Effect of tube current on diagnostic image quality in paediatric cerebral multidetector CT images

被引:16
作者
Ledenius, K. [1 ]
Gustavsson, M. [2 ]
Johansson, S. [2 ]
Stalhammar, F. [2 ]
Wiklund, L-M [2 ]
Thilander-Klang, A. [1 ,3 ]
机构
[1] Gothenburg Univ, Dept Radiat Phys, Sahlgrenska Acad, Sahlgrenska Univ Hosp, SE-41345 Gothenburg, Sweden
[2] Queen Silvia Childrens Hosp, Dept Paediat Radiol & Physiol, SE-41685 Gothenburg, Sweden
[3] Sahlgrens Univ Hosp, Dept Biomed Engn & Med Phys, SE-41345 Gothenburg, Sweden
关键词
RADIATION-DOSE REDUCTION; COMPUTED-TOMOGRAPHY; RATING-SCALES; NOISE; BRAIN;
D O I
10.1259/bjr/24404354
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to investigate the effect of tube current on diagnostic image quality in paediatric cerebral multidetector CT (MDCT) images in order to identify the minimum radiation dose required to reproduce acceptable levels of different diagnostic image qualities. Original digital scanning data (raw data) were selected retrospectively from routine MDCT brain examinations of 25 paediatric patients. All examinations had been performed using axial scanning on an eight-slice MDCT (LightSpeed Ultra, GE Healthcare). Their ages ranged from newborn to 15 years. Quantum noise was added artificially to the raw data representing dose reductions equivalent to steps of 20 mA. Patient identification information was removed. Three experienced radiologists blindly and randomly assessed the resulting images from two different levels of the brain with regard to reproduction of structures and overall image quality. Final data were evaluated using the non-parametric statistical approach of inter-scale concordance. The minimum value of tube current-time product (mAs) required to reproduce an image of sufficient diagnostic quality was established in relation to the age of the patient. The corresponding CT dose index values by volume (CTDI(vol) (mGy)) were also established. In conclusion, acceptable reproduction of low-contrast structures was possible at CTDIvol values down to 20 mGy (patients 1-5 years old). For acceptable reproduction of high-contrast structures, CTDI(vol) values down to 10 mGy were considered possible (patients 1-5 years old). The original image quality for patients under 6 months of age (15 mGy) was found to be inadequate for acceptable reproduction of low-contrast structures.
引用
收藏
页码:313 / 320
页数:8
相关论文
共 28 条
[1]   Dose reduction in pediatric CT: A rational approach [J].
Boone, JM ;
Geraghty, EM ;
Seibert, JA ;
Wootton-Gorges, SL .
RADIOLOGY, 2003, 228 (02) :352-360
[2]   A software tool for increased efficiency in observer performance studies in radiology [J].
Börjesson, S ;
Håkansson, M ;
Båth, M ;
Kheddache, S ;
Svensson, S ;
Tingberg, A ;
Grahn, A ;
Ruschin, M ;
Hemdal, B ;
Mattsson, S ;
Månsson, LG .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :45-52
[3]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[4]   Addition of computer simulated noise to investigate radiation dose and image quality in images with spatial correlation of statistical noise: an example application to X-ray CT of the brain [J].
Britten, AJ ;
Crotty, M ;
Kiremidjian, H ;
Grundy, A ;
Adam, EJ .
BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (916) :323-328
[5]   Radiation dose reduction in paediatric cranial CT [J].
Chan, CY ;
Wong, YC ;
Chau, LF ;
Yu, SK ;
Lau, PC .
PEDIATRIC RADIOLOGY, 1999, 29 (10) :770-775
[6]  
DYBKAER R, 1989, SCAND J CLIN LAB I S, V194, P69
[7]  
European Commission, 1999, 16262 EUR EN EUR COM
[8]   Computer-simulated radiation dose reduction for abdominal multidetector CT of pediatric patients [J].
Frush, DP ;
Slack, CC ;
Hollingsworth, CL ;
Bisset, GS ;
Donnelly, LF ;
Hsieh, J ;
Lavin-Wensell, T ;
Mayo, JR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (05) :1107-1113
[9]  
HAKANSSON M, 2007, P SPIE, V6509
[10]   Effect of low doses of ionising radiation in infancy on cognitive function in adulthood: Swedish population based cohort study [J].
Hall, P ;
Adami, HO ;
Trichopoulos, D ;
Pedersen, NL ;
Lagiou, P ;
Ekbom, A ;
Ingvar, M ;
Lundell, M ;
Granath, F .
BMJ-BRITISH MEDICAL JOURNAL, 2004, 328 (7430) :19-21