Enhancing capabilities of aspiration-type Ion Mobility Spectrometer using a Pulsed Sampling System and a heated transfer line

被引:2
作者
Mikedi, K. [1 ]
Pallis, G. C. [1 ]
Koumoundouros, G. C. [1 ]
Vamvakari, J. [1 ]
Statheropoulos, G. [1 ]
Psarras, G. [1 ]
Moll, V. H. [1 ]
McEntee, R. [1 ]
Statheropoulos, M. [1 ]
机构
[1] NTUA, Sch Chem Engn, Field Analyt Chem & Technol Unit, Athens 15773, Greece
关键词
Ion mobility spectrometry; Aspiration IMS; Sampling system; Closed-loop IMS; Chemical environment generator; GAS-CHROMATOGRAPHY; MASS-SPECTROMETRY; HYPHENATED INSTRUMENTATION; PLASMA CHROMATOGRAPHY; EXPONENTIAL DILUTION; VAPORS; WATER;
D O I
10.1016/j.snb.2015.08.055
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An aspiration-type Ion Mobility Spectrometer (IMS) was connected to an in-house developed Pulsed Sampling System (PSS) through a heated capillary column acting as transfer line, after being modified to a closed loop IMS device. Performance of the whole system was evaluated using acetone gas standards in the concentration range of 200-2000 ppb. Tests were carried out using an in-house developed chemical environment generator for producing acetone concentrations. The system presented long-term stability, linearity, sensitivity and low minimum detectable level; less than 286 ppb for acetone with a closed loop that circulates 1300 mL/min of purified air. Repeatability (RSD) was less than 15% for all concentrations. All data were produced in dry atmosphere inside the IMS; presence of moisture (e.g. when saturation of molecular sieves occurs) seems to severely reduce performance. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:240 / 248
页数:9
相关论文
共 20 条
[1]  
[Anonymous], 2005, ION MOBILITY SPECTRO, DOI DOI 10.1201/9781420038972
[2]  
Arnold NS, 2000, FIELD ANAL CHEM TECH, V4, P219, DOI 10.1002/1520-6521(2000)4:5<219::AID-FACT2>3.0.CO
[3]  
2-7
[4]   DETERMINATION OF 2,4-DICHLOROPHENOXYACETIC ACID IN SOILS BY CAPILLARY GAS-CHROMATOGRAPHY WITH ION-MOBILITY DETECTION [J].
BAIM, MA ;
HILL, HH .
JOURNAL OF CHROMATOGRAPHY, 1983, 279 (NOV) :631-642
[5]   PERFORMANCE ADVANCES IN ION MOBILITY SPECTROMETRY THROUGH COMBINATION WITH HIGH-SPEED VAPOR SAMPLING, PRECONCENTRATION AND SEPARATION TECHNIQUES [J].
DWORZANSKI, JP ;
KIM, MG ;
SNYDER, AP ;
ARNOLD, NS ;
MEUZELAAR, HLC .
ANALYTICA CHIMICA ACTA, 1994, 293 (03) :219-235
[6]   ION MOBILITY SPECTROMETRY AS FLOW-INJECTION DETECTOR AND CONTINUOUS-FLOW MONITOR FOR ANILINE IN HEXANE AND WATER [J].
EICEMAN, GA ;
GARCIAGONZALEZ, L ;
WANG, YF ;
PITTMAN, B ;
BURROUGHS, GE .
TALANTA, 1992, 39 (05) :459-467
[7]   ENHANCED SELECTIVITY IN ION MOBILITY SPECTROMETRY ANALYSIS OF COMPLEX-MIXTURES BY ALTERNATE REAGENT GAS CHEMISTRY [J].
EICEMAN, GA ;
WANG, YF ;
GARCIAGONZALEZ, L ;
HARDEN, CS ;
SHOFF, DB .
ANALYTICA CHIMICA ACTA, 1995, 306 (01) :21-33
[8]   DETECTION OF 2,4,6-TRINITROTOLUENE VAPORS IN AIR BY PLASMA CHROMATOGRAPHY [J].
KARASEK, FW ;
DENNEY, DW .
JOURNAL OF CHROMATOGRAPHY, 1974, 93 (01) :141-147
[9]   IMPROVED EXPONENTIAL DILUTION FLASK FOR GAS CHROMATOGRAPHY [J].
LORENZ, LJ ;
CULP, RA ;
DIXON, RT .
ANALYTICAL CHEMISTRY, 1970, 42 (09) :1119-&
[10]   Ion Mobility Spectrometry and Its Applications in Detection of Chemical Warfare Agents [J].
Makinen, Marko A. ;
Anttalainen, Osmo A. ;
Sillanpaa, Mika E. T. .