Air leakage around endotracheal tube cuffs

被引:46
作者
Dullenkopf, A [1 ]
Schmitz, A [1 ]
Frei, M [1 ]
Gerber, AC [1 ]
Weiss, M [1 ]
机构
[1] Univ Childrens Hosp, Dept Anaesthesia, CH-8032 Zurich, Switzerland
关键词
anaesthesia; intubation; intratracheal; cuff;
D O I
10.1017/S0265021504006064
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background and objective: To compare the recently introduced Microcuff(R) endotracheal tube with conventional tubes in respect of the cuff pressures required to prevent air leakage. Methods: The following tubes (ID 7.0 mm) were compared: Microcuff(R) HVLP ICU, Mallinckrodt HiLo(R), Portex Profile Soft Seal(R), Rusch Super Safety Clear(R) and Sheridan CF(R). Fifty patients undergoing endotracheal intubation with a cuffed tube of internal diameter 7.0 mm were studied. Tracheas were intubated using one of the endotracheal tubes in random order. Cuff pressure to prevent air leakage at standardized ventilator setting (peak inspiratory pressure 20 cmH(2)O/PEEP 5 cmH(2)O/respiratory rate 15 breaths min 1) was assessed by auscultation of audible sounds at the mouth. Patients characteristics and cuff pressures from each brand were compared to the Microcuff(R) group using the Mann-Whitney U-test (P < 0.05 was chosen as the level of statistical significance). Results: Patients' median age (range) was 14.2 (12.0-17.1) yr, body weight 57.5 (40.0-81.9) kg and length 164.9 (146.5-190.0) cm. No significant differences in patients' characteristics were found between groups. Mean cuff pressure (all tubes) required for air sealing was 19.1 (8-42) cmH(2)O. The Microcuff(C) tube required significantly lower sealing pressures (9.5 (8-12) cmH(2)O) compared to the other brands of enclotracheal tube (P < 0.05, Mann-Whitney U-test). Conclusion: The Microcuff(C) enclotracheal tube with its ultra-thin polyurethane cuff membrane required the lowest sealing pressure to prevent air leakage. These features are potentially of interest for long-term intubated patients and for cuffed enclotracheal tubes in children, allowing tracheal sealing at lower cuff pressures implying less damage to the trachea.
引用
收藏
页码:448 / 453
页数:6
相关论文
共 25 条
  • [1] Leakage of fluid around high-volume, low-pressure cuffs - A comparison of four tracheal tubes
    Asai, T
    Shingu, K
    [J]. ANAESTHESIA, 2001, 56 (01) : 38 - 42
  • [2] Gel lubrication of the tracheal tube cuff reduces pulmonary aspiration
    Blunt, MC
    Young, PJ
    Patil, A
    Haddock, A
    [J]. ANESTHESIOLOGY, 2001, 95 (02) : 377 - 381
  • [3] Intracuff Pressure and Tracheal Morbidity - Influence of Filling Cuff with Saline during Nitrous Oxide Anesthesia
    Combes, X
    Schauvliege, F
    Peyrouset, O
    Motamed, C
    Kirov, K
    Dhonneur, G
    Duvaldestin, P
    [J]. ANESTHESIOLOGY, 2001, 95 (05) : 1120 - 1124
  • [4] PREVENTION OF HOSPITAL-ACQUIRED PNEUMONIA - MEASURING EFFECT IN OUNCES, POUNDS, AND TONS
    CRAVEN, DE
    [J]. ANNALS OF INTERNAL MEDICINE, 1995, 122 (03) : 229 - 231
  • [5] CUFFED ENDOTRACHEAL-TUBES - MUCOSAL PRESSURES AND TRACHEAL WALL BLOOD-FLOW
    DOBRIN, P
    CANFIELD, T
    [J]. AMERICAN JOURNAL OF SURGERY, 1977, 133 (05) : 562 - 568
  • [6] DULLENKOPF A, 2003, EUR J ANAESTHESI S30, V20, pA251
  • [7] Ferdinande P., 1995, Acta Oto-Rhino-Laryngologica Belgica, V49, P341
  • [8] Influence of airway pressure on minimum occlusive endotracheal tube cuff pressure
    Guyton, DC
    Barlow, MR
    Besselievre, TR
    [J]. CRITICAL CARE MEDICINE, 1997, 25 (01) : 91 - 94
  • [9] Cuffed tubes in children
    James, I
    [J]. PAEDIATRIC ANAESTHESIA, 2001, 11 (03): : 259 - 263
  • [10] LATTO P, 1997, DIFFICULTIES TRACHEA, P51