The impact of body position on intra-abdominal pressure measurement: A multicenter analysis

被引:96
|
作者
Cheatham, Michael L. [1 ]
De Waele, Jan J. [2 ]
De Laet, Inneke [3 ]
De Keulenaer, Bart [4 ]
Widder, Sandy [5 ]
Kirkpatrick, Andrew W. [6 ,7 ]
Cresswell, Adrian B. [8 ,9 ]
Malbrain, Manu [10 ]
Bodnar, Zsolt [11 ]
Mejia-Mantilla, Jorge H. [12 ]
Reis, Richard [13 ]
Parr, Michael [14 ]
Schulze, Robert [15 ]
Puig, Sonia [16 ]
机构
[1] Orlando Reg Med Ctr Inc, Dept Surg Educ, Orlando, FL USA
[2] Ghent Univ Hosp, Surg Intens Care Unit, B-9000 Ghent, Belgium
[3] Ziekenhuis Netwerk Antwerpen, Intens Care Unit, Antwerp, Belgium
[4] Fremantle Hosp, Intens Care Unit, Fremantle, WA, Australia
[5] Univ Calgary, Dept Surg, Calgary, AB, Canada
[6] Foothills Med Ctr, Dept Crit Care Med, Calgary, AB, Canada
[7] Foothills Med Ctr, Dept Surg, Calgary, AB, Canada
[8] Kings Coll Hosp London, Liver Transplant Surg Serv, London, England
[9] Kings Coll Hosp London, Liver ITU Inst Liver Studies, London, England
[10] Ziekenhuis Netwerk Antwerpen, Intens Care Unit, Antwerp, Belgium
[11] Kenezy Teaching Hosp, Dept Surg, Debrecen, Hungary
[12] Univ Valle, Cali, Colombia
[13] Bratislava Univ, Dept Surg, Bratislava, Slovakia
[14] Liverpool Hosp, Intens Care Unit, Sydney, NSW, Australia
[15] SUNY Downstate Kings Cty Hosp, Dept Surg, Brooklyn, NY USA
[16] Hosp Mar, Barcelona, Spain
关键词
intra-abdominal pressure; intra-abdominal hypertension; abdominal compartment syndrome; intravesicular pressure; intensive care; monitoring; ABDOMINAL COMPARTMENT SYNDROME; INTERNATIONAL-CONFERENCE; CLINICAL EXAMINATION; HYPERTENSION; EXPERTS;
D O I
10.1097/CCM.0b013e3181a021fa
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Elevated intra-abdominal pressure (IAP) is a frequent cause of morbidity and mortality among the critically ill. IAP is most commonly measured using the intravesicular or "bladder" technique. The impact of changes in body position on the accuracy of IAP measurements, such as head of bed elevation to reduce the risk of ventilator-associated pneumonia, remains unclear. Design: Prospective, cohort study. Setting: Twelve international intensive care units. Patients: One hundred thirty-two critically ill medical and surgical patients at risk for intra-abdominal hypertension and abdominal compartment syndrome. Interventions: Triplicate intravesicular pressure measurements were performed at least 4 hours apart with the patient in the supine, 15 degrees, and 30 degrees head of bed elevated positions. The zero reference point was the mid-axillary line at the iliac crest. Measurements and Main Results. Mean IAP values at each head of bed position were significantly different (p < 0.0001). The bias between IAP(supine) and IAP(15)degrees was 1.5 mm Hg (1.3-1.7). The bias between IAP(supine) and IAP(30)degrees was 3.7 mm Hg (3.4-4.0). Conclusions: Head of bed elevation results in clinically significant increases in measured IAP. Consistent body positioning from one IAP measurement to the next is necessary to allow consistent trending of IAP for accurate clinical decision making. Studies that involve IAP measurements should describe the patient's body position so that these values may be properly interpreted. (Grit Care Med 2009; 37:2187-2190)
引用
收藏
页码:2187 / 2190
页数:4
相关论文
共 50 条
  • [31] Estimation of intra-abdominal pressure by bladder pressure measurement: Validity and methodology
    Fusco, MA
    Martin, RS
    Chang, MC
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2001, 50 (02): : 297 - 302
  • [32] Pre-Clinical Validation of A Novel Continuous Intra-Abdominal Pressure Measurement Equipment (SERENNO)
    Tayebi, Salar
    Wise, Robert
    Pourkazemi, Ali
    Stiens, Johan
    Malbrain, Manu L. N. G.
    LIFE-BASEL, 2022, 12 (08):
  • [33] INFLUENCE OF INTRA-ABDOMINAL PRESSURE ON THE BASIC VITAL FUNCTIONS AND FINAL TREATMENT OUTCOME
    Svorcan, Petar
    Stojanovic, Maja
    Stevanovic, Predrag
    Karamarkovic, Aleksandar
    Jankovic, Radmilo
    Ladjevic, Nebojsa
    ACTA CLINICA CROATICA, 2016, 55 (02) : 316 - 322
  • [34] Intra-abdominal Pressure Measurement - Step by Step
    Kuhn, Anja
    Fuchs, Christian
    Hahnenkamp, Klaus
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2021, 146 (18) : 1211 - 1217
  • [35] Harrahill's technique: a simple screening test for intra-abdominal pressure measurement
    Otto, J.
    Binneboesel, M.
    Junge, K.
    Jansen, M.
    Dembinski, R.
    Schumpelick, V.
    Schachtrupp, A.
    HERNIA, 2010, 14 (04) : 415 - 419
  • [36] Comparison of a new device for the continuous intra-gastric measurement of intra-abdominal pressure (CiMon) with direct intra-peritoneal measurements in cirrhotic patients during paracentesis
    Becker, Valentin
    Schmid, Roland M.
    Umgelter, Andreas
    INTENSIVE CARE MEDICINE, 2009, 35 (05) : 948 - 952
  • [37] INTRA-ABDOMINAL PRESSURE MEASUREMENTS IN LATERAL DECUBITUS VERSUS SUPINE POSITION
    De Keulenaer, B. L.
    Cheatham, M. L.
    De Waele, J. J.
    Kimball, E. J.
    Powell, B.
    Davis, W. A.
    Jenkins, I. R.
    ACTA CLINICA BELGICA, 2009, 64 (03) : 210 - 215
  • [38] Intra-abdominal Measurement Techniques: Is There Anything New?
    De Keulenaer, Bart L.
    Regli, Adrian
    Malbrain, Manu L. N. G.
    AMERICAN SURGEON, 2011, 77 (07) : S17 - S22
  • [39] Utility of monitoring intra-abdominal pressure in critically ill children
    Gonzalez, L.
    Rodriguez, R.
    Mencia, S.
    Gil-Ruiz, M.
    Sanavia, E.
    Lopez Herce, J.
    ANALES DE PEDIATRIA, 2012, 77 (04): : 254 - 260
  • [40] Intra-abdominal Pressure Screening of Horses With Colic
    de Paula, Vanessa B.
    Canola, Paulo A.
    Rivera, Gabriela G.
    Bonacin, Yuri S.
    Del Rio, Lara A.
    Canola, Julio C.
    Ferraudo, Antonio S.
    JOURNAL OF EQUINE VETERINARY SCIENCE, 2020, 90