B0-Informed Variable Density Trajectory Design for Enhanced Correction of Off-Resonance Effects in Parallel Transmission

被引:8
|
作者
Schneider, Rainer [1 ,2 ]
Ritter, Dieter [1 ]
Haueisen, Jens [2 ]
Pfeuffer, Josef [1 ]
机构
[1] Siemens Healthcare, MR Applicat Dev, Erlangen, Germany
[2] Ilmenau Univ Technol, Inst Biomed Engn & Informat, Ilmenau, Germany
关键词
parallel transmit; selective excitation; off-resonance; variable density sampling; k-space trajectory; SPATIALLY SELECTIVE EXCITATION; TAILORED RF PULSES; RADIOFREQUENCY PULSE; SPIRAL TRAJECTORIES; SPINAL-CORD; MRI; OPTIMIZATION; ARTIFACTS; FIELDS; ARRAY;
D O I
10.1002/mrm.24780
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo improve B1 and B0 inhomogeneity mitigation performance of spatially selective radio-frequency (RF) pulses in parallel transmission while decreasing RF pulse power. Further enhancement of off-resonance correction for rectilinear spoke-trajectory-based RF pulses with known residual geometric distortions after optimization. MethodsThe appropriate definition of the target magnetization pattern is discussed regarding the maximum physical excitation resolution. Furthermore, a novel variable-density trajectory design is introduced, which subsamples accrued B0 phase error elevations in k-space. A simulation study (echo-planar and spiral 2DRF) at different off-resonance levels and pulse acceleration factors was pursued using data from a whole-body 2-channel parallel transmit 3T MRI system. The new trajectory design for echo-planar 2DRF was validated in human in-vivo experiments. ResultsProper target pattern definition can require spatial filtering, such that RF pulse optimization is prevented from lower excitation performance with significant higher RF power level. The new trajectory design proposed can considerably improve off-resonance compensation, while further reducing the RF power, e.g., 43% less RMSE with 79% less RF power for spoke based pulses. ConclusionThe proposed methods offer significant improvements of the excitation performance (homogeneity and acceleration), while significantly decreasing the RF power. Furthermore, single-channel transmit RF pulse performance can be similarly improved. Magn Reson Med 71:1381-1393, 2014. (c) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:1381 / 1393
页数:13
相关论文
共 3 条
  • [1] Method for B0 Off-Resonance Mapping by Non-Iterative Correction of Phase-Errors (B0-NICE)
    Liu, Junmin
    Drangova, Maria
    MAGNETIC RESONANCE IN MEDICINE, 2015, 74 (04) : 1177 - 1188
  • [2] Robust Time-Shifted Spoke Pulse Design in the Presence of Large B0 Variations with Simultaneous Reduction of Through-Plane Dephasing, B1+Effects, and the Specific Absorption Rate Using Parallel Transmission
    Guerin, Bastien
    Stockmann, Jason P.
    Baboli, Mehran
    Torrado-Carvajal, Angel
    Stenger, Andrew V.
    Wald, Lawrence L.
    MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (02) : 540 - 554
  • [3] Image quality of Gd-EOB-DTPA-enhanced magnetic resonance imaging of the liver using dual-source parallel radiofrequency transmission technology: Comparison with the post-processing correction method for B1 inhomogeneity-induced signal loss
    Takayama, Yukihisa
    Nishie, Akihiro
    Asayama, Yoshiki
    Ishigami, Kousei
    Kakihara, Daisuke
    Nakayama, Tomohiro
    Yoshiura, Takashi
    Obara, Makoto
    Hirakawa, Masakazu
    Honda, Hiroshi
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (11) : 3035 - 3040