested during implant bed preparation: A randomized clinical trial comparing high-speed drilling with irrigation versus low-speed drilling without irrigation
    Bernabeu-Mira, Juan Carlos
    Penarrocha-Diago, Miguel
    Canullo, Luigi
    Camacho-Alonso, Fabio
    Cortes, Arthur Rodriguez Gonzalez
    Penarrocha-Oltra, David
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2024, 26 (04) : 724 - 733
  • [23] Accuracy of Cervical Pedicle Screw Insertion With and Without a Navigation-Linked High-Speed Drill: A Retrospective Clinical Study
    Ono, Kosei
    Murata, Sohei
    Matsushita, Mutsumi
    Shimizu, Yu
    Nakamura, Yusuke
    Yabe, Taisuke
    Ito, Hiromu
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (09)
  • [24] How I do it: en-bloc subaxial cervical laminectomy using a high-speed drill with a footplate attachment
    Kyle B. Mueller
    Kenneth P. Mullinix
    Hector F. Bermudez
    Acta Neurochirurgica, 2020, 162 : 311 - 315
  • [25] Pull-Force Generated by the Revolving High-speed Drill Shaft Results in Unexpected Tissue Wrapping Around the Shaft
    Shimizu, Satoru
    Miyake, Shigeta
    Mochizuki, Takahiro
    Kuroda, Hiroki
    Nakai, Yasunobu
    Kumabe, Toshihiro
    WORLD NEUROSURGERY, 2022, 166 : E353 - E357
  • [26] Heat generation and temperature prediction in metal cutting: A review and implications for high speed machining
    Abukhshim, N. A.
    Mativenga, P. T.
    Sheikh, M. A.
    INTERNATIONAL JOURNAL OF MACHINE TOOLS & MANUFACTURE, 2006, 46 (7-8): : 782 - 800
  • [27] Thermal Analysis of an Ultrasonic Aspirator Micro Claw Tool Compared With Standard High-Speed Drilling During Internal Auditory Canal Opening in a Cadaveric Model
    Maldonado, Justin
    Porto, Edoardo
    Barbero, J. Manuel Revuelta
    Tello, Isaac
    Rodas, Alejandra
    Vivas, Esther X.
    Mattox, Douglas E.
    Solares, C. Arturo
    Muvdi, Tomas Garzon
    Pradilla, Gustavo
    OPERATIVE NEUROSURGERY, 2023, 25 (02) : 183 - 189
  • [28] Aerosol morphology and particle size distribution in orthopaedic bone machining: a laboratory worst-case contamination simulation. Is high-speed bone machining potentially harmful by pollution and quality schemes and what measures could be taken for prevention?
    Putzer, David
    Dammerer, Dietmar
    Huber, Cora
    Boschert, Harald
    Thaler, Martin
    Nogler, Michael
    INTERNATIONAL ORTHOPAEDICS, 2022, 46 (07) : 1647 - 1655