Further study on a flow injection on-line multiplexed sorption preconcentration coupled with flame atomic absorption spectrometry for trace element determination

被引:26
作者
Li, Y
Jiang, Y
Yan, XP [1 ]
机构
[1] Nankai Univ, Coll Chem, State Key Lab Funct Polymer Mat Adsorpt & Seperat, Tianjin 300071, Peoples R China
[2] Nankai Univ, Coll Chem, Res Ctr Analyt Sci, Tianjin 300071, Peoples R China
基金
中国国家自然科学基金;
关键词
multiplexed sorption preconcentration; knotted reactor; chromium; cobalt; nickel; copper; flame atomic absorption spectrometry;
D O I
10.1016/j.talanta.2004.03.045
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A further study on a newly developed flow injection (FI) on-line multiplexed sorption preconcentration (MSP) using a knotted reactor coupled with flame atomic absorption spectrometry (FAAS) was carried out to demonstrate its applicability and limitation for trace element determination. For this purpose, Cr(VI), Cu(II), Ni(II) and Co(II) were selected as the analytes, and detailed comparison was made between the MSP-FAAS and conventional FI on-line sorption preconcentration FAAS in respect to retention efficiency and linear ranges of absorbance versus sample loading flow rate and total preconcentration time. Introduction of an air-flow for removal of the residual solution in the KR after each sub-injection in the MSP procedure played a decisive role in the improvement of retention efficiency. The linearity of absorbance versus sample loading flow rate or total preconcentration time was extended to a more degree for the metal ions with less stability of their PDC (pyrrolidine dithiocarbamate) complexes than those with more stable PDC complexes. It seems that the MSP procedure behaves advantages beyond the inflection points in the diagrams of absorbance versus total preconcentration time and sample loading flow rate obtained by conventional (a single continuous) preconcentration procedure. With a sample loading flow rate of 6.0 ml min(-1) and a total preconcentration time of 180 s, the retention efficiencies were increased from 25, 46, 41 and 63% with a single continuous sorption preconcentration to 44, 78, 65 and 75% with a six sub-injection preconcentration procedure for Cr(VI), Co(II), Ni(II), and Cu(II), respectively. The detection limits were 0.40, 0.33, 0.31 and 0.26 Rg l(-1) for Cr(VI), Co(II), Ni(II), and Cu(II), respectively. The precision (R.S.D.) for eleven replicate determination of 2 mug l(-1) Cr(VI), Co(II) and Ni(II), and 1 mug l(-1) Cu(II), was 2.1, 4.1, 2.6 and 1.7%, respectively. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:758 / 765
页数:8
相关论文
共 11 条
[1]   DETERMINATION OF COPPER IN WATER AND RICE SAMPLES BY FLAME ATOMIC-ABSORPTION SPECTROMETRY WITH FLOW-INJECTION ONLINE ADSORPTION PRECONCENTRATION USING A KNOTTED REACTOR [J].
CHEN, HW ;
XU, SK ;
FANG, ZL .
ANALYTICA CHIMICA ACTA, 1994, 298 (02) :167-173
[2]   DETERMINATION OF CADMIUM IN BIOLOGICAL-MATERIALS BY FLAME ATOMIC-ABSORPTION SPECTROMETRY WITH FLOW-INJECTION ONLINE SORPTION PRECONCENTRATION [J].
FANG, ZL ;
XU, SK ;
DONG, LP ;
LI, WQ .
TALANTA, 1994, 41 (12) :2165-2172
[3]   A flow injection on-line multiplexed sorption preconcentration procedure coupled with flame atomic absorption spectrometry for determination of trace lead in water, tea, and herb medicines [J].
Li, Y ;
Jiang, Y ;
Yan, XP ;
Peng, WJ ;
Wu, YY .
ANALYTICAL CHEMISTRY, 2002, 74 (05) :1075-+
[4]   Selective flow-injection quantification of ultra-trace amounts of Cr(VI) via on-line complexation and preconcentration with APDC followed by determination by electrothermal atomic absorption spectrometry [J].
Nielsen, S ;
Hansen, EH .
ANALYTICA CHIMICA ACTA, 1998, 366 (1-3) :163-176
[5]   Determination of lead in tap water by ICP-AES with flow-injection on-line adsorption preconcentration using a knotted reactor and ultrasonic nebulization [J].
Salonia, JA ;
Wuilloud, RG ;
Gásquez, JA ;
Olsina, RA ;
Martinez, LD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (08) :1239-1243
[6]   Flow injection on-line preconcentration of low levels of Cr(VI) with detection by ETAAS - Comparison of using an open tubular PTFE knotted reactor and a column reactor packed with PTFE beads [J].
Som-Aum, W ;
Liawruangrath, S ;
Hansen, EH .
ANALYTICA CHIMICA ACTA, 2002, 463 (01) :99-109
[7]   Investigation of on-line coupling electrothermal atomic absorption spectrometry with flow injection sorption preconcentration using a knotted reactor for totally automatic determination of lead in water samples [J].
Sperling, M ;
Yan, XP ;
Welz, B .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (14) :1891-1908
[8]   Flow injection on-line preconcentration and separation coupled with atomic (mass) spectrometry for trace element (speciation) analysis based on sorption of organo-metallic complexes in a knotted reactor [J].
Yan, XP ;
Jiang, Y .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2001, 20 (10) :552-562
[9]   Determination of (ultra)trace amounts of antimony(III) in water by flow injection on-line sorption preconcentration in a knotted reactor coupled with electrothermal atomic absorption spectrometry [J].
Yan, XP ;
VanMol, W ;
Adams, F .
ANALYST, 1996, 121 (08) :1061-1067
[10]   Determination of (ultra)trace amounts of arsenic(III) and arsenic(V) in water by inductively coupled plasma mass spectrometry coupled with flow injection on-line sorption preconcentration and separation in a knotted reactor [J].
Yan, XP ;
Kerrich, R ;
Hendry, MJ .
ANALYTICAL CHEMISTRY, 1998, 70 (22) :4736-4742