Feasibility of a beta-gamma digital imaging probe for radioguided surgery

被引:15
|
作者
Tipnis, SV [1 ]
Nagarkar, VV
Shestakova, I
Gaysinskiy, V
Entine, G
Tornai, MP
Stack, BC
机构
[1] Radiat Monitoring Devices Inc, Watertown, MA 02472 USA
[2] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27706 USA
[3] Penn State Univ, Coll Med, Milton S Hershey Med Ctr, Dept Otolaryngol, University Pk, PA 16802 USA
基金
美国国家卫生研究院;
关键词
cesium iodide; charge coupled devices; nuclear imaging; radioguided surgery; surgical probes;
D O I
10.1109/TNS.2003.823022
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We report here on a novel design of a digital, intraoperative imaging probe intended for use,in radio-guided surgical procedures in conjunction with radiolabels such as I-131 and F-18. The probe allows the user to rapidly localize tumors by detecting the highly penetrating gamma rays, and then image the tumor with the short-range beta rays. The system provides a rapid, high-resolution, image of the interrogated area, fulfilling the need for clear delineation of tumors during radio-guided surgical procedures. The beta imaging sensor consists of a micro-columnar CsI(Tl) scintillator screen capable of providing very high detection efficiency, high light output and excellent spatial resolution coupled to a CCD via a flexible, coherent fiberoptic bundle. The gamma sensor is a shielded piece of crystalline CsI(Tl) coupled to a photodiode located behind the image sensitive front end. The feasibility of this design was studied by separately testing the beta imaging and gamma detection components. The operation of the components was characterized with intrinsic performance measurements of count rates, signal-to-noise ratios, spatial resolution, as well as time for acquiring useful images using radionuclide and anthropomorphic phantoms.
引用
收藏
页码:110 / 116
页数:7
相关论文
共 50 条
  • [1] Feasibility studies of an EMCCD-based beta imaging probe for radioguided thyroid surgery
    Shestakova, Irina
    Nagarkar, Vivek V.
    Gaysinskiy, Valeriy
    Entine, Gerald
    Stack, Brendan C.
    Miller, Brian
    HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS AND PENETRATING RADIATION SYSTEMS VIII, 2006, 6319
  • [2] High resolution beta imaging probe for radioguided surgery
    Shestakova, I.
    Gaysinskiy, V. B.
    Thacker, S. C.
    Cool, S.
    Nagarkar, V. V.
    Stack, B. C.
    PENETRATING RADIATION SYSTEMS AND APPLICATIONS VIII, 2007, 6707
  • [3] A comprehensive overview of radioguided surgery using gamma detection probe technology
    Povoski, Stephen P.
    Neff, Ryan L.
    Mojzisik, Cathy M.
    O'Malley, David M.
    Hinkle, George H.
    Hall, Nathan C.
    Murrey, Douglas A., Jr.
    Knopp, Michael V.
    Martin, Edward W., Jr.
    WORLD JOURNAL OF SURGICAL ONCOLOGY, 2009, 7
  • [4] Physical performance of an intraoperative beta probe dedicated to glioma radioguided surgery
    Bogalhas, Frederic
    Menard, Laurent
    Bonzom, Sebatien
    Palfi, Stephane
    Siebert, Rainer
    Duval, Marie-Alix
    Lefebvre, Francoise
    Pinot, Laurent
    Pitre, Stephanie
    Charon, Yves
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) : 833 - 841
  • [5] Radioguided surgery using gamma detection probe technology for resection of cerebral glioma
    Seddighi, Afsoun
    Akbari, Mohammad Esmail
    Seddighi, Amir Saied
    Pirayesh, Elaheh
    Soleymani, Mohammad Mehdi
    Baqdashti, Hesam Rahimi
    Nikouei, Amir
    Zali, Alireza
    Tabatabaei, Seyed Mahmod
    Naimian, Shoeib
    Mellati, Omid
    Sheikhi, Alireza
    HELLENIC JOURNAL OF NUCLEAR MEDICINE, 2015, 18 (03): : 68 - 75
  • [6] Feasibility of beta-particle radioguided surgery for a variety of "nuclear medicine" radionuclides
    Mancini-Terracciano, Carlo
    Donnarumma, Raffaella
    Bencivenga, Gaia
    Bocci, Valerio
    Cartoni, Antonella
    Collamati, Francesco
    Fratoddi, Ilaria
    Giordano, Alessandro
    Indovina, Luca
    Maccora, Daria
    Marafini, Michela
    Mirabelli, Riccardo
    Morganti, Silvio
    Rotili, Dante
    Russomando, Andrea
    Scotognella, Teresa
    Camillocci, Elena Solfaroli
    Toppi, Marco
    Traini, Giacomo
    Venditti, Iole
    Faccini, Riccardo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 43 : 127 - 133
  • [7] Imaging skins: stretchable and conformable on-organ beta particle detectors for radioguided surgery
    Dietsch, S.
    Lindenroth, L.
    Stilli, A.
    Stoyanov, D.
    PROGRESS IN BIOMEDICAL ENGINEERING, 2023, 5 (03):
  • [8] CdTe compact gamma camera for coded aperture imaging in radioguided surgery
    Russo, Paolo
    Di Lillo, Francesca
    Corvino, Veronica
    Frallicciardi, Paola Maria
    Sarno, Antonio
    Mettivier, Giovanni
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 69 : 223 - 232
  • [9] Multimodal imaging in radioguided surgery
    KleinJan G.H.
    Bunschoten A.
    Brouwer O.R.
    van den Berg N.S.
    Valdés-Olmos R.A.
    van Leeuwen F.W.B.
    Clinical and Translational Imaging, 2013, 1 (6) : 433 - 444
  • [10] A prototype small CdTe gamma camera for radioguided surgery and other imaging applications
    Makoto Tsuchimochi
    Harumi Sakahara
    Kazuhide Hayama
    Minoru Funaki
    Ryoichi Ohno
    Takashi Shirahata
    Terje Orskaug
    Gunnar Maehlum
    Koki Yoshioka
    Einar Nygard
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 1605 - 1614