Near-Infrared Diffuse Reflectance Measurement Method Based on Temperature-Insensitive Radial Distance

被引:3
作者
Wu, Minglei [1 ]
Liu, Rong [1 ]
Xu, Kexin [1 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
基金
中国国家自然科学基金; 国家高技术研究发展计划(863计划);
关键词
Near-infrared diffuse reflectance; temperature; absorption coefficient; scattering coefficient; temperature-insensitive radial distance; WAVELENGTH RANGE; GLUCOSE; SPECTROSCOPY; WATER; NM;
D O I
10.1177/0003702818766555
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The variation of temperature is one of the main interference factors that affect the detection accuracy of near-infrared (NIR) diffuse reflectance. In this paper, a measurement method based on temperature-insensitive radial distance was proposed, and its feasibility and effectiveness were verified in Intralipid solutions. First, the possibility of temperature-insensitive radial distance was deduced based on the analytic solution of the steady-state diffusion equation in an infinite media, and the temperature-insensitive radial distances of 3% Intralipid solution in the wavelength range of 1000-1600nm was calculated. Second, a detection system was designed to measure the diffuse reflectance of 3% Intralipid solutions at multiple radial distances with different glucose concentration (0-100mM) and different temperatures (35-40?). Both theoretical calculations and experimental results demonstrated the existence of temperature-insensitive radial distances in the range of 1000-1340nm and 1440-1600nm, and the distances were hardly affected by glucose variations. Finally, the glucose information extracted from the diffuse reflectance of Intralipid solutions at different radial distances under random temperature variations and constant temperature were compared. The result showed that the correlation between the glucose concentration and the diffuse reflectance obtained at the temperature-insensitive radial distance was significantly better than that of other radial distances, which was almost close to the situation of constant temperature. Therefore, the measurement method based on temperature-insensitive radial distance can effectively reduce the influence of temperature variations on NIR diffuse reflectance, and it is expected to improve the accuracy of diffuse reflectance in human body components detection and industrial field analysis.
引用
收藏
页码:1021 / 1028
页数:8
相关论文
共 19 条
  • [1] [Anonymous], 2002, PHARM MED APPL NEARI, DOI DOI 10.1201/9780203910153
  • [2] A primary method for determination of optical parameters of turbid samples and application to intralipid between 550 and 1630nm
    Chen, Cheng
    Lu, Jun Q.
    Ding, Huafeng
    Jacobs, Kenneth M.
    Du, Yong
    Hu, Xin-Hua
    [J]. OPTICS EXPRESS, 2006, 14 (16): : 7420 - 7435
  • [3] The impact of temperature variations on spectroscopic calibration modelling: a comparative study
    Chen, Tao
    Martin, Elaine
    [J]. JOURNAL OF CHEMOMETRICS, 2007, 21 (5-6) : 198 - 207
  • [4] A Reference-Wavelength-Based Method for Improved Analysis of Near-Infrared Spectroscopy
    Chen, Yun
    Chen, Wenliang
    Shi, Zhenzhi
    Yang, Yue
    Xu, Kexin
    [J]. APPLIED SPECTROSCOPY, 2009, 63 (05) : 544 - 548
  • [5] TEMPERATURE-INSENSITIVE NEAR-INFRARED SPECTROSCOPIC MEASUREMENT OF GLUCOSE IN AQUEOUS-SOLUTIONS
    HAZEN, KH
    ARNOLD, MA
    SMALL, GW
    [J]. APPLIED SPECTROSCOPY, 1994, 48 (04) : 477 - 483
  • [6] Influence of temperature on water and aqueous glucose absorption spectra in the near- and mid-infrared regions at physiologically relevant temperatures
    Jensen, PS
    Bak, J
    Andersson-Engels, S
    [J]. APPLIED SPECTROSCOPY, 2003, 57 (01) : 28 - 36
  • [7] Khalil Omar S, 2004, Diabetes Technol Ther, V6, P660, DOI 10.1089/dia.2004.6.660
  • [8] Langford VS, 2001, J PHYS CHEM A, V105, P8916, DOI 10.1021/jp0l0093m
  • [9] Liu H, 1996, J Biomed Opt, V1, P200, DOI 10.1117/12.231370
  • [10] Liu R., 2008, P SPIE