Exploratory survey of image quality on CR digital mammography imaging systems in Mexico

被引:7
|
作者
Gaona, E. [1 ]
Rivera, T. [2 ]
Arreola, M. [3 ]
Franco, J. [1 ]
Molina, N. [1 ]
Alvarez, B. [4 ]
Azorin, C. G. [2 ]
Casian, G. [4 ]
机构
[1] Univ Autonoma Metropolitana Xochimilco, Mexico City 04960, DF, Mexico
[2] IPN, Unidad Legaria, Ctr Invest Ciencia Aplicada & Tecnol Avanzada, Mexico City 11500, DF, Mexico
[3] Univ Florida, Dept Radiol, Gainesville, FL 32610 USA
[4] Hosp Juarez Mexico, Mexico City 07760, DF, Mexico
关键词
CR mammography; Quality image; Dose;
D O I
10.1016/j.apradiso.2013.07.009
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The purpose of this study was to assess the current status of image quality and dose in computed radiographic digital mammography (CRDM) systems. Studies included CRDM systems of various models and manufacturers which dose and image quality comparisons were performed. Due to the recent rise in the use of digital radiographic systems in Mexico, CRDM systems are rapidly replacing conventional film-screen systems without any regard to quality control or image quality standards. Study was conducted in 65 mammography facilities which use CRDM systems in the Mexico City and surrounding States. The systems were tested as used clinically. This means that the dose and beam qualities were selected using the automatic beam selection and photo-timed features. All systems surveyed generate laser film hardcopies for the radiologist to read on a scope or mammographic high luminance light box. It was found that 51 of CRDM systems presented a variety of image artefacts and non-uniformities arising from inadequate acquisition and processing, as well as from the laser printer itself. Undisciplined alteration of image processing settings by the technologist was found to be a serious prevalent problem in 42 facilities. Only four of them showed an image QC program which is periodically monitored by a medical physicist. The Average Glandular Dose (AGD) in the surveyed systems was estimated to have a mean value of 2.4 mGy. To improve image quality in mammography and make more efficient screening mammographic in early detection of breast cancer is required new legislation. (C) 2013 Published by Elsevier Ltd.
引用
收藏
页码:245 / 248
页数:4
相关论文
共 50 条
  • [1] Exploratory survey of initial image quality in new digital mammography units prior to use in patients in Mexico
    Gaona, E.
    Rivera, T.
    Molina-Frechero, N.
    Franco, J. G.
    APPLIED RADIATION AND ISOTOPES, 2018, 141 : 266 - 269
  • [2] LATIN AMERICAN IMAGE QUALITY SURVEY IN DIGITAL MAMMOGRAPHY STUDIES
    Mora, Patricia
    Khoury, Helen
    Bitelli, Regina
    Rosa Quintero, Ana
    Garay, Fernando
    Garcia Aguilar, Juan
    Gamarra, Mirtha
    Ubeda, Carlos
    RADIATION PROTECTION DOSIMETRY, 2017, 174 (01) : 94 - 101
  • [3] TL dosimetry for quality control of CR mammography imaging systems
    Gaona, E.
    Nieto, J. A.
    Gongora, J. A. I. D.
    Arreola, M.
    Enriquez, J. G. F.
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2007, 162 (10-12): : 759 - 763
  • [4] Image quality of digital mammography images produced using wet and dry laser imaging systems
    Al Khalifah, K.
    Brindhaban, A.
    AlArfaj, R.
    Jassim, O.
    RADIOGRAPHY, 2006, 12 (01) : 13 - 19
  • [5] SURVEY OF MAMMOGRAPHY IMAGE QUALITY IN THE UK
    YOUNG, KC
    RAMSDALE, ML
    RUST, A
    RADIATION PROTECTION DOSIMETRY, 1995, 57 (1-4) : 163 - 165
  • [6] Potential image quality in scintillator CCD-based x-ray imaging systems for digital radiography and digital mammography
    Hejazi, S
    Trauernicht, DP
    PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 : 440 - 449
  • [7] Experimental investigation of the dose and image quality characteristics of a digital mammography imaging system
    Huda, W
    Sajewicz, AM
    Ogden, KM
    MEDICAL PHYSICS, 2003, 30 (03) : 442 - 448
  • [8] Improved image quality in digital mammography with image processing
    Baydush, AH
    Floyd, CE
    MEDICAL PHYSICS, 2000, 27 (07) : 1503 - 1508
  • [9] Designing digital imaging systems for mammography
    Roehrig, H
    RADIOLOGY, 1996, 201 : 9404 - 9404
  • [10] Optimization of Image Quality and Dose in Digital Mammography
    Fausto, Agnes M. F.
    Lopes, M. C.
    de Sousa, M. C.
    Furquim, Tania A. C.
    Mol, Anderson W.
    Velasco, Fermin G.
    JOURNAL OF DIGITAL IMAGING, 2017, 30 (02) : 185 - 196