Influence of the screw augmentation technique and a diameter increase on pedicle screw fixation in the osteoporotic spine: pullout versus fatigue testing

被引:86
作者
Kueny, Rebecca A. [1 ]
Kolb, Jan P. [2 ]
Lehmann, Wolfgang [2 ]
Pueschel, Klaus [3 ]
Morlock, Michael M. [1 ]
Huber, Gerd [1 ]
机构
[1] TUHH Hamburg Univ Technol, Inst Biomech, D-21073 Hamburg, Germany
[2] Univ Med Ctr Hamburg Eppendorf, Dept Trauma Hand & Reconstruct Surg, D-20246 Hamburg, Germany
[3] Univ Med Ctr Hamburg Eppendorf, Inst Forens Med, D-22529 Hamburg, Germany
关键词
Pedicle screw; Spine; Cement augmentation; Pullout; Fatigue; Toggle; CEMENT AUGMENTATION; BIOMECHANICAL ANALYSIS; COMPLICATIONS; STRENGTH; LUMBAR; INSTRUMENTATION; STABILIZATION; VERTEBRAE; REVISION;
D O I
10.1007/s00586-014-3476-7
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
For posterior spinal stabilization, loosening of pedicle screws at the bone-screw interface is a clinical complication, especially in the osteoporotic population. Axial pullout testing is the standard pre-clinical testing method for new screw designs although it has questioned clinical relevance. The aim of this study was to determine the fixation strength of three current osteoporotic fixation techniques and to investigate whether or not pullout testing results can directly relate to those of the more physiologic fatigue testing. Thirty-nine osteoporotic, human lumbar vertebrae were instrumented with pedicle screws according to four treatment groups: (1) screw only (control), (2) prefilled augmentation, (3) screw injected augmentation, and (4) unaugmented screws with an increased diameter. Toggle testing was first performed on one pedicle, using a cranial-caudal sinusoidal, cyclic (1.0 Hz) fatigue loading applied at the screw head. The initial compressive forces ranged from 25 to 75 N. Peak force increased stepwise by 25 N every 250 cycles until a 5.4-mm screw head displacement. The contralateral screw then underwent pure axial pullout (5 mm/min). When compared to the control group, screw injected augmentation increased fatigue force (27 %, p = 0.045) while prefilled augmentation reduced fatigue force (-7 %, p = 0.73). Both augmentation techniques increased pullout force compared to the control (ps < 0.04). Increasing the screw diameter by 1 mm increased pullout force (24 %, p = 0.19), fatigue force (5 %, p = 0.73), and induced the least stiffness loss (-29 %) from control. For the osteoporotic spine, screw injected augmentation showed the best biomechanical stability. Although pullout testing was more sensitive, the differences observed were not reflected in the more physiological fatigue testing, thus casting further doubt on the clinical relevance of pullout testing.
引用
收藏
页码:2196 / 2202
页数:7
相关论文
共 24 条
  • [1] ASTM, 2013, F543 STAND SPEC TEST, DOI [10.1520/F0543, DOI 10.1520/F0543]
  • [2] Assessment of different screw augmentation techniques and screw designs in osteoporotic spines
    Becker, S.
    Chavanne, A.
    Spitaler, R.
    Kropik, K.
    Aigner, N.
    Ogon, M.
    Redl, H.
    [J]. EUROPEAN SPINE JOURNAL, 2008, 17 (11) : 1462 - 1469
  • [3] Revision of Cannulated and Perforated Cement-Augmented Pedicle Screws A Biomechanical Study in Human Cadavers
    Bullmann, Viola
    Schmoelz, Werner
    Richter, Marcus
    Grathwohl, Corinna
    Schulte, Tobias L.
    [J]. SPINE, 2010, 35 (19) : E932 - E939
  • [4] Primary pedicle screw augmentation in osteoporotic lumbar vertebrae - Biomechanical analysis of pedicle fixation strength
    Burval, Daniel J.
    McLain, Robert F.
    Milks, Ryan
    Inceoglu, Serkan
    [J]. SPINE, 2007, 32 (10) : 1077 - 1083
  • [5] The biomechanics of pedicle screw-based instrumentation
    Cho, W.
    Cho, S. K.
    Wu, C.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (08): : 1061 - 1065
  • [6] Pedicle Screw Design and Cement Augmentation in Osteoporotic Vertebrae Effects of Fenestrations and Cement Viscosity on Fixation and Extraction
    Choma, Theodore J.
    Pfeiffer, Ferris M.
    Swope, Ryan W.
    Hirner, Jesse P.
    [J]. SPINE, 2012, 37 (26) : E1628 - E1632
  • [7] Biomechanical Analysis of Pedicle Screws in Osteoporotic Bone With Bioactive Cement Augmentation Using Simulated In Vivo Multicomponent Loading
    Choma, Theodore J.
    Frevert, Wesley F.
    Carson, William L.
    Waters, Nicole P.
    Pfeiffer, Ferris M.
    [J]. SPINE, 2011, 36 (06) : 454 - 462
  • [8] Clinical relevance of pull-out strength testing of pedicle screws
    Dawson, JM
    Boschert, P
    Macenski, MM
    Rand, N
    [J]. SPINAL IMPLANTS: ARE WE EVALUATING THEM APPROPRIATELY, 2003, 1431 : 68 - 77
  • [9] COMPLICATIONS ASSOCIATED WITH THE TECHNIQUE OF PEDICLE SCREW FIXATION - A SELECTED SURVEY OF ABS MEMBERS
    ESSES, SI
    SACHS, BL
    DREYZIN, V
    [J]. SPINE, 1993, 18 (15) : 2231 - 2239
  • [10] The use of pedicle-screw internal fixation for the operative treatment of spinal disorders
    Gaines, RW