Shielding stray magnetic fields of open high field MRI magnets

被引:0
作者
Kakugawa, S
Hino, N
Komura, A
Kitamura, M
Takeshima, H
Yatsuo, T
Tazaki, H
机构
[1] Hitachi Ltd, Hitachi Res Lab, Hitachi, Ibaraki 3191292, Japan
[2] Hitachi Med Corp, R&D Ctr, Kashiwa, Chiba 2770804, Japan
关键词
linear programming; magnetic resonance; imaging (MRI); multipole stray field; optimization;
D O I
10.1109/TASC.2004.831023
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed a new magnet for open high field (0.7 T) MRI systems. Since MRI systems are installed in hospital sites, it is important to reduce stray magnetic fields of the magnets. We studied passive, active and hybrid shielding methods and compared them. In order to study feasibility of active shielding, we considered higher order multipole stray fields. We present details of the calculation method and the results.
引用
收藏
页码:1639 / 1642
页数:4
相关论文
共 7 条
[1]   AN OPTIMAL-DESIGN TECHNIQUE FOR MRI SUPERCONDUCTING MAGNETS USING MATHEMATICAL-PROGRAMMING METHOD [J].
ISHIYAMA, A ;
YOKOI, T ;
TAKAMORI, S ;
ONUKI, T .
IEEE TRANSACTIONS ON MAGNETICS, 1988, 24 (02) :922-925
[2]   A study on optimal coil configurations in a split-type superconducting MRI magnet [J].
Kakugawa, S ;
Hino, N ;
Komura, A ;
Kitamura, M ;
Takeshima, H ;
Honmei, T .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) :366-369
[3]   AN OPTIMAL-DESIGN TECHNIQUE FOR COIL CONFIGURATIONS IN IRON-SHIELDED MRI MAGNETS [J].
KITAMURA, M ;
KAKUKAWA, S ;
MORI, K ;
TOMINAKA, T .
IEEE TRANSACTIONS ON MAGNETICS, 1994, 30 (04) :2352-2355
[4]   A DEVELOPMENT OF A PERMANENT-MAGNET ASSEMBLY FOR MRI DEVICES USING ND-FE-B MATERIAL [J].
MIYAMOTO, T ;
SAKURAI, H ;
TAKABAYASHI, H ;
AOKI, M .
IEEE TRANSACTIONS ON MAGNETICS, 1989, 25 (05) :3907-3909
[5]  
SAUZADE MD, 1973, ADV ELECT ELECT PHYS, V34, P42
[6]   SUPERCONDUCTING OPEN-CONFIGURATION MR-IMAGING SYSTEM FOR IMAGE-GUIDED THERAPY [J].
SCHENCK, JF ;
JOLESZ, FA ;
ROEMER, PB ;
CLINE, HE ;
LORENSEN, WE ;
KIKINIS, R ;
SILVERMAN, SG ;
HARDY, CJ ;
BARBER, WD ;
LASKARIS, ET ;
DORRI, B ;
NEWMAN, RW ;
HOLLEY, CE ;
COLLICK, BD ;
DIETZ, DP ;
MACK, DC ;
AINSLIE, MD ;
JASKOLSKI, PL ;
FIGUEIRA, MR ;
LEHN, JCV ;
SOUZA, SP ;
DUMOULIN, CL ;
DARROW, RD ;
STPETERS, RL ;
ROHLING, KW ;
WATKINS, RD ;
EISNER, DR ;
BLUMENFELD, SM ;
VOSBURGH, KG .
RADIOLOGY, 1995, 195 (03) :805-814
[7]   ACCURATE AND CONVENIENT EVALUATION METHOD OF MAGNETIC-FIELD [J].
TAKESHIMA, H .
IEEE TRANSACTIONS ON MAGNETICS, 1990, 26 (05) :2571-2573