Performance evaluation of the 16-module quad-HIDAC small animal PET camera

被引:63
作者
Missimer, J [1 ]
Madi, Z
Honer, M
Keller, C
Schubiger, A
Ametamey, SM
机构
[1] ETH, PSI, Ctr Radiopharmaceut Sci, CH-5232 Villigen, Switzerland
[2] PSI, Div Life Sci, USZ, CH-5232 Villigen, Switzerland
关键词
D O I
10.1088/0031-9155/49/10/016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The quad-HIDAC small animal PET camera is a quadratic array of high-density avalanche chambers; the camera described in this publication consists of 16 modules. We present the system response using point and line sources and a mouse phantom. The quad-HIDAC camera exhibits a count rate stability of better than 1% and linearity of response to coincidences up to 2.2 x 10(5) cps at 16 MBq activity. Corrected for deadtime and random coincidences, the efficiency for the line source is 0.011, of which unscattered coincidences yield 0.009. The scatter fraction originating from the detectors is 0.22. Absorption within the mouse phantom was 20% and the scatter fraction increased to 0.29. Resolution is uniform within the entire field-of-view, which is 28 cm axially and 17 cm radially. Reconstruction of a point source yields a resolution of 1.1 mm FWHM for all three components. The performance of the camera demonstrates its excellent suitability for the functional imaging of small animals.
引用
收藏
页码:2069 / 2081
页数:13
相关论文
共 17 条
[1]  
BELLONI T, 2002, IT ADV SCH L DAVINCI
[2]  
Guzzardi R, 1991, J Nucl Biol Med, V35, P141
[3]  
Honer M, 2004, J NUCL MED, V45, P464
[4]   HALF-LIVES OF 35 RADIONUCLIDES [J].
HOUTERMANS, H ;
MILOSEVIC, O ;
REICHEL, F .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1980, 31 (03) :153-154
[5]   A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals [J].
Jeavons, AP ;
Chandler, RA ;
Dettmar, CAR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (03) :468-473
[6]  
JEAVONS AP, 2002, 10 BOETT COLL 2 WORK
[7]   ANALYTIC 3D IMAGE-RECONSTRUCTION USING ALL DETECTED EVENTS [J].
KINAHAN, PE ;
ROGERS, JG .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :964-968
[8]   Performance evaluation of the microPET R4 PET scanner for rodents [J].
Knoess, C ;
Siegel, S ;
Smith, A ;
Newport, D ;
Richerzhagen, R ;
Winkeler, A ;
Jacobs, A ;
Goble, RN ;
Graf, R ;
Wienhard, K ;
Heiss, WD .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (05) :737-747
[9]  
MANAVAKI R, 2002, M132 IEEE NSS MIC C
[10]  
MISSIMER J, 2001, C HIGH RES IM SMALL