Image Information Obtained Using a Charge-Coupled Device (CCD) Camera During an Immersion Liquid Evaporation Process for Measuring the Refractive Index of Solid Particles

被引:1
|
作者
Niskanen, Ilpo [1 ,2 ]
Sutinen, Veijo [3 ]
Thungstrom, Goeran [2 ]
Raty, Jukka [3 ]
机构
[1] Univ Oulu, Fac Technol, Struct & Construct Technol, Oulu, Finland
[2] Mid Sweden Univ, Dept Elect Design, Sundsvall, Sweden
[3] Univ Oulu, Unit Measurement Technol MITY, Kajaani, Finland
关键词
Refractive index; charge-coupled device camera; CCD camera; liquid mixture; evaporation; image immersion liquid method; calcium fluoride; CaF2; barium fluoride; BaF2;
D O I
10.1177/0003702818756660
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The refractive index is a fundamental physical property of a medium, which can be used for the identification and purity issues of all media. Here we describe a refractive index measurement technique to determine simultaneously the refractive index of different solid particles by monitoring the transmittance of light from a suspension using a charge-coupled device (CCD) camera. An important feature of the measurement is the liquid evaporation process for the refractive index matching of the solid particle and the immersion liquid; this was realized by using a pair of volatile and non-volatile immersion liquids. In this study, refractive indices of calcium fluoride (CaF2) and barium fluoride (BaF2) were determined using the proposed method.