A Novel Semiautomatic Parenchyma Extraction Method for Improved MRI R2*Relaxometry of Iron Loaded Liver

被引:6
|
作者
Feng, Yanqiu [1 ]
Feng, Meiyan [1 ]
Gao, Huashuai [1 ]
Zhang, Xinyuan [1 ]
Xin, Xuegang [1 ]
Feng, Qianjin [1 ]
Chen, Wufan [1 ]
He, Taigang [2 ,3 ]
机构
[1] Southern Med Univ, Sch Biomed Engn, Guangzhou, Guangdong, Peoples R China
[2] St Georges Univ London, Cardiovasc Sci Res Ctr, London, England
[3] Royal Brompton Hosp, Cardiovasc Biomed Res Unit, London SW3 6LY, England
基金
英国惠康基金;
关键词
magnetic resonance imaging; R2*; liver iron; semiautomatic parenchyma extraction; HEPATIC IRON; MAGNETIC-RESONANCE; OVERLOAD; THALASSEMIA; DIAGNOSIS; QUANTIFICATION; THERAPY; STORES;
D O I
10.1002/jmri.24331
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To propose and evaluate an automatic method of extracting parenchyma from a manually delineated whole liver for the R2* measurement of iron load. Materials and Methods: In all, 108 transfusion-dependent patients with a wide range of hepatic iron content were scanned with a multiecho gradient-echo sequence. The R2* was measured by fitting the average signal of liver parenchyma, extracted by the proposed semiautomatic parenchyma extraction (SAPE), traditional manually delineated multiple regions-of-interest (mROIs), and T2* thresholding methods to the noise-corrected monoexponential model. The R2* measurement accuracy of the SAPE method was evaluated through simulation; the intra- and interobserver reproducibility of SAPE, mROI, and T2* thresholding were assessed from the in vivo data using coefficient of variation (CoV). Results: In the simulation, the mean absolute percentage error of R2* measurement using SAPE was 0.23% (range 0.01%-1.09%). In vivo study, the CoVs of intra- and interobserver reproducibility were 0.83%, 1.39% for SAPE, 3.63%, 6.28% for mROI, and 1.62%, 2.66% for T2* thresholding, respectively. Conclusion: The SAPE method provides an accurate and reliable approach to assessing the overall hepatic iron content. The improved magnetic resonance imaging (MRI) R2* reproducibility using the SAPE method may lead to more accurate tissue characterization and increased diagnostic confidence.
引用
收藏
页码:67 / 78
页数:12
相关论文
共 50 条
  • [21] The Effect of Reducing Repetition Time TR on the Measurement of Liver R2 for the Purpose of Measuring Liver Iron Concentration
    Pavitt, Helen L.
    Aydinok, Yesim
    El-Beshlawy, Amal
    Bayraktaroglu, Selen
    Ibrahim, Ahmed S.
    Hamdy, Mona M.
    Pang, Wenjie
    Sharples, Christopher
    St Pierre, Timothy G.
    MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (05) : 1346 - 1351
  • [22] Rapid Look-Up Table Method for Noise-Corrected Curve Fitting in the R2*Mapping of Iron Loaded Liver
    Wang, Changqing
    He, Taigang
    Liu, Xiaoyun
    Zhong, Shouming
    Chen, Wufan
    Feng, Yanqiu
    MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (02) : 865 - 871
  • [23] Histological Scores Validate the Accuracy of Hepatic Iron Load Measured by Signal Intensity Ratio and R2*Relaxometry MRI in Dialysis Patients
    Rostoker, Guy
    Laroudie, Mireille
    Blanc, Raphael
    Griuncelli, Mireille
    Loridon, Christelle
    Lepeytre, Fanny
    Rabate, Clementine
    Cohen, Yves
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (01)
  • [24] R2 relaxometry based MR imaging for estimation of liver iron content: A comparison between two methods
    Juan S. Calle-Toro
    Christian A. Barrera
    Dmitry Khrichenko
    Hansel J. Otero
    Suraj D. Serai
    Abdominal Radiology, 2019, 44 : 3058 - 3068
  • [25] Focal sparing of iron and fat in liver tissue in patients with hemosiderosis: diagnosis with combination of R2*relaxometry and proton density fat fraction calculation by MRI
    Karcaaltincaba, Musturay
    Idilman, Ilkay
    Celik, Azim
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2011, 17 (04) : 323 - 327
  • [26] Assessment of cardiac iron by MRI susceptometry and R2*in patients with thalassemia
    Wang, Zhiyue J.
    Fischer, Roland
    Chu, Zili
    Mahoney, Donald H., Jr.
    Mueller, Brigitta U.
    Muthupillai, Raja
    James, Ellen B.
    Krishnamurthy, Rajesh
    Chung, Taylor
    Padua, Eric
    Vichinsky, Elliott
    Harmatz, Paul
    MAGNETIC RESONANCE IMAGING, 2010, 28 (03) : 363 - 371
  • [27] Comparison of clinical MRI liver iron content measurements using signal intensity ratios, R 2 and R 2*
    Runge, Jurgen H.
    Akkerman, Erik M.
    Troelstra, Marian A.
    Nederveen, Aart J.
    Beuers, Ulrich
    Stoker, Jaap
    ABDOMINAL RADIOLOGY, 2016, 41 (11) : 2123 - 2131
  • [28] R2*Relaxometry for the Quantification of Hepatic Iron Overload: Biopsy-Based Calibration and Comparison with the Literature
    Henninger, B.
    Zoller, H.
    Rauch, S.
    Finkenstedt, A.
    Schocke, M.
    Jaschke, W.
    Kremser, C.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2015, 187 (06): : 472 - 479
  • [29] Liver Iron Content Determination Using a Volumetric Breath-Hold Gradient-Echo Sequence With In-Line R2* Calculation
    Wunderlich, Arthur P.
    Schmidt, Stefan A.
    Mauro, Valeria
    Kneller, Lena
    Kannengiesser, Stephan
    Beer, Meinrad
    Cario, Holger
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2020, 52 (05) : 1550 - 1556
  • [30] Effect of particle size on liver MRI R2* relaxometry: Monte Carlo simulation and phantom studies
    Li, Xiaoben
    Wang, Changqing
    Huang, Jinhong
    Reeder, Scott B.
    Hernando, Diego