Introduction of a novel ultrahigh sensitivity collimator for brain SPECT imaging

被引:2
|
作者
Park, Mi-Ae [1 ,2 ]
Kijewski, Marie Foley [1 ,2 ]
Keijzers, Ronnie [3 ]
Keijzers, Mark [3 ]
Lyon, Morgan C. [1 ,2 ]
Horky, Laura [1 ,2 ]
Moore, Stephen C. [1 ,2 ]
机构
[1] Brigham & Womens Hosp, Dept Radiol, 75 Francis St, Boston, MA 02115 USA
[2] Harvard Med Sch, Boston, MA 02115 USA
[3] Nucl Fields USA, Des Plaines, IA 60018 USA
基金
美国国家卫生研究院;
关键词
high sensitivity brain SPECT; collimator design; DAT imaging; Parkinson's disease;
D O I
10.1118/1.4958962
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Noise levels of brain SPECT images are highest in central regions, due to preferential attenuation of photons emitted from deep structures. To address this problem, the authors have designed a novel collimator for brain SPECT imaging that yields greatly increased sensitivity near the center of the brain without loss of resolution. This hybrid collimator consisted of ultrashort cone-beam holes in the central regions and slant-holes in the periphery (USCB). We evaluated this collimator for quantitative brain imaging tasks. Methods: Owing to the uniqueness of the USCB collimation, the hole pattern required substantial variations in collimator parameters. To utilize the lead-casting technique, the authors designed two supporting plates to position about 37 000 hexagonal, slightly tapered pins. The holes in the supporting plates were modeled to yield the desired focal length, hole length, and septal thickness. To determine the properties of the manufactured collimator and to compute the system matrix, the authors prepared an array of point sources that covered the entire detector area. Each point source contained 32 mu Ci of Tc-99m at the first scan time. The array was imaged for 5 min at each of the 64 shifted locations to yield a 2-mm sampling distance, and hole parameters were calculated. The sensitivity was also measured using a point source placed along the central ray at several distances from the collimator face. High-count projection data from a five-compartment brain phantom were acquired with the three collimators on a dual-head SPECT/CT system. The authors calculated Cramer-Rao bounds on the precision of estimates of striatal and background activity concentration. In order to assess the new collimation system to detect changes in striatal activity, the authors evaluated the precision of measuring a 5% decrease in right putamen activity. The authors also reconstructed images of projection data obtained by summing data from the individual phantom compartments. Results: The sensitivity of the novel cone-beam collimator varied with distance from the detector face; it was higher than that of the fan-beam collimator by factors ranging from 2.7 to 162. Examination of the projections of the point sources revealed that only a few holes were distorted or partially blocked, indicating that the intensive manual fabrication process was very successful. Better reconstructed phantom images were obtained from the USCB + FAN collimator pair than from either LEHR or FAN collimation. For the left caudate, located near the center of the brain, the detected counts were 9.8 (8.3) times higher for UCSB compared with LEHR (FAN), averaged over 60 views. The task-specific SNR for detecting a 5% decrease in putamen uptake was 7.4 for USCB and 3.2 for LEHR. Conclusions: The authors have designed and manufactured a novel collimator for brain SPECT imaging. The sensitivity is much higher than that of a fan-beam collimator. Because of differences between the manufactured collimator and its design, reconstruction of the data requires a measured system matrix. The authors have demonstrated the potential of USCB collimation for improved precision in estimating striatal uptake. The novel collimator may be useful for early detection of Parkinson's disease, and for monitoring therapy response and disease progression. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:4734 / 4741
页数:8
相关论文
共 34 条
  • [1] Performance of a novel collimator for high-sensitivity brain SPECT
    El Fakhri, G
    Ouyang, J
    Zimmerman, RE
    Fischman, AJ
    Kijewski, MF
    MEDICAL PHYSICS, 2006, 33 (01) : 209 - 215
  • [2] Multipinhole collimator with 20 apertures for a brain SPECT application
    Lee, Tzu-Cheng
    Ellin, Justin R.
    Huang, Qiu
    Shrestha, Uttam
    Gullberg, Grant T.
    Seo, Youngho
    MEDICAL PHYSICS, 2014, 41 (11)
  • [3] Collimator Design for a Brain SPECT/MRI Insert
    Salvado, Debora
    Erlandsson, Kjell
    Bousse, Alexandre
    Occhipinti, Michele
    Busca, Paolo
    Fiorini, Carlo
    Hutton, Brian F.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (04) : 1716 - 1724
  • [4] Nonuniform collimator sensitivity: Improved precision for quantitative SPECT
    Kijewski, MF
    Muller, SP
    Moore, SC
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (01) : 151 - 156
  • [5] Parallel-hole collimator concept for stationary SPECT imaging
    Pato, Lara R. V.
    Vandenberghe, Stefaan
    Zedda, Tiziana
    Van Holen, Roel
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (22): : 8791 - 8807
  • [6] Multi-pinhole collimator design in different numbers of projections for brain SPECT
    Huang, Wenbo
    Mok, Greta S. P.
    FRONTIERS IN MEDICINE, 2023, 10
  • [7] Design of a Multi-Pinhole Collimator for I-123 DaTscan Imaging on Dual-Headed SPECT Systems in Combination with a Fan-Beam Collimator
    King, Michael A.
    Mukherjee, Joyeeta M.
    Koenik, Arda
    Zubal, I. George
    Dey, Joyoni
    Licho, Robert
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2016, 63 (01) : 90 - 97
  • [8] PET and SPECT imaging of the brain: a review on the current status of nuclear medicine in Japan
    Kaneta, Tomohiro
    JAPANESE JOURNAL OF RADIOLOGY, 2020, 38 (04) : 343 - 357
  • [9] Brain SPECT as a Biomarker of Neurodegeneration in Dementia in the Era of Molecular Imaging: Still a Valid Option?
    Ferrando, Rodolfo
    Damian, Andres
    FRONTIERS IN NEUROLOGY, 2021, 12