Evaluation of the Monte Carlo method (KTMAN-2) in fluoroscopic dosimetry and comparison with experiment

被引:1
作者
Kim, Minho [1 ]
Lee, Hyounggun [1 ]
Kim, Hyosim [1 ]
Park, Hongmin [1 ]
Lee, Wonho [1 ]
Park, Sungho [2 ]
机构
[1] Korea Univ, Dept Hlth Sci Bioconvergence Engn, Seoul 136103, South Korea
[2] ASAN, Med Ctr, Seoul 138736, South Korea
基金
新加坡国家研究基金会;
关键词
MCNPX; KTMAN-2; Fluoroscopy; TLD; DOSE CALCULATIONS;
D O I
10.3938/jkps.64.936
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
This study evaluated the Monte Carlo method for determining the dose calculation in fluoroscopy by using a realistic human phantom. The dose was calculated by using Monte Carlo N-particle extended (MCNPX) in simulations and was measured by using Korean Typical Man-2 (KTMAN-2) phantom in the experiments. MCNPX is a widely-used simulation tool based on the Monte-Carlo method and uses random sampling. KTMAN-2 is a virtual phantom written in MCNPX language and is based on the typical Korean man. This study was divided into two parts: simulations and experiments. In the former, the spectrum generation program (SRS-78) was used to obtain the output energy spectrum for fluoroscopy; then, each dose to the target organ was calculated using KTMAN-2 with MCNPX. In the latter part, the output of the fluoroscope was calibrated first and TLDs (Thermoluminescent dosimeter) were inserted in the ART (Alderson Radiation Therapy) phantom at the same places as in the simulation. Thus, the phantom was exposed to radiation, and the simulated and the experimental doses were compared. In order to change the simulation unit to the dose unit, we set the normalization factor (NF) for unit conversion. Comparing the simulated with the experimental results, we found most of the values to be similar, which proved the effectiveness of the Monte Carlo method in fluoroscopic dose evaluation. The equipment used in this study included a TLD, a TLD reader, an ART phantom, an ionization chamber and a fluoroscope.
引用
收藏
页码:936 / 940
页数:5
相关论文
共 12 条
  • [1] Abdalla N., 2006, RADIAT PROT DOSIM, V119, P430
  • [2] DOSE REDUCTION IN CT USING BISMUTH SHIELDING: MEASUREMENTS AND MONTE CARLO SIMULATIONS
    Chang, Kyung-Hwan
    Lee, Wonho
    Choo, Dong-Myung
    Lee, Choon-Sik
    Kim, Youhyun
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 138 (04) : 382 - 388
  • [3] Reference dose rates for fluoroscopy guided interventions
    Geleijns, J
    Broerse, JJ
    Hummel, WA
    Schalij, MJ
    [J]. RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) : 135 - 138
  • [4] MONTE CARLO ESTIMATION OF DOSE DIFFERENCE IN LUNG FROM 192Ir BRACHYTHERAPY DUE TO TISSUE INHOMOGENEITY
    Gialousis, G.
    Dimitriadis, A.
    Yakoumakis, E.
    [J]. RADIATION PROTECTION DOSIMETRY, 2011, 147 (1-2) : 287 - 290
  • [5] Goorley T., 2007, LAUR072777 LOS AL NA
  • [6] Modelling and Monte Carlo organ dose calculations for workers walking on ground contaminated with Cs-137 and Co-60 gamma sources
    Han, Bin
    Zhang, Juying
    Na, Yong Hum
    Caracappa, Peter F.
    Xu, X. George
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 141 (03) : 299 - 304
  • [7] Development of the two Korean adult tomographic computational phantoms for organ dosimetry
    Lee, C
    Lee, C
    Park, SH
    Lee, JK
    [J]. MEDICAL PHYSICS, 2006, 33 (02) : 380 - 390
  • [8] Dose conversion coefficients calculated using tomographic phantom, KTMAN-2, for X-ray examination of cardiac catheterisation
    Park, S. H.
    Lee, J. K.
    Lee, C.
    [J]. RADIATION PROTECTION DOSIMETRY, 2008, 128 (03) : 351 - 358
  • [9] Weighting of secondary radiations in organ dose calculations
    Siiskonen, T.
    Tapiovaara, M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 141 (01) : 18 - 26
  • [10] Dose evaluation in fluoroscopy
    Verdun, FR
    Capasso, P
    Valley, JF
    Schnyder, P
    [J]. RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) : 139 - 141