Automated nucleic acid isolation methods for HDV viral load quantification can lead to viral load underestimation

被引:19
作者
Bremer, Birgit [1 ]
Anastasiou, Olympia E. [2 ,3 ]
Ciesek, Sandra [3 ]
Weelemeyer, Heiner [1 ,2 ]
机构
[1] Hannover Med Sch, Dept Gastroenterol Hepatol & Endocrinol, Hannover, Germany
[2] Univ Hosp Essen, Dept Gastroenterol & Hepatol, Essen, Germany
[3] Univ Duisburg Essen, Univ Hosp Essen, Inst Virol, Essen, Germany
关键词
HEPATITIS-B-VIRUS; RNA QUANTIFICATION; EXTRACTION; THERAPY; DRUG; KIT;
D O I
10.3851/IMP3281
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Background: HDV infection is a cause of severe liver disease. Diagnosis and monitoring of HDV RNA are important to patient management. Since 2012, a WHO standard for HDV RNA quantification has been available; however, the impact of RNA extraction methods on HDV viral load quantification has never been evaluated. Methods: The aim of this study was to compare four commonly used automated nucleic acid (NA) extraction methods (AmpliPrep, MagNA Pure, QIAcube QBK and QIAcube VRK) with a manual RNA extraction method (Instant Virus RNA/DNA kit) and evaluate the possible effect of each method on HDV RNA yield with subsequent amplification with the Robogene HDV assay. Serum samples from HDV-positive patients taken before treatment with pegylated interferon-alpha 2a and at treatment weeks 12 and 48 were studied. Results: The automated extraction methods MagNA Pure, Ampliprep and QIAcube VRK extraction led to about 10-fold lower HDV RNA values compared with the manual method of NA extraction, while the difference was smaller with QIAcube QBK (about 6-fold lower). The median viral load was 10,665 IU/ml for the manual method, 445 IU/ml for AmpliPrep, 3,209 IUM for MagNA Pure, 2,060 IU/ml for QIAcube QBK and 3,568 IU/ml for QIAcube VRK. Use of MagNA Pure led to misclassification of two on-treatment samples with low viral load as being false negative. Conclusions: The NA extraction method had a significant impact on the measured HDV viral loads determined by the commonly used Robogene assay, with the manual RNA method yielding consistently higher values of viral load.
引用
收藏
页码:117 / 123
页数:7
相关论文
共 24 条
  • [1] Comparison of 3 nucleic acid isolation methods for the quantification of HIV-1 RNA by Cobas Taqman real-time polymerase chain reaction system
    Alp, Alpaslan
    Hascelik, Guelsen
    [J]. DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2009, 63 (04) : 365 - 371
  • [2] Alp aslan, 2004, Mikrobiyoloji Bulteni, V38, P77
  • [3] Safety and efficacy of REP 2139 and pegylated interferon alfa-2a for treatment-naive patients with chronic hepatitis B virus and hepatitis D virus co-infection (REP 301 and REP 301-LTF): a non-randomised, open-label, phase 2 trial
    Bazinet, Michel
    Pantea, Victor
    Cebotarescu, Valentin
    Cojuhari, Lilia
    Jimbei, Pavlina
    Albrecht, Jeffrey
    Schmid, Peter
    Le Gal, Frederic
    Gordien, Emmanuel
    Krawczyk, Adalbert
    Mijocevic, Hrvoje
    Karimzadeh, Hadi
    Roggendorf, Michael
    Vaillant, Andrew
    [J]. LANCET GASTROENTEROLOGY & HEPATOLOGY, 2017, 2 (12): : 877 - 889
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] Treatment of chronic hepatitis D with the entry inhibitor myrcludex B: First results of a phase Ib/IIa study
    Bogomolov, Pavel
    Alexandrov, Alexander
    Voronkova, Natalia
    Macievich, Maria
    Kokina, Ksenia
    Petrachenkova, Maria
    Lehr, Thorsten
    Lempp, Florian A.
    Wedemeyer, Heiner
    Haag, Mathias
    Schwab, Matthias
    Haefeli, Walter E.
    Blank, Antje
    Urban, Stephan
    [J]. JOURNAL OF HEPATOLOGY, 2016, 65 (03) : 490 - 498
  • [6] Commercial Real-Time Reverse Transcriptase PCR Assays Can Underestimate or Fail to Quantify Hepatitis Delta Virus Viremia
    Brichler, Segolene
    Le Gal, Frederic
    Butt, Afifaa
    Chevret, Sylvie
    Gordien, Emmanuel
    [J]. CLINICAL GASTROENTEROLOGY AND HEPATOLOGY, 2013, 11 (06) : 734 - 740
  • [7] Chudy M., 2013, WHO/BS/2013.2227
  • [8] Performance evaluation of the automated NucliSens easyMAG nucleic acid extraction platform in comparison with QIAamp Mini kit from clinical specimens
    Francesca, Perandin
    Patrizia, Pollara C.
    Franco, Gargiulo
    Carlo, Bonfanti
    Manca, Nino
    [J]. DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2009, 64 (02) : 158 - 165
  • [9] Hepatitis delta virus
    Hughes, Sarah A.
    Wedemeyer, Heiner
    Harrison, Phillip M.
    [J]. LANCET, 2011, 378 (9785) : 73 - 85
  • [10] Kang SH, 2012, ANN CLIN LAB SCI, V42, P370