Theoretical and experimental determination of dosimetric characteristics for ADVANTAGE™ Pd-103 brachytherapy source

被引:18
作者
Meigooni, Ali S. [1 ]
Dini, Sharifeh A. [1 ]
Awan, Shahid B. [1 ]
Dou, Kai [1 ]
Koona, Rafiq A. [1 ]
机构
[1] Univ Kentucky, Med Ctr, Dept Radiat Med, Lexington, KY 40536 USA
关键词
dosimetry; brachytherapy; Pd-103; thermoluminescent dosimetry; TG-43U1;
D O I
10.1016/j.apradiso.2006.03.015
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
ADVANTAGE (TM) Pd-103 brachytherapy source has been recently introduced by IsoAid (TM) for prostate permanent implants. Dosimetric characteristics (Dose rate constant, radial dose function, 2D-, and 1D-anisotropy functions) of this source model have been determined using both theoretical and experimental methods, following the updated TG-43U1 protocol. Derivation of the dose rate constant was based on recent NIST WAFAC calibration performed in accordance with the 1999 Standards. Measurements were performed in Solid Water (TM) using LiF TLD chips and the theoretical calculations were performed in Solid Water (TM) and liquid water phantom materials using PTRAN Monte Carlo code. The results of the Monte Carlo simulation have shown a dose rate constant of 0.69 cGy h(-1) U-1 in liquid water and 0.67 cGy h(-1) U-1 in Solid Water (TM) medium. The measured dose rate constant in Solid Water TM was found to be 0.68 +/- 8% cGyh(-1) U-1, which is in a good agreement (within 5%) to the Monte Carlo simulated data. The 2D- and 1D-anisotropy functions of the ADVANTAGE (TM) Pd-103 source were calculated for radial distances ranging from 0.5 to 5.0 cm. Radial dose function was determined for radial distances ranging from 0.2 to 8.0 cm using line source approximation. All these calculations are based on L-eff equal to 3.61 cm, calculated following TG-43U1 recommendations. The tabulated data for 2D-anisotropy function, 1D-anisotropy function, dose rate constant and radial dose function have been produced for clinical application of this source model. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:881 / 887
页数:7
相关论文
共 19 条
  • [1] Dosimetric characteristics of a new 125I brachytherapy source
    Gearheart, DM
    Drogin, A
    Sowards, K
    Meigooni, AS
    Ibbott, GS
    [J]. MEDICAL PHYSICS, 2000, 27 (10) : 2278 - 2285
  • [2] Prostate brachytherapy seed identification on post-implant TRUS images
    Han, BH
    Wallner, K
    Merrick, G
    Butler, W
    Sutlief, S
    Sylvester, J
    [J]. MEDICAL PHYSICS, 2003, 30 (05) : 898 - 900
  • [3] Hubbell J. H., 2004, TABLES XRAY MASS ATT
  • [4] Experimental determination of dosimetric characteristics of Best® 125I brachytherapy source
    Meigooni, AS
    Gearheart, DM
    Sowards, K
    [J]. MEDICAL PHYSICS, 2000, 27 (09) : 2168 - 2173
  • [5] INSTRUMENTATION AND DOSIMETER-SIZE ARTIFACTS IN QUANTITATIVE THERMOLUMINESCENCE DOSIMETRY OF LOW-DOSE FIELDS
    MEIGOONI, AS
    MISHRA, V
    PANTH, H
    WILLIAMSON, J
    [J]. MEDICAL PHYSICS, 1995, 22 (05) : 555 - 561
  • [6] Dosimetric characteristics of the new RadioCoil™ 103Pd wire line source for use in permanent brachytherapy implants
    Meigooni, AS
    Zhang, H
    Clark, JR
    Rachabatthula, V
    Koona, RA
    [J]. MEDICAL PHYSICS, 2004, 31 (11) : 3095 - 3105
  • [7] Theoretical and experimental determination of dosimetric characteristics for brachyseed™ Pd-103, model Pd-1, source
    Meigooni, AS
    Zhang, HL
    Perry, C
    Dini, SA
    Koona, RA
    [J]. APPLIED RADIATION AND ISOTOPES, 2003, 58 (05) : 533 - 541
  • [8] Experimental determination of the TG-43 dosimetric characteristics of EchoSeed™ model 6733 125I brachytherapy source
    Meigooni, AS
    Dini, SA
    Sowards, K
    Hayes, JL
    Al-Otoom, A
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 939 - 942
  • [9] Experimental and theoretical determination of dosimetric characteristics of IsoAid ADVANTAGE™ 125I brachytherapy source
    Meigooni, AS
    Hayes, JL
    Zhang, HL
    Sowards, K
    [J]. MEDICAL PHYSICS, 2002, 29 (09) : 2152 - 2158
  • [10] Dosimetric characteristics of the Best® double-wall 103Pd brachytherapy source
    Meigooni, AS
    Bharucha, Z
    Yoe-Sein, M
    Sowards, K
    [J]. MEDICAL PHYSICS, 2001, 28 (12) : 2568 - 2575