Evaluation of a Novel Cone Beam Computed Tomography Scanner for Bone Density Examinations in Preoperative 3D Reconstructions and Correlation with Primary Implant Stability

被引:42
|
作者
Sennerby, Lars [1 ,2 ]
Andersson, Peter [2 ]
Pagliani, Luca
Giani, Claudio [3 ]
Moretti, Giacomo [4 ]
Molinari, Massimo [3 ]
Motroni, Alessandro [4 ]
机构
[1] Univ Gothenburg, Dept Oral & Maxillofacial Surg, Inst Odontol, SE-40530 Gothenburg, Sweden
[2] Clin Feltre, Feltre, Italy
[3] de Gotzen Srl, Olgiate Olona, Italy
[4] 3Diemme Srl, Cantu, Italy
关键词
bone density; CBCT; clinical study; implant stability; insertion torque; resonance frequency; RESONANCE FREQUENCY-ANALYSIS; TORQUE; INSERTION; QUALITY;
D O I
10.1111/cid.12193
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
BackgroundBone density examination (BDE) using preoperative cone beam computed tomography (CBCT) might be used to predict primary implant stability in implant patients. PurposeThe aim of the study was to validate a novel CBCT scanner in vitro with regard to BDE in preoperative scans and to analyze the in vivo correlation of CBCT scan results with primary implant stability measurements. Materials and MethodsA CBCT scanner was validated in vitro with regard to spatial uniformity and linearity of CT numbers (Hounsfield units, HU) by using a series of phantoms and plastic and hydroxyapatite specimens of various densities. Forty-nine patients (27 female and 22 male, mean age 55.69.8) were scanned prior to and 1 to 6 months after the placement of 155 dental implants of different lengths and diameters. Mean and peak insertion torque (IT) in Ncm were registered during implant placement. Resonance frequency analysis (RFA) measurements in ISQ units were performed after placement. The second scan was used to export and superimpose the exact positions of bone and implants into the first scan. Virtual probes with the same length as the actual implant were automatically placed at the implant sites, and mean HU values were measured in a 1mm-wide circular corridor from the tip of the threads and out. ResultsThe in vitro validation showed high uniformity and linearity of CT numbers (HU). The clinical study showed significant correlations between bone density and ISQ, mean IT, and peak IT, respectively. ConclusionsThe WhiteFox CBCT scanner measures bone density with high accuracy. There is a correlation between bone density and primary implant stability as assessed with IT and RFA measurements. The findings suggest that BDE may be used as an additional feature in treatment-planning software to estimate primary stability at predetermined implant sites.
引用
收藏
页码:844 / 853
页数:10
相关论文
共 50 条
  • [21] Relationships Between Bone Density Values from Cone Beam Computed Tomography, Maximum Insertion Torque, and Resonance Frequency Analysis at Implant Placement: A Pilot Study
    Angeles Fuster-Torres, Maria
    Penarrocha-Diago, Maria
    Penarrocha-Oltra, David
    Penarrocha-Diago, Miguel
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2011, 26 (05) : 1051 - 1056
  • [22] 3D-evaluation of the maxillary sinus in cone-beam computed tomography
    Luz, Julia
    Greutmann, Dominique
    Wiedemeier, Daniel
    Rostetter, Claudio
    Ruecker, Martin
    Stadlinger, Bernd
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2018, 4
  • [23] 3D-evaluation of the maxillary sinus in cone-beam computed tomography
    Julia Luz
    Dominique Greutmann
    Daniel Wiedemeier
    Claudio Rostetter
    Martin Rücker
    Bernd Stadlinger
    International Journal of Implant Dentistry, 4
  • [24] Correlation of two different devices for the evaluation of primary implant stability depending on dental implant length and bone density: An in vitro study
    Lee, Jungwon
    Lim, Young-Jun
    Ahn, Jin-Soo
    Kim, Bongju
    Baek, Yeon-Wha
    Lim, Bum-Soon
    PLOS ONE, 2024, 19 (05):
  • [25] Evaluation of Post-implant Buccal Bone Resorption Using Cone Beam Computed Tomography: A Clinical Pilot Study
    Vera, Carolina
    De Kok, Ingeborg J.
    Chen, Wenjie
    Reside, Glenn
    Tyndall, Donald
    Cooper, Lyndon F.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2012, 27 (05) : 1249 - 1257
  • [26] Toward improved 3D carotid artery imaging with Adaptive CaRdiac cOne BEAm computed Tomography (ACROBEAT)
    Reynolds, Tess
    Dillon, Owen
    Prinable, Joseph
    Whelan, Brendan
    Keall, Paul J.
    O'Brien, Ricky T.
    MEDICAL PHYSICS, 2020, 47 (11) : 5749 - 5760
  • [27] Influence of object location in cone beam computed tomography (NewTom 5G and 3D Accuitomo 170) on gray value measurements at an implant site
    Parsa, Azin
    Ibrahim, Norliza
    Hassan, Bassam
    van der Stelt, Paul
    Wismeijer, Daniel
    ORAL RADIOLOGY, 2014, 30 (02) : 153 - 159
  • [28] Artifact Reduction in 3D and 4D Cone-Beam Computed Tomography Images With Deep Learning: A Review
    Amirian, Mohammadreza
    Barco, Daniel
    Herzig, Ivo
    Schilling, Frank-Peter
    IEEE ACCESS, 2024, 12 : 10281 - 10295
  • [29] Evaluation of postoperative dental implant primary stability using 3D finite element analysis
    Pammer, David
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2019, 22 (03) : 280 - 287
  • [30] 3D tooth segmentation in cone-beam computed tomography images using distance transform
    Kakehbaraei, Somayeh
    Arvanaghi, Roghayyeh
    Seyedarabi, Hadi
    Esmaeili, Farzad
    Zenouz, Ali Taghavi
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2023, 79