Speech Intelligibility in Noise With a Pinna Effect Imitating Cochlear Implant Processor

被引:31
作者
Wimmer, Wilhelm [1 ,2 ]
Weder, Stefan [2 ]
Caversaccio, Marco [1 ,2 ]
Kompis, Martin [2 ]
机构
[1] Univ Bern, Artificial Hearing Res, ARTORG Ctr Biomed Engn, Inselspital, CH-3010 Bern, Switzerland
[2] Univ Bern, Inselspital, Dept ENT Head & Neck Surg, CH-3010 Bern, Switzerland
基金
瑞士国家科学基金会;
关键词
Directional microphone systems; Oldenburg sentence test; Sonnet; Speech reception threshold; Two-microphone noise reduction; BACKGROUND-NOISE; RECOGNITION; MICROPHONE; PERCEPTION; INTERFERENCE; STRATEGIES; RECIPIENTS; ALGORITHM; OPTIONS; SINGLE;
D O I
10.1097/MAO.0000000000000866
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective:To evaluate the speech intelligibility in noise with a new cochlear implant (CI) processor that uses a pinna effect imitating directional microphone system.Study Design:Prospective experimental study.Setting:Tertiary referral center.Patients:Ten experienced, unilateral CI recipients with bilateral severe-to-profound hearing loss.Intervention:All participants performed speech in noise tests with the Opus 2 processor (omnidirectional microphone mode only) and the newer Sonnet processor (omnidirectional and directional microphone mode).Main Outcome Measure:The speech reception threshold (SRT) in noise was measured in four spatial settings. The test sentences were always presented from the front. The noise was arriving either from the front (S0N0), the ipsilateral side of the CI (S0NIL), the contralateral side of the CI (S0NCL), or the back (S0N180).Results:The directional mode improved the SRTs by 3.6dB (p<0.01), 2.2dB (p<0.01), and 1.3dB (p<0.05) in the S0N180, S0NIL, and S0NCL situations, when compared with the Sonnet in the omnidirectional mode. There was no statistically significant difference in the S0N0 situation. No differences between the Opus 2 and the Sonnet in the omnidirectional mode were observed.Conclusion:Speech intelligibility with the Sonnet system was statistically different to speech recognition with the Opus 2 system suggesting that CI users might profit from the pinna effect imitating directionality mode in noisy environments.
引用
收藏
页码:19 / 23
页数:5
相关论文
共 24 条
  • [1] Blauert J., 1997, SPATIAL HEARING PSYC, P108
  • [2] Evaluation of Different Signal Processing Options in Unilateral and Bilateral Cochlear Freedom Implant Recipients Using R-Space™ Background Noise
    Brockmeyer, Alison M.
    Potts, Lisa G.
    [J]. JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2011, 22 (02) : 65 - 80
  • [3] Advanced Beamformers for Cochlear Implant Users: Acute Measurement of Speech Perception in Challenging Listening Conditions
    Buechner, Andreas
    Dyballa, Karl-Heinz
    Hehrmann, Phillipp
    Fredelake, Stefan
    Lenarz, Thomas
    [J]. PLOS ONE, 2014, 9 (04):
  • [4] Effects of directional microphone and adaptive multichannel noise reduction algorithm on cochlear implant performance
    Chung, King
    Zeng, Fan-Gang
    Acker, Kyle N.
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2006, 120 (04) : 2216 - 2227
  • [5] Using hearing aid adaptive directional microphones to enhance cochlear implant performance
    Chung, King
    Zeng, Fan-Gang
    [J]. HEARING RESEARCH, 2009, 250 (1-2) : 27 - 37
  • [6] Speech Perception for Adult Cochlear Implant Recipients in a Realistic Background Noise: Effectiveness of Preprocessing Strategies and External Options for Improving Speech Recognition in Noise
    Gifford, Rene H.
    Revit, Lawrence J.
    [J]. JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2010, 21 (07) : 441 - 451
  • [7] Two-Microphone Spatial Filtering Improves Speech Reception for Cochlear-Implant Users in Reverberant Conditions With Multiple Noise Sources
    Goldsworthy, Raymond L.
    [J]. TRENDS IN HEARING, 2014, 18
  • [8] INTELLIGIBILITY-WEIGHTED MEASURES OF SPEECH-TO-INTERFERENCE RATIO AND SPEECH SYSTEM PERFORMANCE
    GREENBERG, JE
    PETERSON, PM
    ZUREK, PM
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1993, 94 (05) : 3009 - 3010
  • [9] Database of Multichannel In-Ear and Behind-the-Ear Head-Related and Binaural Room Impulse Responses
    Kayser, H.
    Ewert, S. D.
    Anemueller, J.
    Rohdenburg, T.
    Hohmann, V.
    Kollmeier, B.
    [J]. EURASIP JOURNAL ON ADVANCES IN SIGNAL PROCESSING, 2009,
  • [10] Single and Multiple Microphone Noise Reduction Strategies in Cochlear Implants
    Kokkinakis, Kostas
    Azimi, Behnam
    Hu, Yi
    Friedland, David R.