Development of a real-time feedback algorithm for chest compression during CPR without assuming full chest decompression

被引:7
作者
Dellimore, Gohier K. H. [1 ]
Scheffer, C. [1 ]
机构
[1] Univ Stellenbosch, Dept Mech & Mechatron Engn, Biomed Engn Res Grp, ZA-7602 Matieland, South Africa
关键词
Clinical engineering; Therapeutic and diagnostic systems; Cardiopulmonary resuscitation; Active feedback system; Chest compression; Incomplete chest decompression; RESUSCITATION-COUNCIL GUIDELINES; CARDIOPULMONARY-RESUSCITATION; AUDIOVISUAL FEEDBACK; QUALITY ASSESSMENT; CARDIAC-ARREST; SUPPORT; DEPTH; PERFORMANCE; IMPROVEMENT; MANNEQUIN;
D O I
10.1016/j.resuscitation.2014.03.003
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: To evaluate the performance of a real-time feedback algorithm for chest compression (CC) during cardiopulmonary resuscitation (CPR), which provides accurate estimation of the CC depth based on dual accelerometer signal processing, without assuming full CDC. Also, to explore the influence of incomplete chest decompression (CDC) on the CC depth estimation performance. Methods: The performance of a real-time feedback algorithm for CC during CPR was evaluated by comparison with an offline algorithm using adult CPR manikin CC data obtained under various conditions. Results: The real-time algorithm, using non-causal baselining, delivered comparable CC depth estimation accuracy to the offline algorithm on both soft and hard back support surfaces. In addition, for both algorithms incomplete CDC led to underestimation of the CC depth. Conclusions: CPR feedback systems which utilize an assumption of full CDC may be unreliable especially in long duration CPR events where rescuer fatigue can strongly influence CC quality. In addition, these systems may increase the risk of thoracic and abdominal injury during CPR since rescuers may apply excessive compression forces due to underestimation of the CC depth when incomplete CDC occurs. Hence, there is a strong need for CPR feedback systems to accurately measure CDC in order to improve their clinical effectiveness. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:820 / 825
页数:6
相关论文
共 23 条
  • [1] Compression depth estimation for CPR quality assessment using DSP on accelerometer signals
    Aase, SO
    Myklebust, H
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (03) : 263 - 268
  • [2] CPR quality improvement during in-hospital cardiac arrest using a real-time audiovisual feedback system
    Abella, Benjamin S.
    Edelson, Dana P.
    Kim, Salem
    Retzer, Elizabeth
    Myklebust, Helge
    Barry, Anne M.
    O'Hearn, Nicholas
    Hoek, Terry L. Vanden
    Becker, Lance B.
    [J]. RESUSCITATION, 2007, 73 (01) : 54 - 61
  • [3] Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest
    Abella, BS
    Alvarado, JP
    Myklebust, H
    Edelson, DP
    Barry, A
    O'Hearn, N
    Vanden Hoek, TL
    Becker, LB
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2005, 293 (03): : 305 - 310
  • [4] Chest Compression Quality Over Time in Pediatric Resuscitations
    Badaki-Makun, Oluwakemi
    Nadel, Frances
    Donoghue, Aaron
    McBride, Michael
    Niles, Dana
    Seacrist, Thomas
    Maltese, Matthew
    Zhang, Xuemei
    Paridon, Stephen
    Nadkarni, Vinay M.
    [J]. PEDIATRICS, 2013, 131 (03) : E797 - E804
  • [5] Mechanics of cardiopulmonary resuscitation performed with the patient on a soft bed vs a hard surface
    Boe, JM
    Babbs, CF
    [J]. ACADEMIC EMERGENCY MEDICINE, 1999, 6 (07) : 754 - 757
  • [6] Improvement in timing and effectiveness of external cardiac compressions with a new non-invasive device: the CPR-Ezy
    Boyle, AJ
    Wilson, AM
    Connelly, K
    McGuigan, L
    Wilson, J
    Whitbourn, R
    [J]. RESUSCITATION, 2002, 54 (01) : 63 - 67
  • [7] The impact of backboard size and orientation on sternum-to-spine compression depth and compression stiffness in a manikin study of CPR using two mattress types
    Cloete, G.
    Dellimore, K. H.
    Scheffer, C.
    Smuts, M. S.
    Wallis, L. A.
    [J]. RESUSCITATION, 2011, 82 (08) : 1064 - 1070
  • [8] Cloete G, 2011, IEEE ENG MED BIO, P2484, DOI 10.1109/IEMBS.2011.6090689
  • [9] A modeling approach to the effects of force guided versus depth guided compression during cardiopulmonary resuscitation on different chests and back support surfaces
    Dellimore, K. H.
    Scheffer, C.
    [J]. RESUSCITATION, 2013, 84 (06) : 837 - 842
  • [10] Optimal chest compression in cardiopulmonary resuscitation depends upon thoracic and back support stiffness
    Dellimore, Kiran H.
    Scheffer, Cornie
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2012, 50 (12) : 1269 - 1278