Comparison of T1-weighted 2D TSE, 3D SPGR, and two-point 3D Dixon MRI for automated segmentation of visceral adipose tissue at 3 Tesla

被引:17
作者
Fallah, Faezeh [1 ,2 ]
Machann, Juergen [2 ,3 ,4 ]
Martirosian, Petros [2 ]
Bamberg, Fabian [5 ]
Schick, Fritz [2 ]
Yang, Bin [1 ]
机构
[1] Univ Stuttgart, Inst Signal Proc & Syst Theory, Pfaffenwaldring 47, D-70569 Stuttgart, Germany
[2] Univ Clin Tubingen, Sect Expt Radiol, Dept Diagnost & Intervent Radiol, Tubingen, Germany
[3] Univ Tubingen, Inst Diabet Res & Metab Dis IDM, Helmholtz Ctr Munich, Tubingen, Germany
[4] German Ctr Diabet Res DZD, Tubingen, Germany
[5] Univ Clin Tubingen, Dept Diagnost & Intervent Radiol, Tubingen, Germany
关键词
Visceral adipose tissue; Automatic volumetry; Two-point Dixon imaging; T1-weighted imaging; 3; Tesla; FAT; QUANTIFICATION; WATER; CLASSIFICATION; ACCUMULATION; TOPOGRAPHY; SEPARATION; IDEAL; DEEP;
D O I
10.1007/s10334-016-0588-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To evaluate and compare conventional T1-weighted 2D turbo spin echo (TSE), T1-weighted 3D volumetric interpolated breath-hold examination (VIBE), and two-point 3D Dixon-VIBE sequences for automatic segmentation of visceral adipose tissue (VAT) volume at 3 Tesla by measuring and compensating for errors arising from intensity nonuniformity (INU) and partial volume effects (PVE). The body trunks of 28 volunteers with body mass index values ranging from 18 to 41.2 kg/m(2) (30.02 +/- 6.63 kg/m(2)) were scanned at 3 Tesla using three imaging techniques. Automatic methods were applied to reduce INU and PVE and to segment VAT. The automatically segmented VAT volumes obtained from all acquisitions were then statistically and objectively evaluated against the manually segmented (reference) VAT volumes. Comparing the reference volumes with the VAT volumes automatically segmented over the uncorrected images showed that INU led to an average relative volume difference of -59.22 +/- 11.59, 2.21 +/- 47.04, and -43.05 +/- 5.01 % for the TSE, VIBE, and Dixon images, respectively, while PVE led to average differences of -34.85 +/- 19.85, -15.13 +/- 11.04, and -33.79 +/- 20.38 %. After signal correction, differences of -2.72 +/- 6.60, 34.02 +/- 36.99, and -2.23 +/- 7.58 % were obtained between the reference and the automatically segmented volumes. A paired-sample two-tailed t test revealed no significant difference between the reference and automatically segmented VAT volumes of the corrected TSE (p = 0.614) and Dixon (p = 0.969) images, but showed a significant VAT overestimation using the corrected VIBE images. Under similar imaging conditions and spatial resolution, automatically segmented VAT volumes obtained from the corrected TSE and Dixon images agreed with each other and with the reference volumes. These results demonstrate the efficacy of the signal correction methods and the similar accuracy of TSE and Dixon imaging for automatic volumetry of VAT at 3 Tesla.
引用
收藏
页码:139 / 151
页数:13
相关论文
共 52 条
  • [1] Validation of Volumetric and Single-Slice MRI Adipose Analysis Using a Novel Fully Automated Segmentation Method
    Addeman, Bryan T.
    Kutty, Shelby
    Perkins, Thomas G.
    Soliman, Abraam S.
    Wiens, Curtis N.
    McCurdy, Colin M.
    Beaton, Melanie D.
    Hegele, Robert A.
    McKenzie, Charles A.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2015, 41 (01) : 233 - 241
  • [2] Evaluation of Adipose Tissue Volume Quantification With IDEAL Fat-Water Separation
    Alabousi, Abdullah
    Al-Attar, Salam
    Joy, Tisha R.
    Hegele, Robert A.
    McKenzie, Charles A.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 34 (02) : 474 - 479
  • [3] [Anonymous], 2004, HDB MRI PULSE SEQUEN, DOI 10.1016/B978-0-12-092861-3.X5000-6
  • [4] Whole-Body MR Imaging in the German National Cohort:Rationale, Design, and Technical Background
    Bamberg, Fabian
    Kauczor, Hans-Ulrich
    Weckbach, Sabine
    Schlett, Christopher L.
    Forsting, Michael
    Ladd, Susanne C.
    Greiser, Karin Halina
    Weber, Marc-Andre
    Schulz-Menger, Jeanette
    Niendorf, Thoralf
    Pischon, Tobias
    Caspers, Svenja
    Amunts, Katrin
    Berger, Klaus
    Buelow, Robin
    Hosten, Norbert
    Hegenscheid, Katrin
    Kroencke, Thomas
    Linseisen, Jakob
    Guenther, Matthias
    Hirsch, Jochen G.
    Koehn, Alexander
    Hendel, Thomas
    Wichmann, Heinz-Erich
    Schmidt, Boege
    Joeckel, Karl-Heinz
    Hoffmann, Wolfgang
    Kaaks, Rudolf
    Reiser, Maximilian F.
    Voelzke, Henry
    [J]. RADIOLOGY, 2015, 277 (01) : 206 - 220
  • [5] Fuzzy c-means approach to tissue classification in multimodal medical imaging
    Banerjee, S
    Mukherjee, DP
    Majumdar, DD
    [J]. INFORMATION SCIENCES, 1999, 115 (1-4) : 261 - 279
  • [6] The minimum description length principle in coding and modeling
    Barron, A
    Rissanen, J
    Yu, B
    [J]. IEEE TRANSACTIONS ON INFORMATION THEORY, 1998, 44 (06) : 2743 - 2760
  • [7] Bauer JD, 2015, PLOS ONE, V10, P1
  • [8] Model-based mapping of fat unsaturation and chain length by chemical shift imaginguphantom validation and in vivo feasibility
    Berglund, Johan
    Ahlstrom, Hakan
    Kullberg, Joel
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (06) : 1815 - 1827
  • [9] Boyle Gerard E, 2006, Radiographics, V26, pe24, DOI 10.1148/rg.e24
  • [10] Fat-Suppression Techniques for 3-T MR Imaging of the Musculoskeletal System
    Del Grande, Filippo
    Santini, Francesco
    Herzka, Daniel A.
    Aro, Michael R.
    Dean, Cooper W.
    Gold, Garry E.
    Carrino, John A.
    [J]. RADIOGRAPHICS, 2014, 34 (01) : 217 - 233