Real-time monitoring of spinal cord blood flow with a novel sensor mounted on a cerebrospinal fluid drainage catheter in an animal model

被引:5
|
作者
Hayatsu, Yukihiro [1 ]
Kawamoto, Shunsuke [1 ]
Matsunaga, Tadao [2 ]
Haga, Yoichi [3 ]
Saiki, Yoshikatsu [1 ]
机构
[1] Tohoku Univ, Dept Cardiovasc Surg, Grad Sch Med, Sendai, Miyagi 9808574, Japan
[2] Tohoku Univ, Micro Syst Integrat Ctr, Sendai, Miyagi 9808574, Japan
[3] Tohoku Univ, Grad Sch Biomed Engn, Sendai, Miyagi 9808574, Japan
来源
关键词
AORTIC-ANEURYSM REPAIR; REDUCES PARAPLEGIA; LIFE EXPECTANCY; THORACIC AORTA; PERFUSION; OPERATIONS; SURGERY; INJURY; PAPAVERINE; PROTECTION;
D O I
10.1016/j.jtcvs.2014.04.011
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The aim of our study was to develop a novel monitoring system for spinal cord blood flow (SCBF) to test the efficacy of the SCBF sensor in an animal model. Methods: The sensor system consisted of 2 optical fibers, a pedestal for fiber fixation, and a mirror for the laser reflection and was incorporated into a cerebrospinal fluid drainage catheter. In vivo studies were performed in a swine model (n = 10) to measure SCBF during spinal cord ischemia induced by clamping the descending thoracic aorta and supra-aortic neck vessels, when necessary. A temporary low cardiac output model was also created by inflow clamping of the inferior vena cava to analyze the quantitative changes in SCBF during this maneuver. Results: The developed SCBF monitoring catheter placed intrathecally could detect SCBF in all the swine. The SCBF after aortic crossclamping at the fourth intercostal level exhibited diverse changes reproducibly among the swine, with a >25% reduction in SCBF in 5 pigs, an increase in 3, and no significant changes in 2. Consistent reductions were recorded during inferior vena cava occlusion. The mean SCBF decreased by 32% after inferior vena cava occlusion when the cardiac output had decreased by 27%. Conclusions: We have developed a novel SCBF sensor that could detect real-time changes in spinal cord perfusion in a swine model. The device holds promise to detect imminent ischemia or ensure acceptable blood perfusion in the spinal cord and could further enhance our understanding of spinal cord circulation.
引用
收藏
页码:1726 / 1731
页数:6
相关论文
共 50 条
  • [41] ULTRASONIC PLETHYSMOGRAPHY - A NOVEL METHOD AND DEVICE FOR REAL-TIME, NONINVASIVE AND REPRODUCIBLE MONITORING OF BLOOD/FLUID VOLUME FLUCTUATIONS
    GAVISH, B
    INTERNATIONAL JOURNAL OF MICROCIRCULATION-CLINICAL AND EXPERIMENTAL, 1986, 5 (2-3): : 240 - 240
  • [42] Preventing Spinal Cord Injury in Single-Staged Juxtarenal and Thoracoabdominal Endovascular Aortic Aneurysm Repair: An Evaluation of Cerebrospinal Fluid Catheter Drainage
    Walter, Corinna
    Hirsch, Kornelia
    Heil, Sabine
    Taher, Fadi
    Falkensammer, Juergen
    Gustorff, Burkhard
    Assadian, Afshin
    JOURNAL OF ENDOVASCULAR THERAPY, 2024,
  • [43] A novel camera-based sensor for real-time wastewater quality monitoring
    Antonini, Giorgio
    Pearce, Joshua M.
    Berruti, Franco
    Santoro, Domenico
    WATER PRACTICE AND TECHNOLOGY, 2024, 19 (09) : 3778 - 3793
  • [44] Real-Time Monitoring of Bacteria Clearance From Blood in a Murine Model
    Harrington, Walter N.
    Nolan, Jacqueline
    Nedosekin, Dmitry A.
    Smeltzer, Mark S.
    Zharov, Vladimir P.
    CYTOMETRY PART A, 2020, 97 (07) : 706 - 712
  • [45] Evaluation of conventional and frozen elephant trunk techniques on spinal cord blood flow in an animal model
    Leontyev, Sergey
    Dieterlen, Maja-Theresa
    Halling, Michelle
    Schlegel, Franziska
    Hoyer, Alexandro
    Ossmann, Susann
    Klaeske, Kerstin
    Etz, Christian D.
    Mohr, FriedrichW.
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2017, 52 (03) : 485 - 491
  • [46] Quantitative measurements of spinal cord blood flow of an animal model of relapsing-remitting MS
    Tachrount, M.
    Davies, A.
    Rosianu, F.
    Desai, R.
    Thomas, D.
    Smith, K.
    Golay, X.
    JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2017, 37 : 66 - 67
  • [47] Fast blood flow monitoring in deep tissues with real-time software correlators
    Wang, Detian
    Parthasarathy, Ashwin B.
    Baker, Wesley B.
    Gannon, Kimberly
    Kavuri, Venki
    Ko, Tiffany
    Schenkel, Steven
    Li, Zhe
    Li, Zeren
    Mullen, Michael T.
    Detre, John A.
    Yodh, Arjun G.
    BIOMEDICAL OPTICS EXPRESS, 2016, 7 (03): : 776 - 797
  • [48] Percent positive rate of Lyme real-time polymerase chain reaction in blood, cerebrospinal fluid, synovial fluid, and tissue
    Babady, Ngolela Esther
    Sloan, Lynne M.
    Vetter, Emily A.
    Patel, Robin
    Binnicker, Matthew J.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2008, 62 (04) : 464 - 466
  • [49] A Continuous Opto-Electronic Sensor for Blood Pressure Monitoring with Real-Time System
    Kao, Yung-Hua
    Chao, Paul C. -P.
    Wey, Chin-Long
    PROCEEDINGS OF THE ASME 26TH ANNUAL CONFERENCE ON INFORMATION STORAGE AND PROCESSING SYSTEMS, 2017, 2017,
  • [50] Wireless enzyme sensor system for real-time monitoring of blood glucose levels in fish
    Endo, Hideaki
    Yonemori, Yuki
    Hibi, Kyoko
    Ren, Huifeng
    Hayashi, Tetsuhito
    Tsugawa, Wakako
    Sode, Koji
    BIOSENSORS & BIOELECTRONICS, 2009, 24 (05): : 1417 - 1423