Oxinium modular femoral heads do not reduce polyethylene wear in cemented total hip arthroplasty at five years

被引:32
作者
Jonsson, B. A. [1 ]
Kadar, T. [1 ]
Havelin, L. I. [1 ]
Haugan, K. [1 ]
Espehaug, B. [1 ]
Indrekvam, K. [2 ]
Furnes, O. [1 ]
Hallan, G. [3 ]
机构
[1] Haukeland Hosp, N-5021 Bergen, Norway
[2] Haukeland Hosp, Orthoped Clin, N-5021 Bergen, Norway
[3] Haukeland Hosp, Dept Orthopaed Surg, N-5021 Bergen, Norway
关键词
CROSS-LINKED POLYETHYLENE; MOLECULAR-WEIGHT POLYETHYLENE; OXIDIZED ZIRCONIUM; FOLLOW-UP; OSTEOLYSIS; PROSTHESES; DAMAGE; THA; RADIOSTEREOMETRY; ASSOCIATION;
D O I
10.1302/0301-620X.97B11.36137
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report the five-year outcome of a randomised controlled trial which used radiostereometric analysis (RSA) to assess the influence of surface oxidised zirconium (OxZr, Oxinium) on polyethylene wear in vivo. A total of 120 patients, 85 women and 35 men with a mean age of 70 years (59 to 80) who were scheduled for primary cemented total hip arthroplasty were randomly allocated to four study groups. Patients were blinded to their group assignment and received either a conventional polyethylene (CPE) or a highly cross-linked (HXL) acetabular component of identical design. On the femoral side patients received a 28 mm head made of either cobalt-chromium (CoCr) or OxZr. The proximal head penetration (wear) was measured with repeated RSA examinations over five years. Clinical outcome was measured using the Harris hip score. There was no difference in polyethylene wear between the two head materials when used with either of the two types of acetabular component (p = 0.3 to 0.6). When comparing the two types of polyethylene there was a significant difference in favour of HXLPE, regardless of the head material used (p <0.001). In conclusion, we found no advantage of OxZr over CoCr in terms of polyethylene wear after five years of follow-up. Our findings do not support laboratory results which have shown a reduced rate of wear with OxZr. They do however add to the evidence on the better resistance to wear of HXLPE over CPE.
引用
收藏
页码:1463 / 1469
页数:7
相关论文
共 39 条
[1]  
[Anonymous], ACTA ORTHOP
[2]   Arthroplasty options for the young patient [J].
Bourne, RB ;
Barrack, R ;
Rorabeck, CH ;
Salehi, A ;
Good, V .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (441) :159-167
[3]   Experimental assessment of precision and accuracy of radio stereometric analysis for the determination of polyethylene wear in a total hip replacement model [J].
Bragdon, CR ;
Malchau, H ;
Yuan, XH ;
Perinchief, R ;
Kärrholm, J ;
Börlin, N ;
Estok, DM ;
Harris, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (04) :688-695
[4]  
Charnley J, J Bone Joint Surg, V18r11972, P61
[5]   Highly cross-linked polyethylene in cemented THA -: Randomized study of 61 hips [J].
Digas, G ;
Kärrholm, J ;
Thanner, J ;
Malchau, H ;
Herberts, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :126-138
[6]   5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets -: Two randomized studies using radiostereometric analysis [J].
Digas, Georgios ;
Karrholm, Johan ;
Thanner, Jonas ;
Herberts, Peter .
ACTA ORTHOPAEDICA, 2007, 78 (06) :746-754
[7]   Clinical performance of a durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years [J].
Dorr, LD ;
Wan, ZN ;
Shahrdar, C ;
Sirianni, L ;
Boutary, M ;
Yun, A .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08) :1816-1821
[8]   A literature review of the association between wear rate and osteolysis in total hip arthroplasty [J].
Dumbleton, JH ;
Manley, MT ;
Edidin, AA .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :649-661
[9]   A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners [J].
Engh, C. Anderson, Jr. ;
Stepniewski, Adam S. ;
Ginn, Stuart D. ;
Beykirch, Sarah E. ;
Sychterz-Terefenko, Christi J. ;
Hopper, Robert H., Jr. ;
Engh, Charles A. .
JOURNAL OF ARTHROPLASTY, 2006, 21 (06) :17-25
[10]   18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register Concerns about some newer implants [J].
Espehaug, Birgitte ;
Furnes, Ove ;
Engesaeter, Lars B. ;
Havelin, Leif I. .
ACTA ORTHOPAEDICA, 2009, 80 (04) :402-412