Cadaveric Testing of Robot-Assisted Access to the Internal Auditory Canal for Vestibular Schwannoma Removal

被引:20
作者
Dillon, Neal P. [1 ]
Balachandran, Ramya [2 ]
Siebold, Michael A. [3 ]
Webster, Robert J., III [1 ]
Wanna, George B. [2 ]
Labadie, Robert F. [2 ]
机构
[1] Vanderbilt Univ, Med Ctr, Mech Engn, PMB 351592,2301 Vanderbilt Pl, Nashville, TN 37235 USA
[2] Vanderbilt Univ, Med Ctr, Otolaryngol, PMB 351592,2301 Vanderbilt Pl, Nashville, TN 37235 USA
[3] Vanderbilt Univ, Elect Engn, PMB 351592,2301 Vanderbilt Pl, Nashville, TN 37235 USA
基金
美国国家卫生研究院;
关键词
Hypothesis: An image-guided robotic system can safely; perform the bulk removal of bone during the translabyrinthine; approach to vestibular schwannoma (VS); Background: The translabyrinthine approach to VS; removal involves extensive manual milling in the temporal; bone to gain access to the internal auditory canal (IAC); for tumor resection. This bone removal is time consuming; and challenging due to the presence of vital anatomy (e.g; facial nerve) embedded within the temporal bone. A; robotic system can use preoperative imaging and segmentations; to guide a surgical drill to remove a prescribed; volume of bone; thereby preserving the surgeon for the; more delicate work of opening the IAC and resecting the; tumor; Methods: Fresh human cadaver heads were used in the; experiments. For each trial; the desired bone resection; volume was planned on a preoperative computed tomography; (CT) image; the steps in the proposed clinical workflow were undertaken; and the robot was programmed to mill the; specified volume. A postoperative CT scan was acquired for; evaluation of the accuracy of the milled cavity and; examination of vital anatomy; Results: In all experimental trials; the facial nerve and; chorda tympani were preserved. The root mean squared; surface accuracy of the milled cavities ranged from 0.23 to; 0.65 mm and the milling time ranged from 32.7 to; 57.0; minute; Conclusion: This work shows feasibility of using a robotassisted; approach for VS removal surgery. Further testing and; system improvements are necessary to enable clinical; translation of this technology; IN-VITRO; ACCURACY EVALUATION; COCHLEAR ACCESS; VALIDATION; SURGERY; IMAGES; SYSTEM; FRAMES; CT;
D O I
10.1097/MAO.0000000000001324
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Hypothesis: An image-guided robotic system can safely perform the bulk removal of bone during the translabyrinthine approach to vestibular schwannoma (VS). Background: The translabyrinthine approach to VS removal involves extensive manual milling in the temporal bone to gain access to the internal auditory canal (IAC) for tumor resection. This bone removal is time consuming and challenging due to the presence of vital anatomy (e.g., facial nerve) embedded within the temporal bone. A robotic system can use preoperative imaging and segmentations to guide a surgical drill to remove a prescribed volume of bone, thereby preserving the surgeon for the more delicate work of opening the IAC and resecting the tumor. Methods: Fresh human cadaver heads were used in the experiments. For each trial, the desired bone resection volume was planned on a preoperative computed tomography (CT) image, the steps in the proposed clinical workflow were undertaken, and the robot was programmed to mill the specified volume. A postoperative CT scan was acquired for evaluation of the accuracy of the milled cavity and examination of vital anatomy. Results: In all experimental trials, the facial nerve and chorda tympani were preserved. The root mean squared surface accuracy of the milled cavities ranged from 0.23 to 0.65 mm and the milling time ranged from 32.7 to 57.0 minute. Conclusion: This work shows feasibility of using a robotassisted approach for VS removal surgery. Further testing and system improvements are necessary to enable clinical translation of this technology. © 2017 Otology & Neurotology, Inc.
引用
收藏
页码:441 / 447
页数:7
相关论文
共 25 条
[11]  
Hoon Lim, 2011, Proceedings of the 2011 IEEE International Conference on Mechatronics and Automation (ICMA 2011), P549, DOI 10.1109/ICMA.2011.5985720
[12]   Configuration optimization and experimental accuracy evaluation of a bone-attached, parallel robot for skull surgery [J].
Kobler, Jan-Philipp ;
Nuelle, Kathrin ;
Lexow, G. Jakob ;
Rau, Thomas S. ;
Majdani, Omid ;
Kahrs, Lueder A. ;
Kotlarski, Jens ;
Ortmaier, Tobias .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2016, 11 (03) :421-436
[13]   Design and analysis of a head-mounted parallel kinematic device for skull surgery [J].
Kobler, Jan-Philipp ;
Kotlarski, Jens ;
Oeltjen, Julian ;
Baron, Stephan ;
Ortmaier, Tobias .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2012, 7 (01) :137-149
[14]   Design of a Bone-Attached Parallel Robot for Percutaneous Cochlear Implantation [J].
Kratchman, Louis B. ;
Blachon, Gregoire S. ;
Withrow, Thomas J. ;
Balachandran, Ramya ;
Labadie, Robert F. ;
Webster, Robert J., III .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (10) :2904-2910
[15]   Minimally Invasive Image-Guided Cochlear Implantation Surgery: First Report of Clinical Implementation [J].
Labadie, Robert F. ;
Balachandran, Ramya ;
Noble, Jack H. ;
Blachon, Gregoire S. ;
Mitchell, Jason E. ;
Reda, Fitsum A. ;
Dawant, Benoit M. ;
Fitzpatrick, J. Michael .
LARYNGOSCOPE, 2014, 124 (08) :1915-1922
[16]   Clinical Validation Study of Percutaneous Cochlear Access Using Patient-Customized Microstereotactic Frames [J].
Labadie, Robert F. ;
Balachandran, Ramya ;
Mitchell, Jason E. ;
Noble, Jack H. ;
Majdani, Omid ;
Haynes, David S. ;
Bennett, Marc L. ;
Dawant, Benoit M. ;
Fitzpatrick, J. Michael .
OTOLOGY & NEUROTOLOGY, 2010, 31 (01) :94-99
[17]   Customized, rapid-production microstereotactic table for surgical targeting: description of concept and in vitro validation [J].
Labadie, Robert F. ;
Mitchell, Jason ;
Balachandran, Ramya ;
Fitzpatrick, J. Michael .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2009, 4 (03) :273-280
[18]   Multimodality image registration by maximization of mutual information [J].
Maes, F ;
Collignon, A ;
Vandermeulen, D ;
Marchal, G ;
Suetens, P .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :187-198
[19]  
McBrayer KL, 2016, SPIE MED IMAGING 201
[20]   Automatic Identification and 3D Rendering of Temporal Bone Anatomy [J].
Noble, Jack H. ;
Dawant, Benoit M. ;
Warren, Frank M. ;
Labadie, Robert F. .
OTOLOGY & NEUROTOLOGY, 2009, 30 (04) :436-442