A novel thermal desorption spectroscopy apparatus

被引:8
|
作者
Castro, FJ [1 ]
Meyer, G
机构
[1] Comis Nacl Energia Atom, Ctr Atom Bariloche, RA-8400 San Carlos De Bariloche, Rio Negro, Argentina
[2] Inst Balseiro, UNC, RA-8400 San Carlos De Bariloche, Rio Negro, Argentina
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2000年 / 71卷 / 05期
关键词
D O I
10.1063/1.1150594
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We present a new experimental setup designed to perform measurements of thermal desorption spectroscopy of hydrogen released from metallic samples. The distinctive features of the proposed arrangement are a compact reactor and the use of a mass flow meter to measure the flux of gas desorbed from the sample. A complete set of hydrogen desorption spectra from Pd samples has been recorded to test the equipment. The spectra show very good reproducibility, a high signal-to-noise ratio, and agree very well with results reported in the literature obtained using more sophisticated equipment. These features encourage the use of this low cost setup to perform a fast and reliable characterization of the processes controling hydrogen desorption from metallic compounds. (C) 2000 American Institute of Physics. [S0034-6748(00)04105-8].
引用
收藏
页码:2131 / 2133
页数:3