Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance

被引:13
作者
Jin, Hosang [1 ]
Keeling, Vance P. [1 ]
Johnson, Daniel A. [1 ]
Ahmad, Salahuddin [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Dept Radiat Oncol, Oklahoma City, OK 73104 USA
关键词
MapCHECK; 2; angular dependence; RapidArc QA; calibration field size; VMAT QA; MODULATED ARC THERAPY; STEP-AND-SHOOT; QUANTITATIVE-EVALUATION; RADIATION-THERAPY; DIODE-ARRAY; IMRT; TOMOTHERAPY; VERIFICATIONS; SOFTWARE; DEVICE;
D O I
10.1120/jacmp.v15i3.4638
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to investigate an effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc QA for 6, 8, 10, and 15 MV. The angular dependence was investigated by comparing MapCHECK 2 measurements in MapPHAN-MC2 to the corresponding Eclipse calculations every 10 using 10 x 10 cm(2) and 3 x 3 cm(2) fields. Fourteen patients were selected to make RapidArc plans using the four energies, and verification plans were delivered to two phantom setups: MapCHECK 2/MapPHAN phantom (MapPHAN QA) and MapCHECK 2 on an isocentric mounting fixture (IMF QA). Migration of MapCHECK 2 on IMF was simulated by splitting arcs every 10 degrees and displacing an isocenter of each partial arc in the Eclipse system (IMFACTUAL QA). To investigate the effect of calibration field size, MapCHECK 2 was calibrated by two field sizes (10 x 10 cm(2) and 3 x 3 cm(2)) and applied to all QA measurements. The. test was implemented using criteria of 1%/1 mm, 2%/2 mm, and 3%/3 mm. A mean dose of all compared points for each plan was compared with respect to a mean effective field size of the RapidArc plan. The angular dependence was considerably high at gantry angles of 90 degrees +/- 10 degrees and 270 degrees +/- 10 degrees (for 10 x 10/3 x 3 cm(2) at 90 degrees, 30.6% +/- 6.6%/33.4% +/- 5.8% (6 MV), 17.3% +/- 5.3%/15.0% +/- 6.8% (8 MV), 8.9% +/- 2.9%/7.8% +/- 3.2% (10 MV), and 2.2% +/- 2.3%/-1.3% +/- 2.6% (15 MV)). For 6 MV, the angular dependence significantly deteriorated the gamma passing rate for plans of large field size in MapPHAN QA (< 90% using 3%/3 mm); however, these plans passed the. test in IMFACTUAL QA (> 95%). The different calibration field sizes did not make any significant dose difference for both MapPHAN QA and IMFACTUAL QA. For 8, 10, and 15 MV, the angular dependence does not make any clinically meaningful impact on MapPHAN QA. Both MapPHAN QA and IMFACTUAL QA presented clinically acceptable. passing rates using 3%/3 mm. MapPHAN QA showed better passing rates than IMFACTUAL QA for the tighter criteria. The 10 x 10 cm(2) calibration showed better agreement for plans of small effective field size (< 5 x 5 cm(2)) in MapPHAN QA. There was no statistical difference between IMF QA and IMFACTUAL QA. In conclusion, MapPHAN QA is not recommended for plans of large field size, especially for 6 MV, and MapCHECK 2 should be calibrated using a field size similar to a mean effective field size of a RapidArc plan for better agreement for IMF QA.
引用
收藏
页码:80 / 92
页数:13
相关论文
共 30 条
[1]   Consistency and reproducibility of the VMAT plan delivery using three independent validation methods [J].
Chandraraj, Varatharaj ;
Stathakis, Sotirios ;
Manickam, Ravikumar ;
Esquivel, Carlos ;
Supe, Sanjay S. ;
Papanikolaou, Nikos .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (01) :129-140
[2]   Volumetric-modulated arc radiotherapy for carcinomas of the anal canal: A treatment planning comparison with fixed field IMRT [J].
Clivio, Alessandro ;
Fogliata, Antonella ;
Franzetti-Pellanda, Alessandra ;
Nicolini, Giorgia ;
Vanetti, Eugenio ;
Wyttenbach, Rolf ;
Cozzi, Luca .
RADIOTHERAPY AND ONCOLOGY, 2009, 92 (01) :118-124
[3]   ASSESSING THE ROLE OF VOLUMETRIC MODULATED ARC THERAPY (VMAT) RELATIVE TO IMRT AND HELICAL TOMOTHERAPY IN THE MANAGEMENT OF LOCALIZED, LOCALLY ADVANCED, AND POST-OPERATIVE PROSTATE CANCER [J].
Davidson, Melanie T. M. ;
Blake, Samuel J. ;
Batchelar, Deidre L. ;
Cheung, Patrick ;
Mah, Katherine .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (05) :1550-1558
[4]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[5]   RapidArc quality assurance through MapCHECK [J].
Gloi, Aime M. ;
Buchanan, Robert E. ;
Zuge, Corrie L. ;
Goettler, AnnDrea M. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :39-47
[6]   Is a single arc sufficient in volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes? [J].
Guckenberger, Matthias ;
Richter, Anne ;
Krieger, Thomas ;
Wilbert, Juergen ;
Baier, Kurt ;
Flentje, Michael .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) :259-265
[7]   Clinical practice and evaluation of electronic portal imaging device for VMAT quality assurance [J].
Huang, Yen-Cho ;
Yeh, Chien-Yi ;
Yeh, Jih-Hsiang ;
Lo, Ching-Jung ;
Tsai, Ping-Fang ;
Hung, Chih-Hung ;
Tsai, Chieh-Sheng ;
Chen, Chen-Yuan .
MEDICAL DOSIMETRY, 2013, 38 (01) :35-41
[8]   Quality assurance methodology for Varian RapidArc treatment plans [J].
Iftimia, Ileana ;
Cirino, Eileen T. ;
Xiong, Li ;
Mower, Herbert W. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (04) :130-143
[9]   A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery [J].
Jursinic, PA ;
Nelms, BE .
MEDICAL PHYSICS, 2003, 30 (05) :870-879
[10]   MapCHECK used for rotational IMRT measurements: Step-and-shoot, Tomotherapy, RapidArc [J].
Jursinic, Paul A. ;
Sharma, Renu ;
Reuter, Jim .
MEDICAL PHYSICS, 2010, 37 (06) :2837-2846