Design of a permanent magnet with a mechanical sweep suitable for variable-temperature continuous-wave and pulsed EPR spectroscopy

被引:22
作者
Bauer, C. [1 ]
Raich, H. [1 ]
Jeschke, G. [1 ]
Bluemler, P. [1 ]
机构
[1] Max Planck Inst Polymer Res, D-55021 Mainz, Germany
关键词
ESR; Electron spin resonance; Electron paramagnetic resonance; Copper; ESEEM; ENDOR; Nested Halbach dipoles; PORTABLE ESR SPECTROMETER; DISTANCE MEASUREMENTS; MINIATURE MAGNET; RESONANCE; NMR; BIOMACROMOLECULES; DOSIMETRY; SYSTEM; RANGE;
D O I
10.1016/j.jmr.2009.02.010
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A magnetic system is introduced which consists of three nested rings of permanent magnets of a Halbach dipolar layout and is capable for EPR spectroscopy. Two of the rings can be rotated independently to adjust the magnetic flux in the center and even allow for mechanical field sweeps. The presented prototype achieves a magnetic flux range of 0.0282-0.3013 T with a minimal sweep of 0.15 mT and homogeneity of about 10(-3). First applications with CW and pulsed Mims ENDOR as well as ESEEM experiments on a sample of a glycine single crystal doped with 1% copper nitrate demonstrate that flux range, sweep accuracy and homogeneity of this prototype is sufficient for EPR experiments on most solid samples. Together with a recently improved design magnets can be build which could serve as compact and easily transportable replacement of standard electromagnets with negligible consumption of power or coolants. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:222 / 227
页数:6
相关论文
共 23 条
[1]   Portable X-band system for solution state dynamic nuclear polarization [J].
Armstrong, Brandon D. ;
Lingwood, Mark D. ;
McCarney, Evan R. ;
Brown, Elliott R. ;
Bluemler, Peter ;
Han, Songi .
JOURNAL OF MAGNETIC RESONANCE, 2008, 191 (02) :273-281
[2]   Mobile sensor for high resolution NMR spectroscopy and imaging [J].
Danieli, Ernesto ;
Mauler, Joerg ;
Perlo, Juan ;
Bluemich, Bernhard ;
Casanova, Federico .
JOURNAL OF MAGNETIC RESONANCE, 2009, 198 (01) :80-87
[3]   LIGAND ENDOR SPECTRA OF CU(II) IMPURITY COMPLEXES IN ALPHA-GLYCINE CRYSTALS [J].
FUJIMOTO, M ;
MCDOWELL, CA ;
TAKUI, T .
JOURNAL OF CHEMICAL PHYSICS, 1979, 70 (08) :3694-3701
[4]   DESIGN OF PERMANENT MULTIPOLE MAGNETS WITH ORIENTED RARE-EARTH COBALT MATERIAL [J].
HALBACH, K .
NUCLEAR INSTRUMENTS & METHODS, 1980, 169 (01) :1-10
[5]   STRONG RARE-EARTH COBALT QUADRUPOLES [J].
HALBACH, K .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (03) :3882-3884
[6]  
IKEYA M, 1989, APPL RADIAT ISOTOPES, V40, P845, DOI 10.1016/0883-2889(89)90005-1
[7]   AN ELECTRON-SPIN-RESONANCE SYSTEM FOR INVIVO HUMAN TOOTH DOSIMETRY [J].
ISHII, H ;
IKEYA, M .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1990, 29 (05) :871-875
[8]  
Jeschke G, 2002, CHEMPHYSCHEM, V3, P927, DOI 10.1002/1439-7641(20021115)3:11<927::AID-CPHC927>3.0.CO
[9]  
2-Q
[10]   Distance measurements on spin-labelled biomacromolecules by pulsed electron paramagnetic resonance [J].
Jeschke, Gunnar ;
Polyhach, Yevhen .
PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 2007, 9 (16) :1895-1910