The use of suture anchors in repair of the ruptured patellar tendon - A biomechanical study

被引:58
作者
Bushnell, Brandon D. [1 ]
Byram, Ian R. [1 ]
Weinhold, Paul S. [1 ]
Creighton, R. Alex [1 ]
机构
[1] Univ N Carolina, Sch Med, Dept Orthopaed Surg, Chapel Hill, NC 27599 USA
关键词
patellar tendon repair; suture anchors; transpatellar tunnels; extensor mechanism;
D O I
10.1177/0363546506287489
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Rupture of the patellar tendon is a disabling injury that usually requires surgical treatment. The standard method of repair involves placing suture loops through transpatellar tunnels. The use of suture anchors in patellar tendon repair has not been previously described. Hypothesis: No difference exists in the amount of gap formation during cyclic loading or in ultimate load-to-failure strength between repairs performed with anchors and those performed with 2 types of transpatellar sutures. Study Design: Controlled laboratory study. Methods: Six matched pairs of cadaveric knees were tested in a custom biomechanical apparatus based on an established model. Repairs were performed using either suture anchors with No. 2 FiberWire or transpatellar suture tunnels using 2 different types of suture-No. 5 Ethibond and No. 2 FiberWire. Gap formation across the repair site during 250 cycles of extension as well as ramp-up load to failure were measured for each repair. Results: The mean total gap formation across the repair site at 250 cycles was 4.1 +/- 1.9 mm for the suture anchor group, 6.7 +/- 1.8 mm for the FiberWire tunnel group, and 8.5 +/- 2.7 mm for the Ethibond tunnel group. Mean load to failure was 779 +/- 183 N, 730 +/- 83 N, and 763 +/- 231 N, respectively. Conclusion: Significantly less gap formation throughout 250 cycles (P = .009) and no difference in load to failure occurred with patellar tendon repairs performed with suture anchors as compared with repairs performed with transpatellar tunnels. Clinical Relevance: The newly described method, using suture anchors for repair of patellar tendon ruptures, may be clinically equal or superior to the established method of using transpatellar tunnels.
引用
收藏
页码:1492 / 1499
页数:8
相关论文
共 45 条
  • [1] Archdeacon MT, 2003, SKELETAL TRAUMA BASI, P2013
  • [2] AZAR FM, 2003, CAMPBELLS OPERATIVE, P2449
  • [3] Internal fixation strength of suture anchors - Update 1997
    Barber, FA
    Herbert, MA
    Click, JN
    [J]. ARTHROSCOPY, 1997, 13 (03): : 355 - 362
  • [4] Sutures and suture anchors: Update 2003
    Barber, FA
    Herbert, MA
    Richards, DP
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2003, 19 (09) : 985 - 990
  • [5] Suture anchor strength revisited
    Barber, FA
    Herbert, MA
    Click, JN
    [J]. ARTHROSCOPY, 1996, 12 (01): : 32 - 38
  • [6] BARBER FA, 2004, TXB ARTHROSCOPY, P15
  • [7] Traumatic patella tendon rupture: early mobilisation following surgical repair
    Bhargava, SP
    Hynes, MC
    Dowell, JK
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2004, 35 (01): : 76 - 79
  • [8] Cyclic loading of rotator cuff repairs: An in vitro biomechanical comparison of bioabsorbable tacks with transosseous sutures
    Bicknell, RT
    Harwood, C
    Ferreira, L
    King, GJW
    Johnson, JA
    Faber, K
    Drosdowech, D
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2005, 21 (07) : 875 - 880
  • [9] Cyclic loading of anchor-based rotator cuff repairs: Confirmation of the tension overload phenomenon and comparison of suture anchor fixation with transosseous fixation
    Burkhart, SS
    Pagan, JLD
    Wirth, MA
    Athanasiou, KA
    [J]. ARTHROSCOPY, 1997, 13 (06): : 720 - 724
  • [10] In vitro analysis of rotator cuff repairs: A comparison of arthroscopically inserted tacks or anchors with open transosseous repairs
    Chhabra, A
    Goradia, VK
    Francke, EI
    Baer, GS
    Monahan, T
    Kline, AJ
    Miller, MD
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2005, 21 (03) : 323 - 327