On the preparation of efficiency calibration standards for gamma-ray spectrometers

被引:14
作者
Taskaeva, M [1 ]
Taskaev, E [1 ]
Penev, I [1 ]
机构
[1] KOZLODUY NPP,KOZLODUY,BULGARIA
关键词
D O I
10.1016/S0969-8043(96)00096-6
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A procedure for preparation of low-activity efficiency calibration standards in different geometries and having different densities for gamma-ray spectrometers has been developed. Natural organic and inorganic materials are used as matrices. The densities of the matrices used vary from 0.4 to 1.5 g/cm(3). The activity distribution homogeneity in the matrix after preparation and after using them for 9 months was studied. The matrix stability to settling down and selfgrinding in the process of continued usage is examined. The standards have been prepared using certified mixed-nuclide gamma-ray standard solutions, containing Am-241, Cd-109, Ce-139, Co-57, Co-60, Cs-137, Hg-203, Sn-113, Sr-85 and Y-88 and mono-nuclide solution containing (210)pb from EUROSTANDARD, traceable to the Czech Metrological Institute. Copyright (C) 1996 Elsevier Science Ltd.
引用
收藏
页码:981 / 990
页数:10
相关论文
共 10 条
[1]  
[Anonymous], RADIOACT RADIOCHEM
[2]   A RELIABLE AND ACCURATE PROCEDURE FOR PREPARING LOW-ACTIVITY EFFICIENCY CALIBRATION STANDARDS FOR GERMANIUM GAMMA-RAY SPECTROMETERS [J].
CROUDACE, IW .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-LETTERS, 1991, 153 (02) :151-162
[3]  
MCFARLAND RC, 1994, RADIOACT RADIOCHEM, V5, P6
[4]  
MCFARLAND RC, 1991, RADIOACT RADIOCHEM, V2, P4
[5]  
MCFARLAND RC, 1991, RADIOACT RADIOCHEM, V2, P6
[6]  
MCFARLAND RC, 1990, RADIOACT RADIOCHEM, V1, P18
[7]  
MCFARLAND RC, 1992, RADIOACT RADIOCHEM, V3, P6
[8]  
MCFARLAND RC, 1993, RADIOACT RADIOCHEM, V4, P4
[9]  
Sanderson C. G., 1993, RADIOACT RADIOCHEM, V4, P36
[10]  