Ultrasound-based vessel wall tracking:: An auto-correlation technique with RF center frequency estimation

被引:78
作者
Rabben, SI
Bjærum, S
Sorhus, V
Torp, H
机构
[1] Univ Oslo, Inst Surg Res, Oslo, Norway
[2] GE Vingmed Ultrasound, Horten, Norway
[3] Natl Hosp, SimSurg & Intervent Ctr, Oslo, Norway
[4] Norwegian Univ Sci & Technol, Dept Physiol & Biomed Engn, N-7034 Trondheim, Norway
关键词
ultrasound; Doppler; vessel wall motion; echo tracking; autocorrelation; cross-correlation;
D O I
10.1016/S0301-5629(02)00487-8
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Vessel diameter is related to the distending blood pressure, and is used in estimations of vessel stiffness parameters. The vessel walls can be tracked by integrating wall velocities estimated by ultrasound (US) Doppler techniques. The purpose of this work was to evaluate the performance of the modified autocorrelation estimator when applied on vessel wall motion. As opposed to the conventional autocorrelation method that only estimates the mean Doppler frequency, the modified autocorrelation method estimates both the mean Doppler frequency and the radiofrequency (RF) center frequency. To make a systematic evaluation of the estimator, we performed computer simulations of vessel wall motion, where pulse bandwidth, signal-to-noise ratio (SNR), signal-to-reverberation ratio, packet size and sample volume were varied. As reference, we also analyzed the conventional autocorrelation method and the cross-correlation method with parabolic interpolation. Under the simulation conditions considered here, the modified autocorrelation method had the lowest bias and variance of the estimators. When integrating velocity estimates over several cardiac cycles, the resulting tissue displacement curves might drift. This drift is directly related to the magnitude of the estimator bias and variance. Hence, the modified autocorrelation method should be the preferred choice of method. (E-mail: s.i.rabben@klinmed.uio.no) (C) 2002 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:507 / 517
页数:11
相关论文
共 27 条
[1]   NONINVASIVE ESTIMATES OF AORTIC ROOT PRESSURES - EXTERNAL SUBCLAVIAN ARTERIAL PULSE TRACING CALIBRATED BY OSCILLOMETRICALLY DETERMINED BRACHIAL ARTERIAL PRESSURES [J].
AAKHUS, S ;
TORP, H ;
HAUGLAND, T ;
HATLE, L .
CLINICAL PHYSIOLOGY, 1993, 13 (06) :573-586
[2]   DIAMETER OF INTACT CAROTID ARTERY IN MAN AND ITS CHANGE WITH PULSE PRESSURE [J].
ARNDT, JO ;
KLAUSKE, J ;
MERSCH, F .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1968, 301 (03) :230-&
[3]   CALCULATION OF PULSE-WAVE VELOCITY USING CROSS-CORRELATION - EFFECTS OF REFLEXES IN THE ARTERIAL TREE [J].
BENTHIN, M ;
DAHL, P ;
RUZICKA, R ;
LINDSTROM, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (05) :461-469
[4]   TIME DOMAIN FORMULATION OF PULSE-DOPPLER ULTRASOUND AND BLOOD VELOCITY ESTIMATION BY CROSS-CORRELATION [J].
BONNEFOUS, O ;
PESQUE, P .
ULTRASONIC IMAGING, 1986, 8 (02) :73-85
[5]   A radio frequency domain complex cross-correlation model to estimate blood flow velocity and tissue motion by means of ultrasound [J].
Brands, PJ ;
Hoeks, APG ;
Ledoux, LAF ;
Reneman, RS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (06) :911-920
[6]   A noninvasive method to estimate pulse wave velocity in arteries locally by means of ultrasound [J].
Brands, PJ ;
Willigers, JM ;
Ledoux, LAF ;
Reneman, RS ;
Hoeks, APG .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (09) :1325-1335
[7]   USE OF THE CALIBRATED CAROTID PULSE TRACING FOR CALCULATION OF LEFT-VENTRICULAR PRESSURE AND WALL STRESS THROUGHOUT EJECTION [J].
COLAN, SD ;
BOROW, KM ;
NEUMANN, A .
AMERICAN HEART JOURNAL, 1985, 109 (06) :1306-1310
[8]   DETERMINATION OF TISSUE MOTION VELOCITY BY CORRELATION INTERPOLATION OF PULSED ULTRASONIC ECHO SIGNALS [J].
DEJONG, PGM ;
ARTS, T ;
HOEKS, APG ;
RENEMAN, RS .
ULTRASONIC IMAGING, 1990, 12 (02) :84-98
[9]  
DUCK FA, 1990, PHYSICAL PROPERTIES
[10]   PULSE-WAVE RECORDING - DEVELOPMENT OF A METHOD FOR INVESTIGATING FETAL CIRCULATION INUTERO [J].
ERIKSEN, PS ;
GENNSER, G ;
LINDSTROM, K ;
BENTHIN, M ;
DAHL, P .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1985, 9 (01) :18-27