In Vivo Contact Force Analysis and Correlation With Tissue Impedance During Left Atrial Mapping and Catheter Ablation of Atrial Fibrillation

被引:48
作者
Makimoto, Hisaki [1 ]
Lin, Tina [1 ]
Rillig, Andreas [1 ]
Metzner, Andreas [1 ]
Wohlmuth, Peter [2 ]
Arya, Anita [1 ]
Antz, Matthias [1 ]
Mathew, Shibu [1 ]
Deiss, Sebastian [1 ]
Wissner, Erik [1 ]
Rausch, Peter [1 ]
Bardyszewski, Aleksander [1 ]
Kamioka, Masashi [1 ]
Li, Xuping [1 ]
Kuck, Karl-Heinz [1 ]
Ouyang, Feifan [1 ]
Tilz, Roland Richard [1 ]
机构
[1] Asklepios Klin St Georg, Dept Cardiol, D-20099 Hamburg, Germany
[2] Asklepios Klin St Georg, D-20099 Hamburg, Germany
关键词
RADIOFREQUENCY LESION SIZE; PULMONARY VEIN ISOLATION; DOUBLE-LASSO TECHNIQUE; ELECTRODE; TOCCATA;
D O I
10.1161/CIRCEP.113.000556
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-The aim of this study was to evaluate in vivo contact force (CF) and the correlation of CF with impedance during left atrial 3-dimensional electroanatomical mapping and ablation. Methods and Results-CF during point-by-point left atrial mapping was assessed in 30 patients undergoing atrial fibrillation ablation. Operators were blinded to the real-time CF data. Data were analyzed according to 11 predefined areas in the left atrial and 6 segments around the ipsilateral pulmonary veins. A total of 3475 mapping and 878 ablation points were analyzed. Median CF during mapping was 14.0g (6.5-26.2; q1-q3), ranging from 5.1g at the ridge to 29.8g at the roof. Median CF at the ridge and mitral isthmus were 5.1g and 6.9g, respectively. Extremely high CF ≥100g was noted in 24 points (0.7%). Median CFs during ablation around the right and left pulmonary veins were 22.8g (12.6-37.9; q1-q3) and 12.3g (6.9-30.2; q1-q3), respectively. The lowest median CFs were recorded at the anterior-superior and anterior- inferior segments of the left pulmonary veins (7.2g and 7.9g). Impedance values during mapping and impedance fall during ablation correlated with the applied CF (R2=0.16; P<0.001 and R2=0.04; P<0.001) although there was significant overlap. Conclusions-Excessively high and low CF values can be observed during left atrial mapping and ablation. The low CF obtained at the mitral isthmus and anterior segments of the left pulmonary veins may explain why reconnection after ablation occurs more frequently at these sites. CF and impedance do correlate; however, the impedance for a given CF ranges widely, limiting its use in clinical practice. © 2013 American Heart Association, Inc.
引用
收藏
页码:46 / 54
页数:9
相关论文
共 18 条
[1]   Radiofrequency catheter ablation with the split-tip electrode in the temperature-controlled mode [J].
Antz, M ;
Otomo, K ;
Nakagawa, H ;
Yamanashi, WS ;
Jackman, WM ;
Kuck, KH .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (12) :1765-1773
[2]   HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: Recommendations for personnel, policy, procedures and follow-up [J].
Calkins, Hugh ;
Brugada, Josep ;
Packer, Douglas L. ;
Cappato, Riccardo ;
Chen, Shih-Ann ;
Crijns, Harry J. G. ;
Damiano, Ralph J., Jr. ;
Davies, D. Wyn ;
Haines, David E. ;
Haissaguerre, Michel ;
Lesaka, Yoshito ;
Jackman, Warren ;
Jais, Pierre ;
Kottkamp, Hans ;
Kuck, Karl Heinz ;
Lindsay, Bruce D. ;
Marchlinski, Francis E. ;
McCarthy, Patrick M. ;
Mont, J. Lluis ;
Morady, Fred ;
Nademanee, Koonlawee ;
Natale, Andrea ;
Pappone, Carlo ;
Prystowsky, Eric ;
Raviele, Antonio ;
Ruskin, Jeremy N. ;
Shemin, Richard J. .
EUROPACE, 2007, 9 (06) :335-379
[3]   Relationship Between Catheter Forces, Lesion Characteristics, "Popping," and Char Formation: Experience with Robotic Navigation System [J].
Di Biase, Luigi ;
Natale, Andrea ;
Barrett, Conor ;
Tan, Carmela ;
Elayi, Claude S. ;
Ching, Chi Keong ;
Wang, Paul ;
Al-Ahmad, Amin ;
Arruda, Mauricio ;
Burkhardt, J. David ;
Wisnoskey, Brian J. ;
Chowdhury, Punam ;
De Marco, Shari ;
Armaganijan, Luciana ;
Litwak, Kenneth N. ;
Schweikert, Robert A. ;
Cummings, Jennifer E. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2009, 20 (04) :436-440
[4]   Ipsilateral pulmonary vein isolation performed by a single continuous circular lesion: role of pulmonary vein mapping during ablation [J].
Fuernkranz, Alexander ;
Julian, K. R. Chun Julian ;
Schmidt, Boris ;
Wohlmuth, Peter ;
Tilz, Roland ;
Kuck, Karl-Heinz ;
Ouyang, Feifan .
EUROPACE, 2011, 13 (07) :935-941
[5]   Interrelation of tissue temperature versus flow velocity in two different kinds of temperature-controlled catheter radiofrequency energy applications [J].
Gumbrecht, S ;
Neuzner, J ;
Pitschner, HF .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 1998, 2 (02) :211-219
[6]   A novel radiofrequency ablation catheter using contact force sensing: Toccata study [J].
Kuck, Karl-Heinz ;
Reddy, Vivek Y. ;
Schmidt, Boris ;
Natale, Andrea ;
Neuzil, Petr ;
Saoudi, Nadir ;
Kautzner, Josef ;
Herrera, Claudia ;
Hindricks, Gerhard ;
Jais, Pierre ;
Nakagawa, Hiroshi ;
Lambert, Hendrik ;
Shah, Dipen C. .
HEART RHYTHM, 2012, 9 (01) :18-23
[7]   Prospective Characterization of Catheter-Tissue Contact Force at Different Anatomic Sites During Antral Pulmonary Vein Isolation [J].
Kumar, Saurabh ;
Morton, Joseph B. ;
Lee, Justin ;
Halloran, Karen ;
Spence, Steven J. ;
Gorelik, Alexandra ;
Hepworth, Graham ;
Kistler, Peter M. ;
Kalman, Jonathan M. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2012, 5 (06) :1124-1129
[8]   Inverse relationship between electrode size and lesion size during radiofrequency ablation with active electrode cooling [J].
Nakagawa, H ;
Wittkampf, FHM ;
Yamanashi, WS ;
Pitha, JV ;
Imai, S ;
Campbell, B ;
Arruda, M ;
Lazzara, R ;
Jackman, WM .
CIRCULATION, 1998, 98 (05) :458-465
[9]   COMPARISON OF IN-VIVO TISSUE TEMPERATURE PROFILE AND LESION GEOMETRY FOR RADIOFREQUENCY ABLATION WITH A SALINE-IRRIGATED ELECTRODE VERSUS TEMPERATURE CONTROL IN A CANINE THIGH MUSCLE PREPARATION [J].
NAKAGAWA, H ;
YAMANASHI, WS ;
PITHA, JV ;
ARRUDA, M ;
WANG, XZ ;
OHTOMO, K ;
BECKMAN, KJ ;
MCCLELLAND, JH ;
LAZZARA, R ;
JACKMAN, WM .
CIRCULATION, 1995, 91 (08) :2264-2273
[10]   A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention [J].
Okumura, Yasuo ;
Johnson, Susan B. ;
Bunch, T. Jared ;
Henz, Benhur D. ;
O'Brien, Christine J. ;
Packer, Douglas L. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2008, 19 (06) :632-640