Monte carlo simulation study of a novel parallel-hole collimator with a CdTe pixelated semiconductor SPECT system

被引:1
|
作者
Lee, Young-Jin [1 ]
Kim, Dae-Hong [1 ]
Rhee, Yong-Chun [1 ]
Kim, Hee-Joung [1 ]
机构
[1] Yonsei Univ, Coll Hlth Sci, Dept Radiol Sci, Wonju 220710, South Korea
基金
新加坡国家研究基金会;
关键词
Cadmium telluride (CdTe) pixelated semiconductor detector; Single-photon-emission computed tomography (SPECT); Novel parallel-hole collimator; Geant4 Application for Tomographic Emission (GATE) simulation; NUCLEAR-MEDICINE; GAMMA-CAMERA; GATE; RECONSTRUCTION; FEASIBILITY; PERFORMANCE; TOMOGRAPHY; VALIDATION; DETECTORS;
D O I
10.3938/jkps.64.1737
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Recently, many studies have investigated the use of a pixelated semiconductor detector to improve spatial resolution. The purpose of this study was to evaluate novel parallel-hole collimator geometric designs with a CdTe pixelated semiconductor single-photon-emission computed tomography (SPECT) system. The pixelated semiconductor detector was modeled as a PID 350 detector (Ajat Oy Ltd., Finland) with small pixels (0.35 x 0.35 mm(2)) by using Geant4 Application for Tomographic Emission (GATE) software. We designed a novel parallel-hole collimator consisting of two overlapping parallel-hole collimators. Each hole size was four times that of the pixelated parallel-hole collimator. The overlap ratios of these collimators were 1:1, 1:2, 2:1, 1:3, 3:1, 1:4, and 4:1. To evaluate the performance of this system, we evaluated the sensitivity and the spatial resolution. The results for our new parallel-hole collimator indicated that the evaluated sensitivity averages using overlap ratios of 1:1, 1:2, 2:1, 1:3, 3:1, 1:4, and 4:1 were 4.45, 7.56, 7.51, 12.76, 12.65, 20.01, and 19.90 times higher, respectively, than those of the pixelated parallel-hole collimator. The evaluated averages of the spatial resolution varied depending on the source-to-collimator distances. In conclusion, we successfully designed a novel parallel-hole collimator with various overlap ratios of the collimator septal heights with a CdTe pixelated semiconductor SPECT system. Based on our results, we recommend using this collimator with a CdTe pixelated semiconductor SPECT system.
引用
收藏
页码:1737 / 1744
页数:8
相关论文
共 50 条
  • [1] Monte carlo simulation study of a novel parallel-hole collimator with a CdTe pixelated semiconductor SPECT system
    Young-Jin Lee
    Dae-Hong Kim
    Yong-Chun Rhee
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2014, 64 : 1737 - 1744
  • [2] A Monte Carlo simulation study of the feasibility of a high resolution parallel-hole collimator with a CdTe pixelated semiconductor SPECT system
    Lee, Y. -J.
    Park, S. -J.
    Lee, S. -W.
    Kim, D. -H.
    Kim, Y. -S.
    Jo, B. -D.
    Kim, H. -J.
    JOURNAL OF INSTRUMENTATION, 2013, 8
  • [3] The Effect of High-resolution Parallel-hole Collimator Materials with a Pixelated Semiconductor SPECT System at Equivalent Sensitivities: Monte Carlo Simulation Studies
    Lee, Young-Jin
    Kim, Dae-Hong
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (07) : 1055 - 1062
  • [4] The effect of high-resolution parallel-hole collimator materials with a pixelated semiconductor SPECT system at equivalent sensitivities: Monte Carlo simulation studies
    Young-Jin Lee
    Dae-Hong Kim
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2014, 64 : 1055 - 1062
  • [5] Comparison of ultra-high-resolution parallel-hole collimator materials based on the CdTe pixelated semiconductor SPECT system
    Lee, Young-Jin
    Ryu, Hyun-Ju
    Lee, Seung-Wan
    Park, Su-Jin
    Kim, Hee-Joung
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 713 : 33 - 39
  • [6] Characterization of parallel-hole collimator using Monte Carlo Simulation
    Pandey, Anil Kumar
    Sharma, Sanjay Kumar
    Karunanithi, Sellam
    Kumar, Praveen
    Bal, Chandrasekhar
    Kumar, Rakesh
    INDIAN JOURNAL OF NUCLEAR MEDICINE, 2015, 30 (02): : 128 - 134
  • [7] Optimization of an ultra-high-resolution parallel-hole collimator for CdTe semiconductor SPECT system
    Lee, Y. -J.
    Ryu, H. -J.
    Cho, H. -M.
    Lee, S. -W.
    Choi, Y. -N.
    Kim, H. -J.
    JOURNAL OF INSTRUMENTATION, 2013, 8
  • [8] Optimization of the SPECT systems based on a CdTe pixelated semiconductor detector using novel parallel-hole collimators
    Lee, Y. -J.
    Park, S. -J.
    Kim, D. -H.
    Kim, H. -J.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [9] Performance evaluation of high-resolution square parallel-hole collimators with a CZT room temperature pixelated semiconductor SPECT system: a Monte Carlo simulation study
    Lee, Y.
    Kang, W.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [10] Evaluation of the performance of parallel-hole collimator for high resolution small animal SPECT: A Monte Carlo study
    Sadremomtaz, Alireza
    Telikani, Zeinab
    IRANIAN JOURNAL OF NUCLEAR MEDICINE, 2016, 24 (02): : 136 - 143