Uncertainty of high resolution inductively coupled plasma mass spectrometry based aerosol measurements

被引:18
|
作者
Yenisoy-Karakas, Serpil [1 ]
Gaga, Eftade O. [2 ]
Cankur, Oktay [3 ]
Karakas, Duran [4 ]
机构
[1] Abant Izzet Baysal Univ, Fac Art & Sci, Dept Chem, TR-14280 Bolu, Turkey
[2] Anadolu Univ, Fac Engn & Architecture, Dept Environm Engn, Eskisehir, Turkey
[3] TUBITAK UME, Natl Metrol Inst, TR-41470 Gebze, Turkey
[4] Abant Izzet Baysal Univ, Fac Architecture & Engn, Dept Environm Engn, TR-14280 Bolu, Turkey
关键词
Elements; Aerosols; High resolution inductively coupled plasma mass spectrometry; Method validation; Microwave digestion; Uncertainty; SAMPLES;
D O I
10.1016/j.talanta.2009.05.044
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Simultaneous and multi-elemental method was optimized to determine 35 elements in 274 coarse and fine aerosol samples by using HR-ICP-MS. The procedures were validated. Measurement uncertainties of all elements including sampling uncertainties were calculated by applying bottom-up approach. The average recoveries obtained for each element ranged between 79% and 129% using NIST SRM 1648, urban dust. The calculated uncertainties of the analytical methods were between 2.9% and 18% for both sample types. The major contributions to the uncertainty budget come from the calibration curves, repeatability and volume of air. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:1298 / 1305
页数:8
相关论文
共 50 条