Application of electron-attachment reactions to enhance selectivity of electron-capture detector for nitroaromatic explosives

被引:1
作者
Gehrke, M [1 ]
Kapila, S
Flanigan, V
机构
[1] Univ Missouri, Ctr Environm Sci & Technol, Rolla, MO 65409 USA
[2] Univ Missouri, Dept Chem, Rolla, MO 65409 USA
关键词
electron-capture detection; detection; GC; instrumentation; explosives; nitroaromatic compounds; halogenated compounds; nitrotoluenes; nitrobenzenes; chlorobenzenes; pesticides;
D O I
10.1016/S0021-9673(00)00123-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The differences in the extent of electron-attachment reactions between thermal electrons and selected classes of organic molecules with high electron affinities were investigated. The investigations showed that interactions of thermal electrons with nitroaromatic compounds lead to the formation of neutral products with very low electron affinities. By contrast, a number of other analytes with high electron affinities such as polyhalogenated organic compounds, lead to products with high electron affinities. This difference was exploited to differentiate between nitroaromatic and polychlorinated organic compounds with a tandem arrangement consisting of two electron-capture detectors connected in series with an electron-attachment reactor. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:133 / 139
页数:7
相关论文
共 14 条
  • [1] SELECTIVE DETECTION OF BROMINATED HYDROCARBONS BY THE PHOTODETACHMENT-MODULATED ELECTRON-CAPTURE DETECTOR
    BOGNAR, JA
    KNIGHTON, WB
    GRIMSRUD, EP
    [J]. ANALYTICAL CHEMISTRY, 1992, 64 (20) : 2451 - 2455
  • [2] PERFORMANCE OF A SOLUTE SWITCHED ELECTRON-CAPTURE DETECTOR FOR PESTICIDES
    BOSHOFF, PR
    HOPKINS, BJ
    [J]. JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1979, 17 (10) : 588 - 594
  • [3] Mass spectrometric characterization of electron attachment reaction products
    Gehrke, M
    Kapila, S
    Nam, P
    Flanigan, V
    [J]. DETECTION AND REMEDIATION TECHNOLOGIES FOR MINES AND MINELIKE TARGETS IV, PTS 1 AND 2, 1999, 3710 : 299 - 306
  • [4] GRIMSRUD EP, 1992, DETECTORS CAPILLARY
  • [5] HANISCH RC, 1985, PB85158145GAR, P82
  • [6] VOLTAGE EFFECTS IN A DC ELECTRON-CAPTURE DETECTOR
    KAPILA, S
    AUE, WA
    [J]. JOURNAL OF CHROMATOGRAPHY, 1975, 108 (01): : 13 - 21
  • [7] KAPILA S, 1979, THESIS DALHOUSIE U
  • [8] KEITH LH, 1983, P AWWA WAT QUAL TECH, P197
  • [9] SOLUTE SWITCHING AND DETECTION BY SYNCHRONOUS DEMODULATION IN GAS-CHROMATOGRAPHY
    LOVELOCK, JE
    [J]. JOURNAL OF CHROMATOGRAPHY, 1975, 112 (OCT29): : 29 - 36
  • [10] ELECTRON-CAPTURE DETECTOR - THEORY AND PRACTICE