Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review

被引:46
作者
Jabran, Ali [1 ]
Peach, Chris [1 ,2 ]
Ren, Lei [1 ]
机构
[1] Univ Manchester, Sch Mech Aerosp & Civil Engn, Manchester, Lancs, England
[2] Univ Hosp South Manchester, Dept Shoulder & Elbow Surg, Manchester, Lancs, England
来源
BIOMEDICAL ENGINEERING ONLINE | 2018年 / 17卷
基金
英国工程与自然科学研究理事会;
关键词
Proximal humerus fractures; Locking plates; Open reduction internal fixation; Biomechanical testing; SURGICAL NECK FRACTURES; INTRAMEDULLARY FIBULAR ALLOGRAFT; CALCIUM-PHOSPHATE CEMENT; HUMAN SHOULDER SPECIMENS; MEDIAL CORTICAL SUPPORT; FINITE-ELEMENT-ANALYSIS; ANGULAR STABLE PLATE; ROTATOR CUFF SUTURES; IN-VITRO ASSESSMENT; LOCKING PLATE;
D O I
10.1186/s12938-018-0479-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Proximal humerus fractures are the third most common in the human body but their management remains controversial. Open reduction and internal fixation with plates is one of the leading modes of operative treatment for these fractures. The development of technologies and techniques for these plates, during the recent decades, promise a bright future for their clinical use. A comprehensive review of in vitro biomechanical studies is needed for the comparison of plates' mechanical performance and the testing methodologies. This will not only guide clinicians with plate selection but also with the design of future in vitro biomechanical studies. This review was aimed to systematically categorise and review the in vitro biomechanical studies of these plates based on their protocols and discuss their results. The technologies and techniques investigated in these studies were categorised and compared to reach a census where possible. Methods and results: Web of Science and Scopus database search yielded 62 studies. Out of these, 51 performed axial loading, torsion, bending and/or combined bending and axial loading while 11 simulated complex glenohumeral movements by using tendons. Loading conditions and set-up, failure criteria and performance parameters, as well as results for each study, were reviewed. Only two studies tested four-part fracture model while the rest investigated two-and three-part fractures. In ten studies, synthetic humeri were tested instead of cadaveric ones. In addition to load-displacement data, three-dimensional motion analysis systems, digital image correlation and acoustic emission testing have been used for measurement. Conclusions: Overall, PHILOS was the most tested plate and locking plates demonstrated better mechanical performance than non-locking ones. Conflicting results have been published for their comparison with non-locking blade plates and polyaxial locking screws. Augmentation with cement [calcium phosphate or poly(methyl methacrylate)] or allografts (fibular and femoral head) was found to improve bone-plate constructs' mechanical performance. Controversy still lies over the use of rigid and semi-rigid implants and the insertion of inferomedial screws for calcar region support. This review will guide the design of in vitro and in silico biomechanical tests and also supplement the study of clinical literature.
引用
收藏
页数:30
相关论文
共 126 条
  • [1] Treatment of proximal humeral fractures with Polarus nail fixation
    Agel, J
    Jones, CB
    Sonzone, AG
    Camuso, M
    Henley, MB
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2004, 13 (02) : 191 - 195
  • [2] Biomechanical Contribution of Tension-Reducing Rotator Cuff Sutures in 3-Part Proximal Humerus Fractures
    Arvesen, John E.
    Gill, Stephen W.
    Sinatra, Philip M.
    Eng, Michael
    Bledsoe, Gary
    Kaar, Scott G.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2016, 30 (08) : E262 - E266
  • [3] Impact of shoulder rotation on neck-shaft angle: A clinical study
    Assuncao, J. H.
    Malavolta, E. A.
    Beraldo, R. A.
    Gracitelli, M. E. C.
    Bordalo-Rodrigues, M.
    Ferreira Neto, A. A.
    [J]. ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2017, 103 (06) : 865 - 868
  • [4] The biomechanical performance of locking plate fixation with intramedullary fibular strut graft augmentation in the treatment of unstable fractures of the proximal humerus
    Bae, J. -H.
    Oh, J. -K.
    Chon, C. -S.
    Oh, C. -W.
    Hwang, J. -H.
    Yoon, Y. -C.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (07): : 937 - 941
  • [5] Bai L, 2014, J ORTHOP TRAUMA, V28, P452, DOI 10.1097/BOT.0000000000000057
  • [6] Basic epidemiology of fractures of the upper and lower limb among Americans over 65 years of age
    Baron, JA
    Karagas, M
    Barrett, J
    Kniffin, W
    Malenka, D
    Mayor, M
    Keller, RB
    [J]. EPIDEMIOLOGY, 1996, 7 (06) : 612 - 618
  • [7] The epidemiology of peripheral fractures
    Baron, JA
    Barrett, JA
    Karagas, MR
    [J]. BONE, 1996, 18 (03) : S209 - S213
  • [8] In vivo glenohumeral contact forces -: Measurements in the first patient 7 months postoperatively
    Bergmann, G.
    Graichen, F.
    Bender, A.
    Kaab, M.
    Rohlmann, A.
    Westerhoff, P.
    [J]. JOURNAL OF BIOMECHANICS, 2007, 40 (10) : 2139 - 2149
  • [9] In vivo gleno-humeral joint loads during forward flexion and abduction
    Bergmann, G.
    Graichen, F.
    Bender, A.
    Rohlmann, A.
    Halder, A.
    Beier, A.
    Westerhoff, P.
    [J]. JOURNAL OF BIOMECHANICS, 2011, 44 (08) : 1543 - 1552
  • [10] The three-dimensional geometry of the proximal humerus - Implications for surgical technique and prosthetic design
    Boileau, P
    Walch, G
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (05): : 857 - 865