Design and testing of conductive polysilicon beam leads for use in a high-density biomedical connector

被引:2
作者
Nikles, S
Bradley, R
Bledsoe, S
Najafi, K
机构
[1] Univ Michigan, Ctr Wireless Integrated Microsyst, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Biomed Engn, Ann Arbor, MI 48109 USA
[3] Univ Michigan, Sch Dent, Dept Biol & Mat Sci, Ann Arbor, MI 48109 USA
[4] Univ Michigan, Dept Otolaryngol, Ann Arbor, MI 48109 USA
关键词
D O I
10.1088/0960-1317/14/7/017
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A study was conducted to measure and characterize the reliability of polysilicon cantilever beams with electroplated gold pads for use in a high-density biomedical connector. In this design, an array of beams is brought into contact with a corresponding array of 30 mum high gold bumps, forming electrical connection. Analytical computations of multi-layer beams were performed, including the effects of residual stresses. Beam leads of various lengths and widths with electroplated gold contact pads on their ends were tested over 1000 cycles to determine their mechanical reliability, and to measure their contact resistance with gold bumps on a separate substrate. The dimensions of the polysilicon beam that produce the least breakage were determined to be 400 mum long by 125 mum wide. For a beam having a calculated contact force of similar to100 muN, the initial contact resistance was 764 mOmega. After 1000 connect/disconnect cycles, beams of this type had an average final contact resistance of 1.598 Omega. These results demonstrate that very high-density connectors with high mechanical reliability and low-contact resistance can be fabricated.
引用
收藏
页码:957 / 968
页数:12
相关论文
共 34 条
[11]  
Halg B., 1990, Proceedings. IEEE Micro Electro Mechanical Systems. An Investigation of Micro Structures, Sensors, Actuators, Machines and Robots (Cat. No.90CH2832-4), P172, DOI 10.1109/MEMSYS.1990.110271
[12]   SILICON RIBBON CABLES FOR CHRONICALLY IMPLANTABLE MICROELECTRODE ARRAYS [J].
HETKE, JF ;
LUND, JL ;
NAJAFI, K ;
WISE, KD ;
ANDERSON, DJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (04) :314-321
[13]   Contact physics of gold microcontacts for MEMS switches [J].
Hyman, D ;
Mehregany, M .
IEEE TRANSACTIONS ON COMPONENTS AND PACKAGING TECHNOLOGIES, 1999, 22 (03) :357-364
[14]  
JAMIESON B, 2000, ANN BIOMED ENG S1, V28, pS112
[15]   INFLUENCE OF SURFACE-COATINGS ON ELASTICITY, RESIDUAL-STRESSES, AND FRACTURE PROPERTIES OF SILICON MICROELEMENTS [J].
JOHANSSON, S ;
ERICSON, F ;
SCHWEITZ, JA .
JOURNAL OF APPLIED PHYSICS, 1989, 65 (01) :122-128
[16]   Influence of substrates on the elastic reaction of films for the microindentation tests [J].
Kim, MT .
THIN SOLID FILMS, 1996, 283 (1-2) :12-16
[17]  
KOVACS GTA, 1998, MICROMACHINED TRANSD, pR20
[18]   The effects of post-deposition processes on polysilicon Young's modulus [J].
Lee, S ;
Cho, C ;
Kim, J ;
Park, S ;
Yi, S ;
Kim, J ;
Cho, DD .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1998, 8 (04) :330-337
[19]  
MUHLSTEIN CL, 2001, MECH PROPERTIES STRU, pR7
[20]   A HIGH-YIELD IC-COMPATIBLE MULTICHANNEL RECORDING ARRAY [J].
NAJAFI, K ;
WISE, KD ;
MOCHIZUKI, T .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1985, 32 (07) :1206-1211