A quantitative three-dimensional dose attenuation analysis around Fletcher-Suit-Delclos due to stainless steel tube for high-dose-rate brachytherapy by Monte Carlo calculations

被引:8
|
作者
Parsai, E. Ishmael [1 ]
Zhang, Zhengdong [1 ]
John, J. Feldmeier [1 ]
机构
[1] Univ Toledo, Dept Radiat Oncol, Toledo, OH 43614 USA
关键词
FSD applicator; Attenuation; Brachytherapy; Monte Carlo simulation;
D O I
10.1016/j.brachy.2008.11.012
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The commercially available brachytherapy treatment-planning systems today, usually neglects the attenuation effect from stainless steel (SS) tube when Fletcher-Suit-Delclos (FSD) is used in treatment of cervical and endometrial cancers. This could lead to potential inaccuracies in computing dwell times and dose distribution. A more accurate analysis quantifying the level of attenuation for high-dose-rate (HDR) iridium 192 radionuclide ((192)Ir) source is presented through Monte Carlo simulation verified by measurement. METHODS AND MATERIALS: In this investigation a general Monte Carlo N-Particles (MCNP) transport code was used to construct a typical geometry of FSD through simulation and compare the doses delivered to point A in Manchester System with and without the SS tubing, A quantitative assessment of inaccuracies in delivered dose vs. the computed dose is presented. In addition, this investigation expanded to examine the attenuation-corrected radial and anisotropy dose functions in a form parallel to the updated AAPM Task Group No. 43 Report (AAPM TG-43) formalism. This will delineate quantitatively the inaccuracies in dose distributions in three-dimensional space. The changes in dose deposition and distribution caused by increased attenuation coefficient resulted from presence of SS are quantified using MCNP Monte Carlo simulations in coupled photon/electron transport. The source geometry was that of the Vari Source wire model VS2000. The FSD was that of the Varian medical system. In this model, the bending angles of tandem and colpostats are 15 degrees and 120 degrees, respectively. We assigned 10 dwell positions to the tandem and 4 dwell positions to right and left colpostats or ovoids to represent a typical treatment case. Typical dose delivered to point A was determined according to Manchester dosimetry system. RESULTS AND CONCLUSIONS: Based on our computations, the reduction of dose to point A was shown to be at least 3%. So this effect presented by SS-FSD systems on patient dose is of concern. (C) 2009 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:318 / 323
页数:6
相关论文
共 12 条
  • [1] Three-dimensional quantitative dose reduction analysis in MammoSite balloon by Monte Carlo calculations
    Zhang, Zhengdong
    Parsai, E. Ishmael
    Feldmeier, John J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (04): : 139 - 151
  • [2] High-dose-rate brachytherapy:: Dose escalation in three-dimensional miniorgans of the human bronchial wall
    Kotsianos, D
    Bach, D
    Gamarra, F
    Haimerl, W
    Knüchel, R
    Lang, SM
    Pfeifer, KJ
    Huber, RM
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (05): : 1267 - 1273
  • [3] Definitive three-dimensional high-dose-rate brachytherapy for inoperable endometrial cancer
    Draghini, Lorena
    Maranzano, Ernesto
    Casale, Michelina
    Trippa, Fabio
    Anselmo, Paola
    Arcidiacono, Fabio
    Fabiani, Stefania
    Italiani, Marco
    Chirico, Luigia
    Muti, Marco
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2017, 9 (02) : 118 - 123
  • [4] Dosimetry comparison between TG-43 and Monte Carlo calculations using the Freiburg flap for skin high-dose-rate brachytherapy
    Vijande, Javier
    Ballester, Facundo
    Ouhib, Zoubir
    Granero, Domingo
    Carmen Pujades-Claumarchirant, M.
    Perez-Calatayud, Jose
    BRACHYTHERAPY, 2012, 11 (06) : 528 - 535
  • [5] CONE BEAM CT-BASED THREE-DIMENSIONAL PLANNING IN HIGH-DOSE-RATE BRACHYTHERAPY FOR CERVICAL CANCER
    Al-Halabi, Hani
    Portelance, Lorraine
    Duclos, Marie
    Reniers, Brigitte
    Bahoric, Boris
    Souhami, Luis
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 77 (04): : 1092 - 1097
  • [6] Three-dimensional image-guided combined intracavitary and interstitial high-dose-rate brachytherapy in cervical cancer: A systematic review
    Li, Fei
    Lu, Shuangchen
    Zhao, Hongfu
    Mu, Xin
    Mao, Zhuang
    BRACHYTHERAPY, 2021, 20 (01) : 85 - 94
  • [7] Dosimetric impact of applicator displacement on three-dimensional image-guided high-dose-rate brachytherapy treatments for cervical cancer
    Jayarathna, Sandun
    Hoang, Matthew
    Badkul, Rajeev
    Hoover, Andrew
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2023, 15 (05) : 334 - 343
  • [8] Three-dimensional source position verification in image-guided high-dose-rate brachytherapy using an XCT-based gel dosimeter
    Yonemura, Miki
    Tachibana, Hidenobu
    Kojima, Toru
    Seki, Kazuya
    Nakaichi, Tetsu
    Rachi, Toshiya
    Tachibana, Rie
    Akimoto, Tetsuo
    MEDICAL PHYSICS, 2025, 52 (02) : 1243 - 1255
  • [9] Dosimetric comparison between three dimensional treatment planning system, Monte Carlo simulation and gel dosimetry in nasopharynx phantom for high dose rate brachytherapy
    Fazli, Zeynab
    Sadeghi, Mahdi
    Zahmatkesh, M. H.
    Mahdavi, Seied Rabei
    Tenreiro, Claudio
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2013, 9 (03) : 402 - 409
  • [10] Influence of transitioning of planning techniques in high-dose-rate brachytherapy monotherapy for clinically localized prostate cancer from two- to three-dimensional planning
    Yamazaki, Hideya
    Masui, Koji
    Suzuk, Gen
    Nakamur, Satoaki
    Yoshida, Ken
    Kotsuma, Tadayuki
    Tanaka, Eiichi
    Otani, Keisuke
    Yoshioka, Yasuo
    Ogawa, Kazuhiko
    BRACHYTHERAPY, 2019, 18 (05) : 589 - 597