Memory effect in deuterium analysis by continuous flow isotope ratio measurement

被引:29
作者
Olsen, Jesper
Seierstad, Inger
Vinther, Bo
Johnsen, Sigfus
Heinemeier, Jan
机构
[1] Aarhus Univ, Dept Phys & Astron, DK-8000 Aarhus C, Denmark
[2] Niels Bohr Inst Astron Phys & Geophys, DK-2100 Copenhagen, Denmark
关键词
delta D deuterium measurements; CF-IRMS; memory effect; correction;
D O I
10.1016/j.ijms.2006.05.004
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
The method of deuterium analysis (delta D) measurements has previously been shown to suffer from intersample memory effects. In this study the memory effect of delta D continuous flow isotope ratio mass spectrometry (CF-IRMS) employing the chromium reduction technique is investigated with the goal of optimising the system configuration. By optimising the system configuration, the magnitude of the memory effect has been reduced from 6% to 1-2%. It is shown that the memory effect displays a long range behaviour which can be described by a sum of two exponential decay functions, indicating two main sources with different time constants. Furthermore, a correction algorithm which takes into account the observed long range nature of the memory effect is developed and tested against waters of known isotopic composition. (C) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:44 / 52
页数:9