Quantification of granisetron in human plasma by liquid chromatography coupled to electrospray tandem mass spectrometry

被引:13
作者
Nirogi, Ramakrishna V. S. [1 ]
Kandikere, Vishwottam N. [1 ]
Shukla, Manoj [1 ]
Mudigonda, Koteshwara [1 ]
Maurya, Santosh [1 ]
Boosi, Ravikumar [1 ]
机构
[1] Suven Life Sci Ltd, Biopharmaceut Res, Hyderabad 500034, Andhra Pradesh, India
关键词
granisetron; electrospray ionization; liquid chromatography-tandem mass spectrometry; pharmacokinetic study;
D O I
10.1002/bmc.613
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple, sensitive and rapid high-performance liquid chromatography/electrospray ionization tandem mass spectrometry method was developed and validated for the assay of granisetron in human plasma. Following liquid-liquid extraction, the analytes were separated using an isocratic mobile phase on a reversed-phase C-18 column and analyzed by MS in the multiple reaction monitoring mode using the respective [M+H](+) ions, m/z 313/138 for granisetron and m/z 409/228 for the IS. The assay exhibited a linear dynamic range of 0.1-20 ng/mL for granisetron in human plasma. The lower limit of quantification was 100 pg/mL with a relative standard deviation of less than 5%. Acceptable precision and accuracy were obtained for concentrations over the standard curve range. A run time of 2.0 min for each sample made it possible to analyze more than 400 human plasma samples per day. The validated method has been successfully used to analyze human plasma samples for application in pharmacokinetic, bioavailability or bioequivalence studies. Copyright (c) 2006 John Wiley & Sons, Ltd.
引用
收藏
页码:888 / 897
页数:10
相关论文
共 30 条
[11]  
Jemal M, 2000, BIOMED CHROMATOGR, V14, P422, DOI 10.1002/1099-0801(200010)14:6<422::AID-BMC25>3.0.CO
[12]  
2-I
[13]   SIMULTANEOUS DETERMINATION OF GRANISETRON AND 7-HYDROXYGRANISETRON IN HUMAN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE DETECTION [J].
KUDOH, S ;
SATO, T ;
OKADA, H ;
KUMAKURA, H ;
NAKAMURA, H .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 660 (01) :205-210
[14]   Pharmacokinetics, tolerability and efficacy of bolus-dose vs continuous-infusion granisetron in the prevention of vomiting in patients undergoing haematopoietic stem-cell transplantation -: A randomised, double-blind evaluation [J].
Lazarus, HM ;
O'Riordan, MA ;
Creger, RJ ;
Myers, CM ;
Blumer, JL ;
Reed, MD .
CLINICAL DRUG INVESTIGATION, 1999, 17 (02) :155-165
[15]   Progress in liquid chromatography-mass spectrometry instrumentation and its impact on high-throughput screening [J].
Niessen, WMA .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1000 (1-2) :413-436
[16]   Sensitive and selective liquid chromatography-tandem mass spectrometry method for the quantification of azithromycin in human plasma [J].
Nirogi, RVS ;
Kandikere, VN ;
Shukla, M ;
Mudigonda, K ;
Maurya, S ;
Boosi, R ;
Yerramilli, A .
ANALYTICA CHIMICA ACTA, 2005, 553 (1-2) :1-8
[17]   High-performance liquid chromatographic determination of granisetron in human plasma [J].
Pinguet, F ;
Bressolle, F ;
Martel, P ;
Salabert, D ;
Astre, C .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 675 (01) :99-105
[18]   GRANISETRON - A REVIEW OF ITS PHARMACOLOGICAL PROPERTIES AND THERAPEUTIC USE AS AN ANTIEMETIC [J].
PLOSKER, GL ;
GOA, KL .
DRUGS, 1991, 42 (05) :805-824
[19]   Liquid chromatography-negative ion electrospray tandem mass spectrometry method for the quantification of tacrolimus in human plasma and its bioanalytical applications [J].
Ramakrishna, NVS ;
Vishwottam, KN ;
Puran, S ;
Manoj, S ;
Santosh, M ;
Wishu, S ;
Koteshwara, M ;
Chidambara, J ;
Gopinadh, B ;
Sumatha, B .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 805 (01) :13-20
[20]   Rapid, simple and highly sensitive LC-ESI-MS/MS method for the quantification of tamsulosin in human plasma [J].
Ramakrishna, NVS ;
Vishwottam, KN ;
Manoj, S ;
Koteshwara, M ;
Wishu, S ;
Varma, DP .
BIOMEDICAL CHROMATOGRAPHY, 2005, 19 (10) :709-719