Validation of a Low Dose Simulation Technique for Computed Tomography Images

被引:28
作者
Muenzel, Daniela [1 ]
Koehler, Thomas [2 ]
Brown, Kevin [3 ]
Zabic, Stanislav [3 ]
Fingerle, Alexander A. [1 ]
Waldt, Simone [1 ]
Bendik, Edgar [1 ]
Zahel, Tina [1 ]
Schneider, Armin [4 ]
Dobritz, Martin [1 ]
Rummeny, Ernst J. [1 ]
Noel, Peter B. [1 ]
机构
[1] Tech Univ Munich, Dept Radiol, D-80290 Munich, Germany
[2] Philips Technol GmbH, Innovat Technol, Hamburg, Germany
[3] Philips Healthcare, Cleveland, OH USA
[4] Tech Univ Munich, MITI, Intervent Res Grp, D-80290 Munich, Germany
关键词
ITERATIVE RECONSTRUCTION ALGORITHM; TUBE CURRENT MODULATION; RADIATION-EXPOSURE; CT; REDUCTION; QUALITY; RISKS; NOISE; MANAGEMENT; IMPACT;
D O I
10.1371/journal.pone.0107843
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: Evaluation of a new software tool for generation of simulated low-dose computed tomography (CT) images from an original higher dose scan. Materials and Methods: Original CT scan data (100 mAs, 80 mAs, 60 mAs, 40 mAs, 20 mAs, 10 mAs; 100 kV) of a swine were acquired (approved by the regional governmental commission for animal protection). Simulations of CT acquisition with a lower dose (simulated 10-80 mAs) were calculated using a low-dose simulation algorithm. The simulations were compared to the originals of the same dose level with regard to density values and image noise. Four radiologists assessed the realistic visual appearance of the simulated images. Results: Image characteristics of simulated low dose scans were similar to the originals. Mean overall discrepancy of image noise and CT values was -1.2% (range -9% to 3.2%) and -0.2% (range -8.2% to 3.2%), respectively, p>0.05. Confidence intervals of discrepancies ranged between 0.9-10.2 HU (noise) and 1.9-13.4 HU (CT values), without significant differences (p>0.05). Subjective observer evaluation of image appearance showed no visually detectable difference. Conclusion: Simulated low dose images showed excellent agreement with the originals concerning image noise, CT density values, and subjective assessment of the visual appearance of the simulated images. An authentic low-dose simulation opens up opportunity with regard to staff education, protocol optimization and introduction of new techniques.
引用
收藏
页数:8
相关论文
共 29 条
[1]  
Amis E Stephen Jr, 2007, J Am Coll Radiol, V4, P272, DOI 10.1016/j.jacr.2007.03.002
[2]   Radiation Exposure and Image Quality of Normal Computed Tomography Brain Images Acquired With Automated and Organ-Based Tube Current Modulation Multiband Filtering and Iterative Reconstruction [J].
Becker, Hans-Christoph ;
Augart, Dominik ;
Karpitschka, Martina ;
Ulzheimer, Stefan ;
Bamberg, Fabian ;
Morhard, Dominik ;
Neumaier, Klement ;
Graser, Anno ;
Johnson, Thorsten ;
Reiser, Maximilian .
INVESTIGATIVE RADIOLOGY, 2012, 47 (03) :202-207
[3]   Iterative reconstruction methods in X-ray CT [J].
Beister, Marcel ;
Kolditz, Daniel ;
Kalender, Willi A. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :94-108
[4]   Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know [J].
Brenner, DJ ;
Doll, R ;
Goodhead, DT ;
Hall, EJ ;
Land, CE ;
Little, JB ;
Lubin, JH ;
Preston, DL ;
Preston, RJ ;
Puskin, JS ;
Ron, E ;
Sachs, RK ;
Samet, JM ;
Setlow, RB ;
Zaider, M .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2003, 100 (24) :13761-13766
[5]   Addition of computer simulated noise to investigate radiation dose and image quality in images with spatial correlation of statistical noise: an example application to X-ray CT of the brain [J].
Britten, AJ ;
Crotty, M ;
Kiremidjian, H ;
Grundy, A ;
Adam, EJ .
BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (916) :323-328
[6]   Filtered Back Projection, Adaptive Statistical Iterative Reconstruction, and a Model-based Iterative Reconstruction in Abdominal CT: An Experimental Clinical Study [J].
Deak, Zsuzsanna ;
Grimm, Jochen M. ;
Treitl, Marcus ;
Geyer, Lucas L. ;
Linsenmaier, Ulrich ;
Koerner, Markus ;
Reiser, Maximilian F. ;
Wirth, Stefan .
RADIOLOGY, 2013, 266 (01) :197-206
[7]   Computer-simulated radiation dose reduction for abdominal multidetector CT of pediatric patients [J].
Frush, DP ;
Slack, CC ;
Hollingsworth, CL ;
Bisset, GS ;
Donnelly, LF ;
Hsieh, J ;
Lavin-Wensell, T ;
Mayo, JR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (05) :1107-1113
[8]   Dose reduction in CT by anatomically adapted tube current modulation. 1. Simulation studies [J].
Gies, M ;
Kalender, WA ;
Wolf, H ;
Suess, C ;
Madsen, MT .
MEDICAL PHYSICS, 1999, 26 (11) :2235-2247
[9]   Impact of iterative reconstruction on CNR and SNR in dynamic myocardial perfusion imaging in an animal model [J].
Gramer, B. M. ;
Muenzel, D. ;
Leber, V. ;
von Thaden, A. -K. ;
Feussner, H. ;
Schneider, A. ;
Vembar, M. ;
Soni, N. ;
Rummeny, E. J. ;
Huber, A. M. .
EUROPEAN RADIOLOGY, 2012, 22 (12) :2654-2661
[10]   Assessment of an iterative reconstruction algorithm (SAFIRE) on image quality in pediatric cardiac CT datasets [J].
Han, B. Kelly ;
Grant, Katharine L. R. ;
Garberich, Ross ;
Sedlmair, Martin ;
Lindberg, Jana ;
Lesser, John R. .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2012, 6 (03) :200-204