Determination of silicate in freshwaters using flow injection with luminol chemiluminescence detection

被引:15
作者
Yaqoob, M
Nabi, A
Worsfold, PJ
机构
[1] Univ Plymouth, Sch Earth Ocean & Environm Sci, Plymouth PL4 8AA, Devon, England
[2] Univ Plymouth, Plymouth Environm Res Ctr, Plymouth PL4 8AA, Devon, England
[3] Univ Baluchistan, Dept Chem, Quetta, Pakistan
关键词
flow injection analysis; chemiluminescence; silicate; heteropoly acid; freshwaters;
D O I
10.1016/j.aca.2004.05.076
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection (171) manifold based on luminol chemiluminescence (CL) detection for the determination of silicate in freshwaters is described. The molybdosilicic heteropoly acid formed by silicate and ammonium molybdate in acidic conditions generated CL response via the oxidation of luminol. The calibration graph was linear over the range 0.35-140 mug Si L-1 (r(2) = 0.9974) at 25 degreesC with relative standard deviations (n = 5) in the range 1.0-3.8%. The practical limit of detection was 0.35 mug Si L-1 with a sample throughput of 80 h(-1). Potential interferences in freshwaters were removed by passing the sample stream through in-line iminodiacetate chelating (to remove cations) and anion exchange (to remove phosphate) micro-columns. The method was applied to freshwater samples from the upper Tamar catchment and the results obtained (140 +/- 9 to 1493 +/- 47 mug Si L-1) were in reasonable agreement with the results obtained using a segmented flow analyser with spectrophotometric detection (182 +/- 9 to 1652 +/- 23 mug Si L-1) as the reference method. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:137 / 142
页数:6
相关论文
共 31 条
[1]   UV digestion of seawater samples prior to the determination of copper using flow injection with chemiluminescence detection [J].
Achterberg, EP ;
Braungardt, CB ;
Sandford, RC ;
Worsfold, PJ .
ANALYTICA CHIMICA ACTA, 2001, 440 (01) :27-36
[2]  
*AM PUB ASS AM WAT, 1985, STAND METH EX WAT WA
[3]   Determination of sub-nanomolar levels of iron in seawater using flow injection with chemiluminescence detection [J].
Bowie, AR ;
Achterberg, EP ;
Mantoura, RFC ;
Worsfold, PJ .
ANALYTICA CHIMICA ACTA, 1998, 361 (03) :189-200
[4]  
Bowie AR, 1996, J BIOLUM CHEMILUM, V11, P61, DOI 10.1002/(SICI)1099-1271(199603)11:2<61::AID-BIO406>3.0.CO
[5]  
2-O
[6]   Real-time monitoring of picomolar concentrations of iron(II) in marine waters using automated flow injection-chemiluminescence instrumentation [J].
Bowie, AR ;
Achterberg, EP ;
Sedwick, PN ;
Ussher, S ;
Worsfold, PJ .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2002, 36 (21) :4600-4607
[7]   AUTOMATIC DETERMINATION OF SILICATE-SILICON IN NATURAL WATERS WITH SPECIAL REFERENCE TO SEA WATER [J].
BREWER, PG ;
RILEY, JP .
ANALYTICA CHIMICA ACTA, 1966, 35 (04) :514-&
[8]   Determination of cobalt and iron in estuarine and coastal waters using flow injection with chemiluminescence detection [J].
Cannizzaro, V ;
Bowie, AR ;
Sax, A ;
Achterberg, EP ;
Worsfold, PJ .
ANALYST, 2000, 125 (01) :51-57
[9]  
Fujiwara T, 1996, APPL ORGANOMET CHEM, V10, P675, DOI 10.1002/(SICI)1099-0739(199611)10:9<675::AID-AOC531>3.0.CO
[10]  
2-#