Exploratory survey of image quality on CR digital mammography imaging systems in Mexico

被引:7
|
作者
Gaona, E. [1 ]
Rivera, T. [2 ]
Arreola, M. [3 ]
Franco, J. [1 ]
Molina, N. [1 ]
Alvarez, B. [4 ]
Azorin, C. G. [2 ]
Casian, G. [4 ]
机构
[1] Univ Autonoma Metropolitana Xochimilco, Mexico City 04960, DF, Mexico
[2] IPN, Unidad Legaria, Ctr Invest Ciencia Aplicada & Tecnol Avanzada, Mexico City 11500, DF, Mexico
[3] Univ Florida, Dept Radiol, Gainesville, FL 32610 USA
[4] Hosp Juarez Mexico, Mexico City 07760, DF, Mexico
关键词
CR mammography; Quality image; Dose;
D O I
10.1016/j.apradiso.2013.07.009
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The purpose of this study was to assess the current status of image quality and dose in computed radiographic digital mammography (CRDM) systems. Studies included CRDM systems of various models and manufacturers which dose and image quality comparisons were performed. Due to the recent rise in the use of digital radiographic systems in Mexico, CRDM systems are rapidly replacing conventional film-screen systems without any regard to quality control or image quality standards. Study was conducted in 65 mammography facilities which use CRDM systems in the Mexico City and surrounding States. The systems were tested as used clinically. This means that the dose and beam qualities were selected using the automatic beam selection and photo-timed features. All systems surveyed generate laser film hardcopies for the radiologist to read on a scope or mammographic high luminance light box. It was found that 51 of CRDM systems presented a variety of image artefacts and non-uniformities arising from inadequate acquisition and processing, as well as from the laser printer itself. Undisciplined alteration of image processing settings by the technologist was found to be a serious prevalent problem in 42 facilities. Only four of them showed an image QC program which is periodically monitored by a medical physicist. The Average Glandular Dose (AGD) in the surveyed systems was estimated to have a mean value of 2.4 mGy. To improve image quality in mammography and make more efficient screening mammographic in early detection of breast cancer is required new legislation. (C) 2013 Published by Elsevier Ltd.
引用
收藏
页码:245 / 248
页数:4
相关论文
共 50 条
  • [21] The influence of software filtering in digital mammography image quality
    Michail, C.
    Spyropoulou, V.
    Kalyvas, N.
    Valais, I.
    Dimitropoulos, N.
    Fountos, G.
    Kandarakis, I.
    Panayiotakis, G.
    JOURNAL OF INSTRUMENTATION, 2009, 4
  • [22] Study on patient dosimetry and image quality in digital mammography
    Xavier A.C.S.
    Andrade M.E.A.
    Pinto B.V.-C.C.
    de Barros V.S.M.
    Kramer R.
    Khoury H.J.
    Andrade, Marcos Ely Almeida (marcos.ely@gmail.com), 1600, Brazilian Society of Biomedical Engineering (33): : 138 - 143
  • [23] Evaluation and Comparison of Image Quality Figures for Digital Mammography
    Tsai, H. Y.
    Liu, H. L.
    Hwang, Y. S.
    Chen, C. C.
    MEDICAL PHYSICS, 2009, 36 (06) : 2451 - +
  • [24] Figure of Image Quality and Information Capacity in Digital Mammography
    Michail, Christos M.
    Kalyvas, Nektarios E.
    Valais, Ioannis G.
    Fudos, Ioannis P.
    Fountos, George P.
    Dimitropoulos, Nikos
    Koulouras, Grigorios
    Kandris, Dionisis
    Samarakou, Maria
    Kandarakis, Ioannis S.
    BIOMED RESEARCH INTERNATIONAL, 2014, 2014
  • [25] Effect of image quality on calcification detection in digital mammography
    Warren, Lucy M.
    Mackenzie, Alistair
    Cooke, Julie
    Given-Wilson, Rosalind M.
    Wallis, Matthew G.
    Chakraborty, Dev P.
    Dance, David R.
    Bosmans, Hilde
    Young, Kenneth C.
    MEDICAL PHYSICS, 2012, 39 (06) : 3202 - 3213
  • [26] Initial image quality and clinical experience with new CR digital mammography system:: A phantom and clinical study
    Gaona, Enrique
    Alvarez Alfonso, Beatriz Y.
    Casian Castellanos, Gustavo
    Franco Enriquez, Jesus Gabriel
    MEDICAL PHYSICS, 2008, 1032 : 237 - 239
  • [27] Practical Evaluation of Image Quality in Computed Radiographic (CR) Imaging Systems
    Desai, Nikunj
    Singh, Abhinav
    Valentino, Daniel J.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [28] Optimization of image quality and average glandular dose in CR mammography
    Satoh, Kimihiko
    Kuwabara, Takao
    Yasuda, Hiroaki
    Arakawa, Satoshi
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [29] CR Mammography: Image Quality Measurement and Model Calculation for Needle vs. Powder Imaging Plate
    Vandenbroucke, Dirk A. N.
    Leblans, Paul J. R.
    DIGITAL MAMMOGRAPHY, 2010, 6136 : 219 - 226
  • [30] A statistical alternative to current measures of image quality in digital mammography
    Caldwell, D.
    Baldelli, P.
    Phelan, N.
    Kenny, P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (03):