Feasibility of a Portable Bihormonal Closed-Loop System to Control Glucose Excursions at Home Under Free-Living Conditions for 48 Hours

被引:55
作者
van Bon, Arianne C. [1 ]
Luijf, Yoeri M. [1 ]
Koebrugge, Rob [2 ]
Koops, Robin [2 ]
Hoekstra, Joost B. L. [1 ]
DeVries, J. Hans [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Internal Med, NL-1105 AZ Amsterdam, Netherlands
[2] Inreda Diabet BV, Goor, Netherlands
关键词
ARTIFICIAL PANCREAS; HYPOGLYCEMIA; GLUCAGON; INSULIN; FREQUENCY; AWARENESS; DELIVERY; ADULTS;
D O I
10.1089/dia.2013.0166
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: This study assessed the feasibility of a portable bihormonal closed-loop system at home. Subjects and Methods: Sixteen pump-treated patients with type 1 diabetes received 48 h of closed-loop therapy with a telemonitored insulin- and glucagon-delivering closed-loop system and 48 h of patient-managed open-loop therapy. Results: Owing to technical problems in five cases, only 11 patients could be analyzed. Whereas median (interquartile range) glucose levels were not significantly different during Day 1 of open-loop control (OL1) from closed-loop control (CL1) (8.27 [0.83] mmol/L vs. 8.84 [1.47] mmol/L; P=0.206), they were significantly lower during Day 2 of closed-loop control (CL2) versus open-loop control (OL2) (7.70 [2.29] mmol/L vs. 8.84 [0.87] mmol/L; P=0.027). Time spent in euglycemia (3.9-10 mmol/L) was comparable with 67.2% (38.5%) in OL1 versus 79.2% (16.9%) in CL1 (P=0.189) and 66.0% (29.8%) in OL2 versus 76.5% (23.9%) in CL2 (P=0.162). Time spent in hypoglycemia (<3.9 mmol/L) was comparable on Day 1 of control (OL1, 0.68% [8.68%]; CL1, 2.08% [7.61%]; P=0.593) but significantly higher during Day 2 of control (OL2, 0.00% [11.07%]; CL2, 2.8% [9.8%]; P=0.0172) (P=0.017). Conclusions: Bihormonal closed-loop control is feasible at home, with comparable time in euglycemia to open-loop control and significantly lower median glucose levels on Day 2 of control at the expense of more time in hypoglycemia, albeit still at a very low percentage of time.
引用
收藏
页码:131 / 136
页数:6
相关论文
共 16 条
[1]   Fully Integrated Artificial Pancreas in Type 1 Diabetes: Modular Closed-Loop Glucose Control Maintains Near Normoglycemia [J].
Breton, Marc ;
Farret, Anne ;
Bruttomesso, Daniela ;
Anderson, Stacey ;
Magni, Lalo ;
Patek, Stephen ;
Man, Chiara Dalla ;
Place, Jerome ;
Demartini, Susan ;
Del Favero, Simone ;
Toffanin, Chiara ;
Hughes-Karvetski, Colleen ;
Dassau, Eyal ;
Zisser, Howard ;
Doyle, Francis J., III ;
De Nicolao, Giuseppe ;
Avogaro, Angelo ;
Cobelli, Claudio ;
Renard, Eric ;
Kovatchev, Boris .
DIABETES, 2012, 61 (09) :2230-2237
[2]   Novel Use of Glucagon in a Closed-Loop System for Prevention of Hypoglycemia in Type 1 Diabetes [J].
Castle, Jessica R. ;
Engle, Julia M. ;
Youssef, Joseph E. L. ;
Massoud, Ryan G. ;
Yuen, Kevin C. J. ;
Kagan, Ryland ;
Ward, W. Kenneth .
DIABETES CARE, 2010, 33 (06) :1282-1287
[3]  
Clarke William L, 2009, J Diabetes Sci Technol, V3, P1031
[4]   REDUCED AWARENESS OF HYPOGLYCEMIA IN ADULTS WITH IDDM - A PROSPECTIVE-STUDY OF HYPOGLYCEMIC FREQUENCY AND ASSOCIATED SYMPTOMS [J].
CLARKE, WL ;
COX, DJ ;
GONDERFREDERICK, LA ;
JULIAN, D ;
SCHLUNDT, D ;
POLONSKY, W .
DIABETES CARE, 1995, 18 (04) :517-522
[5]   Improved glycaemic control in type 1 diabetes patients following participation per se in a clinical trial -: mechanisms and implications [J].
DeVries, JH ;
Snoek, FJ ;
Kostense, PJ ;
Heine, RJ .
DIABETES-METABOLISM RESEARCH AND REVIEWS, 2003, 19 (05) :357-362
[6]   FREQUENCY OF SEVERE HYPOGLYCEMIA IN PATIENTS WITH TYPE-1 DIABETES WITH IMPAIRED AWARENESS OF HYPOGLYCEMIA [J].
GOLD, AE ;
MACLEOD, KM ;
FRIER, BM .
DIABETES CARE, 1994, 17 (07) :697-703
[7]   Pharmacokinetic and glucodynamic comparisons of recombinant and animal-source glucagon after IV, IM, and SC injection in healthy volunteers [J].
Graf, CJ ;
Woodworth, JR ;
Seger, ME ;
Holcombe, JH ;
Bowsher, RR ;
Lynch, R .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1999, 88 (10) :991-995
[8]   Glucose-responsive insulin and glucagon delivery (dual-hormone artificial pancreas) in adults with type 1 diabetes: a randomized crossover controlled trial [J].
Haidar, Ahmad ;
Legault, Laurent ;
Dallaire, Maryse ;
Alkhateeb, Ammar ;
Coriati, AdSLe ;
Messier, Virginie ;
Cheng, Peiyao ;
Millette, Maude ;
Boulet, Benoit ;
Rabasa-Lhoret, Remi .
CANADIAN MEDICAL ASSOCIATION JOURNAL, 2013, 185 (04) :297-305
[9]   Closed-loop insulin delivery: from bench to clinical practice [J].
Hovorka, Roman .
NATURE REVIEWS ENDOCRINOLOGY, 2011, 7 (07) :385-395
[10]   Manual closed-loop insulin delivery in children and adolescents with type 1 diabetes: a phase 2 randomised crossover trial [J].
Hovorka, Roman ;
Allen, Janet M. ;
Elleri, Daniela ;
Chassin, Ludovic J. ;
Harris, Julie ;
Xing, Dongyuan ;
Kollman, Craig ;
Hovorka, Tomas ;
Larsen, Anne Mette F. ;
Nodale, Marianna ;
De Palma, Alessandro ;
Wilinska, Malgorzata E. ;
Acerini, Carlo L. ;
Dunger, David B. .
LANCET, 2010, 375 (9716) :743-751