Temperature dependences of LaBr3(Ce), LaCl3(Ce) and NaI(Tl) scintillators

被引:142
作者
Moszynski, M. [1 ]
Nassalski, A.
Syntfeld-Kazuch, A.
Szczesniak, T.
Czarnacki, W.
Wolski, D.
Pausch, G.
Stein, J.
机构
[1] Soltan Inst Nucl Studies, PL-05400 Otwock, Poland
[2] Target Syst Elect GmbH, D-42651 Solingen, Germany
关键词
NaI(Tl); LaCl3 and LaBr3 scintillators; thermal instability of light output; decay time constants of light pulses; energy resolution; photomultiplier thermal instability;
D O I
10.1016/j.nima.2006.06.039
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The temperature dependence of light output, energy resolution and decay time constants of the light pulses of NaI(TI), LaCl3(Ce) (LaCl3) and LaBr3(Ce) (LaBr3) crystals were measured over the temperature range of -30 to 60 degrees C. In the study of the light output, the number of photoelectrons produced by the scintillators in the XP2020 photomultiplier was measured and corrected for by the temperature dependence of the quantum efficiency determined for 360 and 420nm, respectively. It showed a high stability of the light output of LaBr3 of about 0.01%/degrees C and a comparable uniformity of LaCl3 at a long peaking time of 12 mu s. The well-known thermal instability of NaI(TI) was confirmed at a short peaking time of 2 mu s. However, a much better stability of NaI(TI) at low temperatures was observed for a long peaking time. The study of the decay of light pulses from LaCl3 and LaBr3 crystals confirmed earlier measurements, while NaI(TI) showed a complex behavior at different temperatures. At low temperatures a strong contribution of a slow component of up to 60% of the total light was observed, while at elevated temperatures a well-known initial slow decay was replaced by a delayed maximum and the slow component became insignificant.. The results of the study of energy resolution seem to be correlated with the variation of both the light output and a dependence of the decay time constants of the light pulses at changing temperature. This is particularly interesting in the case of NaI(TI), where different dependencies of the energy resolution as a function of temperature for different peaking times in the spectroscopy amplifier were found. Tests of the XP2020 PMT itself showed that the thermal instability of the gain of the dynode structure of about-0.4%/degrees C is a dominating effect. The opposite effect on an increasing quantum efficiency, partly compensating for the gain instability, was observed above 10 degrees C for the longer wavelength of 420nm. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:739 / 751
页数:13
相关论文
共 27 条
[1]   Comparison of LaCl3:Ce and NaI(Tl) scintillators in γ-ray spectrometry [J].
Balcerzyk, M ;
Moszynski, M ;
Kapusta, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 537 (1-2) :50-56
[2]   Energy resolution and light yield non-proportionality of ZnSe: Te scintillator studied by large area avalanche photodiodes and photomultipliers [J].
Balcerzyk, M ;
Klamra, W ;
Moszynski, M ;
Kapusta, M ;
Szawlowski, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 482 (03) :720-727
[3]   MEASUREMENT OF TIME DEPENDENCE OF SCINTILLATION INTENSITY BY A DELAYED-COINCIDENCE METHOD [J].
BOLLINGER, L ;
THOMAS, GE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1961, 32 (09) :1044-+
[4]   Non-proportionality in the scintillation response and the energy resolution obtainable with scintillation crystals [J].
Dorenbos, P ;
deHaas, JTM ;
vanEijk, CWE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (06) :2190-2202
[5]   Gamma ray spectroscopy with a ⊘19 x19 mm3 LaBr3 : 0.5% Ce3+ scintillator [J].
Dorenbos, P ;
de Haas, JTM ;
van Eijk, CWE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (03) :1289-1296
[6]   Gamma ray spectroscopy with CsI(Tl) scintillator coupled to silicon drift chamber [J].
Fiorini, C ;
Longoni, A ;
Perotti, F ;
Labanti, C ;
Lechner, P ;
Struder, L .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (06) :2553-2560
[7]  
Flyckt S.O., 2002, Photomultiplier tubes principles and applications
[8]   Luminescence in solids. [J].
Gurney, RW ;
Mott, NF .
TRANSACTIONS OF THE FARADAY SOCIETY, 1939, 35 (01) :0069-0073
[9]   TUKAN - An 8K pulse height analyzer and multi-channel scaler with a PCI or a USB interface [J].
Guzik, Z ;
Borsuk, S ;
Traczyk, K ;
Plominski, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) :231-235
[10]  
IANAKIEV K, 2005, IEEE NSSMIC C PUERT