Potentiometric Determination of La(III) Using Chitosan Modified Carbon Paste Electrode with An experimental Design

被引:19
|
作者
Rajabi, Negar [1 ]
Masrournia, Mahboubeh [1 ]
Abedi, Mohamadreza [2 ]
机构
[1] Islamic Azad Univ, Dept Chem, Mashhad Branch, Mashhad, Razavi Khorasan, Iran
[2] Islamic Azad Univ, Dept Appl Chem, Quchan Branch, Quchan, Iran
来源
CHEMICAL METHODOLOGIES | 2020年 / 4卷 / 05期
关键词
Lanthanum selective electrode; Multi-walled carbon nanotubes; Chitosan; Modified carbon paste electrode; MEMBRANE-ELECTRODE; VOLTAMMETRIC SENSOR; FOLIC-ACID; LIQUID; PHARMACEUTICALS; NANOSTRUCTURE; CROWN;
D O I
10.22034/chemm.2020.109975
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
In this work, D-optimal mixture design was applied as an experimental design to screen and optimize the effect of the composition of a carbon paste on its performance. In this study, chitosan was offered as a ligand for the determination of La(III) with the carbon paste electrode. In the fabricated sensor, the greatest nernstian slope was achieved on the optimized mixture composition by experimental design: 60.00% w/w powder graphite, 1.53% w/w multi-walled carbon nanotubes (MWCNTs), 29.79% w/w paraffin oil, and 8.68% w/w chitosan. MWCNT was used for increasing the electrode response to La(III). The quadratic fitting pattern based on D-optimal model was used to find the desirability functions of the suggested design to assess the cross-interferences and the interactions between the factors. The fabricated sensor for determining the La(III) ions demonstrated a maximum nernstian slope equal to 19.70 mV.decade-1 along a linear range from 1.0x10(-6) to 1.0x10(-2) mol/L-1 and detection limit of 10(-6) mol/L-1. The designed sensor was successfully tested in the pH range of 2-9 with suitable selectivity, fast response time (about 30 s) and long lifetime (over 2 months) was obtained without any deviation. The offered electrode was used successfully as an indicator in the potentiometric titration of La(III) with EDTA. Copyright (C) 2020 by SPC (Sami Publishing Company)
引用
收藏
页码:660 / 670
页数:11
相关论文
共 50 条
  • [21] Potentiometric determination of monohydrogen arsenate by zeolite-modified carbon-paste electrode
    Mazloum-Ardakani, Mohammad
    Karimi, M. A.
    Mashhadizadeh, M. H.
    Pesteh, M.
    Azimi, M. S.
    Kazemian, H.
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2007, 87 (04) : 285 - 294
  • [22] Potentiometric Determination of Oxymetazoline Hydrochloride Using Modified Carbon Paste Electrodes
    Issa, Yousry M.
    Khalil, Mohamed M.
    Zayed, Sayed I. M.
    Mustafa, Naglaa
    PORTUGALIAE ELECTROCHIMICA ACTA, 2021, 39 (05) : 381 - 392
  • [23] Cyclam modified carbon paste electrode as a potentiometric sensor for determination of cobalt(II) ions
    Pouretedal, HR
    Keshavarz, MH
    CHEMICAL RESEARCH IN CHINESE UNIVERSITIES, 2005, 21 (01) : 28 - 31
  • [24] DETERMINATION OF CHROMIUM(VI) AND CHROMIUM(III) BY USING A DIPHENYLCARBAZIDE-MODIFIED CARBON PASTE ELECTRODE
    PANIAGUA, AR
    VAZQUEZ, MD
    TASCON, ML
    BATANERO, PS
    ELECTROANALYSIS, 1993, 5 (02) : 155 - 163
  • [25] Carbon paste electrode modified with chromium thiopental for the potentiometric flow injection analysis of chromium (III)
    Youssef, A. F. A.
    Issa, Y. M.
    Mohamed, M. S.
    TOXICOLOGICAL AND ENVIRONMENTAL CHEMISTRY, 2012, 94 (02): : 220 - 238
  • [26] Potentiometric determination of nitrazepam in pharmaceutical dosage form and spiked plasma using a novel modified carbon paste electrode
    Azab, Marwa M.
    Rizk, Mohamed
    Mohamed, Omaima M.
    Farag, Amir Shaaban
    ELECTROANALYSIS, 2024, 36 (11)
  • [27] Determination of gold using clay modified carbon paste electrode
    Navrátilová, Z
    Kula, P
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 367 (04): : 369 - 372
  • [28] Determination of gold using clay modified carbon paste electrode
    Z. Navrátilová
    Petr Kula
    Fresenius' Journal of Analytical Chemistry, 2000, 367 : 369 - 372
  • [29] Catalytic determination of perchlorate using a modified carbon paste electrode
    Neuhold, CG
    Kalcher, K
    Cai, XH
    Raber, G
    ANALYTICAL LETTERS, 1996, 29 (10) : 1685 - 1704
  • [30] Voltammetric determination of bisphenol A using modified carbon paste electrode
    Jahani, Peyman Mohammadzadeh
    Beitollahi, Hadi
    Nejad, Fariba Garkani
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2024,