摘要
Purpose: This cadaver study evaluates the use of pre-bent 3-dimensional titanium mesh implants for orbital floor and medial wall reconstruction. Design: Nonrandomized comparative study. Participants: Eight human cadaveric heads (n = 16 orbits). Methods: Transcutaneous incisions were used to expose both orbital floors in each cadaveric head. Unilateral orbital floor and medial wall fractures were generated in each specimen. The contralateral orbit remained uninjured. The fractures then were repaired with pre-bent titanium mesh fan plates molded from aluminum templates presenting different sizes. The templates were generated from topographical computed tomography (CT) data previously obtained from normal subjects. The accuracy of orbital reconstruction was evaluated with postoperative CT scans. Main Outcome Measures: The mean value of the distances between the implant and the bony orbit was evaluated. Results: The mean distance between all 16 plates and their respective orbital floors/medial walls was 0.81 +/- 0.74 mm. Mean values were 0.68 +/- 0.63 mm for the unfractured side and 0.93 +/- 0.82 mm for the fractured side. No significant differences were found between orbits when evaluated for side of injury, gender, or size of defect. Conclusion: Pre-bent 3-dimensional titanium mesh implants provide accurate reconstruction of orbital floor and medial orbital wall fractures. The mean implant error was < 1 mm for all orbits studied. (c) 2006 by the American Academy of Ophthalmology.
引用
收藏
页码:1863 / 1868
页数:6
相关论文
共 45 条
[1]   ORBITAL VOLUME MEASUREMENTS IN ENOPHTHALMOS USING 3-DIMENSIONAL CT IMAGING [J].
BITE, U ;
JACKSON, IT ;
FORBES, GS ;
GEHRING, DG .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1985, 75 (04) :502-507
[2]   Reconstruction of orbital floor fracture with polyglactin 910/polydioxanon patch (Ethisorb):: A retrospective study [J].
Büchel, P ;
Rahal, A ;
Seto, C ;
Iizuka, T .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2005, 63 (05) :646-650
[3]   Internal fixation in trapdoor-type orbital blowout fracture [J].
Burm, JS .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2005, 116 (04) :962-970
[4]   Pure orbital blowout fracture: New concepts and importance of medial orbital blowout fracture [J].
Burm, JS ;
Chung, CH ;
Oh, SJ .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1999, 103 (07) :1839-1849
[5]  
Burnstine Michael A, 2003, Curr Opin Ophthalmol, V14, P236, DOI 10.1097/00055735-200310000-00002
[6]  
Chang Eli L, 2004, Curr Opin Ophthalmol, V15, P411, DOI 10.1097/01.icu.0000137854.37950.fb
[7]   The accuracy of stereolithography in planning craniofacial bone replacement [J].
Chang, PSH ;
Parker, TH ;
Patrick, CW ;
Miller, MJ .
JOURNAL OF CRANIOFACIAL SURGERY, 2003, 14 (02) :164-170
[8]   Assessment of internal orbital reconstructions for pure blowout fractures: Cranial bone grafts versus titanium mesh [J].
Ellis, E ;
Tan, YH .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2003, 61 (04) :442-453
[9]   AN ANALYSIS OF 2,067 CASES OF ZYGOMATICO-ORBITAL FRACTURE [J].
ELLIS, E ;
ELATTAR, A ;
MOOS, KF .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1985, 43 (06) :417-428
[10]   Single-step fronto-orbital resection and reconstruction with individual resection template and corresponding titanium implant:: a new method of computer-aided surgery [J].
Eufinger, H ;
Wittkampf, ARM ;
Wehmöller, M ;
Zonneveld, FW .
JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 1998, 26 (06) :373-378