On-line solid-phase extraction thermal desorption gas chromatography with ion trap detection tandem mass spectrometry for the analysis of microcontaminants in water

被引:0
|
作者
Louter, AJH [1 ]
vanDoornmalen, J [1 ]
Vreuls, JJ [1 ]
Brinkman, UAT [1 ]
机构
[1] FREE UNIV AMSTERDAM, DEPT ANALYT CHEM, NL-1081 HV AMSTERDAM, NETHERLANDS
来源
HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY | 1996年 / 19卷 / 12期
关键词
solid-phase extraction thermal desorption; gas chromatography; ion trap detection; tandem mass spectrometry; water samples; microcontaminants;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An improved set-up for solid-phase extraction with thermal desorption coupled on-line to gas chromatography (SPETD-GC) is presented, It includes a newly designed liner for a programmable temperature vaporizer BTV) and an improved water elimination system, The SPETD procedure now includes a washing step with HPLC-grade water to prevent degradation of analytes due to interaction with remaining sample constituents, The system was used to analyze surface and tap water samples over a ii-month period, No decrease of chromatographic or trace-enrichment performance was observed, and a liner packed with Tenax GR could be used for at least 150 analyses, The SPETD module was coupled to GC,vith ion-trap detection for mass spectrometric (MS) and MS/MS detection, The linearity and repeatability of the procedure for several pesticides which were tested in the 0.5-10 mu g/1 range were fully satisfactory (1 mu g/l, RSD range 5-11%; n = 5), When using sample volumes of 0.1 ml only, detection limits were as low as 0.1-0.2 mu g/l. As an example, the confirmation and quantification of a suspected pesticide in a real-life sample using electron impact and positive chemical ionization in both the MS and MS/MS mode is shown.
引用
收藏
页码:679 / 685
页数:7
相关论文
共 50 条
  • [1] Determination of microcontaminants in sediments by on-line solid-phase extraction-gas chromatography-mass spectrometry
    Slobodník, J
    Ramalho, S
    van Baar, BM
    Louter, AJH
    Brinkman, UAT
    CHEMOSPHERE, 2000, 41 (09) : 1469 - 1478
  • [2] Use of a drying cartridge in on-line solid-phase extraction gas chromatography mass spectrometry
    Hankemeier, T
    Louter, AJH
    Dalluge, J
    Vreuls, RJJ
    Brinkman, UAT
    HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY, 1998, 21 (08): : 450 - 456
  • [3] On-line solid-phase extraction gas chromatography ion trap tandem mass spectrometric detection for the nanogram per liter analysis of trace pollutants in aqueous samples
    Verma, KK
    Louter, AJH
    Jain, A
    Pocurull, E
    Vreuls, JJ
    Brinkman, UAT
    CHROMATOGRAPHIA, 1997, 44 (7-8) : 372 - 380
  • [4] Determination of drugs of abuse in water by solid-phase extraction, derivatisation and gas chromatography-ion trap-tandem mass spectrometry
    Gonzalez-Marino, Iria
    Benito Quintana, Jose
    Rodriguez, Isaac
    Cela, Rafael
    JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (11) : 1748 - 1760
  • [5] Data correlation in on-line solid-phase extraction gas chromatography atomic emission mass spectrometric detection of unknown microcontaminants
    Hankemeier, T
    Rozenbrand, J
    Abhadur, M
    Vreuls, JJ
    Brinkman, UAT
    CHROMATOGRAPHIA, 1998, 48 (3-4) : 273 - 283
  • [6] On-line solid-phase extraction-gas chromatography-ion trap tandem mass spectrometric detection for the nanogram per liter analysis of trace pollutants in aqueous samples
    K. K. Verma
    A. J. H. Louter
    A. Jain
    E. Pocurull
    J. J. Vreuls
    U. A. Th. Brinkman
    Chromatographia, 1997, 44 : 372 - 380
  • [7] Sensitive determination of polycyclic aromatic hydrocarbons in water samples using monolithic capillary solid-phase extraction and on-line thermal desorption prior to gas chromatography-mass spectrometry
    Galan-Cano, Francisco
    Bernabe-Zafon, Virginia
    Lucena, Rafael
    Cardenas, Soledad
    Manuel Herrero-Martinez, Jose
    Ramis-Ramos, Guillermo
    Valcarcel, Miguel
    JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (14) : 1802 - 1807
  • [8] Data correlation in on-line solid-phase extraction-gas chromatography-atomic emission/mass spectrometric detection of unknown microcontaminants
    Th. Hankemeier
    J. Rozenbrand
    M. Abhadur
    J. J. Vreuls
    U. A. Th. Brinkman
    Chromatographia, 1998, 48 : 273 - 283
  • [9] On-line solid-phase extraction liquid chromatography, particle beam mass spectrometry, and gas chromatography mass spectrometry of carbamate pesticides
    Slobodnik, J
    HoekstraOussoren, SJF
    Jager, ME
    Honing, M
    vanBaar, BLM
    Brinkman, UAT
    ANALYST, 1996, 121 (09) : 1327 - 1334
  • [10] An improved approach for on-line solid-phase extraction gas chromatography
    Louter, AJH
    Ramalho, S
    Vreuls, RJJ
    Jahr, D
    Brinkman, UAT
    JOURNAL OF MICROCOLUMN SEPARATIONS, 1996, 8 (07) : 469 - 477