共 1 条
A widely applicable plasma renin activity assay by LC-MS/MS with offline solid phase extraction
被引:15
作者:
Owen, Laura J.
[1
,2
]
Adaway, Joanne
[1
,2
]
Morris, Karen
[1
]
Lockhart, Stephen
[1
]
Keevil, Brian G.
[1
,2
]
机构:
[1] Univ S Manchester Hosp, Dept Biochem, Manchester M23 9LT, Lancs, England
[2] Univ Manchester, Manchester Acad Hlth Sci Ctr, Manchester, Lancs, England
关键词:
LC-MS/MS;
solid phase extraction;
Renin activity;
D O I:
10.1177/0004563214524558
中图分类号:
R446 [实验室诊断];
R-33 [实验医学、医学实验];
学科分类号:
1001 ;
摘要:
Background The measurement of plasma renin activity using LC-MS/MS is an attractive alternative to radioimmunoassay techniques. A published method for renin activity by LC-MS/MS uses equipment that is not available in all LC-MS/MS laboratories. Here, we present an offline modification to allow users of LC-MS/MS to perform this analysis without specialist equipment. Methods Samples were prepared in duplicate and incubated for 6.5 h and 24 h. Solid phase extraction was performed offline using Waters Oasis MAX ion-exchange 96-well plate. A method comparison was performed between this assay and a previously published assay using on-line solid phase extraction. Results The offline method for plasma rennin activity gave similar results to the on-line method across the concentration range. Analytical performance was also found to be comparable and the use of a 24 h incubation for low samples was also deemed unnecessary. Discussion The method described is an acceptable alternative to a previously published assay which does not require the use of highly specialist equipment and will be applicable to the majority of LC-MS/MS users.
引用
收藏
页码:409 / 411
页数:3